Probe calibration for freehand 3-D ultrasound

被引:87
作者
Lindseth, F [1 ]
Tangen, GA [1 ]
Lango, T [1 ]
Bang, J [1 ]
机构
[1] SINTEF, Unimed Ultrasound, N-7465 Trondheim, Norway
关键词
utrasound probe calibration; freehand 3-D ultrasound reconstruction; ultrasound-based surgical navigation; image-guided therapy; phantom study; accuracy evaluation; automatic point detection; image processing;
D O I
10.1016/S0301-5629(03)01012-3
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Ultrasound (US) probe calibration establishes the rigid body transformation between the US image and a tracking device attached to the probe. This is an important requirement for correct 3-D reconstruction of freehand US images and, thus, for accurate surgical navigation based on US. In this study, we evaluated three methods for probe calibration, based on a single-point phantom, a wire-cross phantom requiring 2-D alignment and a wire phantom for freehand scanning. The processing of acquired data is fairly common to these methods and, to a great extent, based on automated procedures. The evaluation is based on quality measures in 2-D and 3-D reconstructed data. With each of the three methods, we calibrated a linear-array probe, a phased-array sector probe and an intraoperative probe. The freehand method performed best, with a 3-D navigation accuracy of 0.6 mm for one of the probes. This indicates that clinical accuracy in the order of I mm may be achieved in US-based surgical navigation.
引用
收藏
页码:1607 / 1623
页数:17
相关论文
共 30 条
  • [1] LEAST-SQUARES FITTING OF 2 3-D POINT SETS
    ARUN, KS
    HUANG, TS
    BLOSTEIN, SD
    [J]. IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 1987, 9 (05) : 699 - 700
  • [2] Three-dimensional freehand ultrasound: Image reconstruction and volume analysis
    Barry, CD
    Allott, CP
    John, NW
    Mellor, PM
    Arundel, PA
    Thomson, DS
    Waterton, JC
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1997, 23 (08) : 1209 - 1224
  • [3] Dynamic three-dimensional freehand echocardiography using raw digital ultrasound data
    Berg, S
    Torp, H
    Martens, D
    Steen, E
    Samstad, S
    Hoivik, I
    Olstad, B
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1999, 25 (05) : 745 - 753
  • [4] Blackall JM, 2000, LECT NOTES COMPUT SC, V1935, P462
  • [5] Calibration of three-dimensional ultrasound images for image-guided radiation therapy
    Bouchet, LG
    Meeks, SL
    Goodchild, G
    Bova, FJ
    Buatti, JM
    Friedman, WA
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (02) : 559 - 577
  • [6] 3D ULTRASONIC IMAGE FEATURE LOCALIZATION BASED ON MAGNETIC SCANHEAD TRACKING - IN-VITRO CALIBRATION AND VALIDATION
    DETMER, PR
    BASHEIN, G
    HODGES, T
    BEACH, KW
    FILER, EP
    BURNS, DH
    STRANDNESS, DE
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1994, 20 (09) : 923 - 936
  • [7] Duck A., 1990, PHYS PROPERTIES TISS
  • [8] Gobbi DG, 1999, LECT NOTES COMPUT SC, V1679, P920
  • [9] SonoWand, an ultrasound-based neuronavigation system
    Gronningsaeter, A
    Kleven, A
    Ommedal, S
    Aarseth, TE
    Lie, T
    Lindseth, F
    Lango, T
    Unsgard, G
    [J]. NEUROSURGERY, 2000, 47 (06) : 1373 - 1379
  • [10] HARTOV A, 1999, NEUROSURG FOCUS, V6