Problems with the estimation of urine protein by automated assays

被引:26
作者
Dube, J
Girouard, J
Leclerc, P
Douville, P
机构
[1] CHUS Fleurimont, Fleurimont, PQ J1H 5N4, Canada
[2] Hop Enfants Jesus, CHA, Serv Biochim, Quebec City, PQ G1J 1Z4, Canada
[3] Hop St Sacrement, CHA, Quebec City, PQ G1S 4L8, Canada
[4] Hop Hotel Dieu, CHUQL, Serv Biochim, Quebec City, PQ G1R 2J6, Canada
关键词
urine; protein; measurement; automated; comparison; bias; precision; linearity;
D O I
10.1016/j.clinbiochem.2004.12.010
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: Most clinical laboratories replaced their manual precipitation techniques for the determination of urinary protein with automated dye binding assays or benzethonium chloride-turbidimetric assays. Few studies have validated these assays for the measurement of urinary proteins in the normal range. Design and methods: This study compares four automated assays for the measurement of urinary protein to a manual Ponceau S/TCA precipitation assay. We evaluated the linearity, the precision, the analytical sensitivity, the accuracy and the recovery of different proteins for each assay. Results: All assays showed good linearity with the theoretical concentration of albumin present in the sample. The coefficient of variation was below 10% at a concentration of 0.142 g/L. However, the manual Ponceau S/TCA assay demonstrated superior analytical sensitivity. Accuracy determinations showed a variable positive bias and poor correlations at concentrations below 0.1 g/L when compared to the Ponceau S/TCA assay. Small molecular weight peptides particularly affected the pyrogallol red assays but other urinary components also interfered with the automated assays. Conclusions: Most automated assays show high imprecision and poor accuracy for the measurement of urinary protein in the normal range. The Ponceau S/TCA offers a precise and accurate manual alternative to these automated assays. (c) 2005 The Canadian Society of Clinical Chemists. All rights reserved.
引用
收藏
页码:479 / 485
页数:7
相关论文
共 20 条
[1]   IMPROVED BETWEEN-LABORATORY AGREEMENT FOR SPECIFIC PROTEIN ASSAYS IN SERUM FOLLOWING INTRODUCTION OF A COMMON REFERENCE PREPARATION (SPS-01) DEMONSTRATED IN AN EXTERNAL QUALITY ASSESSMENT SCHEME [J].
CHAMBERS, RE ;
BULLOCK, DG ;
WHICHER, JT .
CLINICA CHIMICA ACTA, 1987, 164 (02) :189-200
[2]   URINARY TOTAL PROTEIN ESTIMATION - FACT OR FICTION [J].
CHAMBERS, RE ;
BULLOCK, DG ;
WHICHER, JT .
NEPHRON, 1989, 53 (01) :33-36
[3]  
DILENA BA, 1983, CLIN CHEM, V29, P553
[4]  
GLENN GC, 1980, AM J CLIN PATHOL, V74, P531
[5]   Markers of diabetic nephropathy [J].
Hong, CY ;
Chia, KS .
JOURNAL OF DIABETES AND ITS COMPLICATIONS, 1998, 12 (01) :43-60
[6]  
IWATA J, 1979, CLIN CHEM, V25, P1317
[7]  
KAWAKAMI H, 1990, CLIN NEPHROL, V33, P232
[8]   PROTEINURIA [J].
KIM, MS .
CLINICS IN LABORATORY MEDICINE, 1988, 8 (03) :527-540
[9]  
LEMANN J, 1987, CLIN CHEM, V33, P297
[10]  
Lynch PLM, 1998, CLIN CHEM, V44, P674