Probe calibration for freehand 3-D ultrasound

被引:87
作者
Lindseth, F [1 ]
Tangen, GA [1 ]
Lango, T [1 ]
Bang, J [1 ]
机构
[1] SINTEF, Unimed Ultrasound, N-7465 Trondheim, Norway
关键词
utrasound probe calibration; freehand 3-D ultrasound reconstruction; ultrasound-based surgical navigation; image-guided therapy; phantom study; accuracy evaluation; automatic point detection; image processing;
D O I
10.1016/S0301-5629(03)01012-3
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Ultrasound (US) probe calibration establishes the rigid body transformation between the US image and a tracking device attached to the probe. This is an important requirement for correct 3-D reconstruction of freehand US images and, thus, for accurate surgical navigation based on US. In this study, we evaluated three methods for probe calibration, based on a single-point phantom, a wire-cross phantom requiring 2-D alignment and a wire phantom for freehand scanning. The processing of acquired data is fairly common to these methods and, to a great extent, based on automated procedures. The evaluation is based on quality measures in 2-D and 3-D reconstructed data. With each of the three methods, we calibrated a linear-array probe, a phased-array sector probe and an intraoperative probe. The freehand method performed best, with a 3-D navigation accuracy of 0.6 mm for one of the probes. This indicates that clinical accuracy in the order of I mm may be achieved in US-based surgical navigation.
引用
收藏
页码:1607 / 1623
页数:17
相关论文
共 30 条
  • [11] HOOK I, 2003, PROBE CALIBRATION 3D
  • [12] LANGO T, 2000, THESIS NORWEGIAN U S
  • [13] System for quantitative three-dimensional echocardiography of the left ventricle based on a magnetic-field position and orientation sensing system
    Legget, ME
    Leotta, DF
    Bolson, EL
    McDonald, JA
    Martin, RW
    Li, XN
    Otto, CM
    Sheehan, FH
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1998, 45 (04) : 494 - 504
  • [14] Performance of a miniature magnetic position sensor for three-dimensional ultrasound imaging
    Leotta, DF
    Detmer, PR
    Martin, RW
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1997, 23 (04) : 597 - 609
  • [15] LEOTTA DF, 1995, P IEEE ULTRASON S, V2, P1415
  • [16] A robust and automatic method for evaluating accuracy in 3-D ultrasound-based navigation
    Lindseth, F
    Bang, J
    Lango, T
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2003, 29 (10) : 1439 - 1452
  • [17] Lindseth Frank, 2002, Comput Aided Surg, V7, P197
  • [18] LIU J, 1998, P IEEE, V2, P839
  • [19] Beam calibration without a phantom for creating a 3-D freehand ultrasound system
    Muratore, DM
    Galloway, RL
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2001, 27 (11) : 1557 - 1566
  • [20] Pagoulatos N, 1999, IEEE Trans Inf Technol Biomed, V3, P278, DOI 10.1109/4233.809172