Comparison of the non-invasive Nexfin® monitor with conventional methods for the measurement of arterial blood pressure in moderate risk orthopaedic surgery patients

被引:37
作者
Balzer, Felix [1 ,2 ]
Habicher, Marit [1 ,2 ]
Sander, Michael [3 ]
Sterr, Julian [4 ]
Scholz, Stephanie [1 ,2 ]
Feldheiser, Aarne [1 ,2 ]
Mueller, Michael [2 ,5 ]
Perka, Carsten [2 ,5 ]
Treskatsch, Sascha [1 ,2 ]
机构
[1] Charite, Campus Charite Mitte, Dept Anaesthesiol & Intens Care Med, Charitepl, Berlin, Germany
[2] Charite, Campus Virchow Klinikum, Charitepl, Berlin, Germany
[3] Univ Klinikum Giessen & Marburg GmbH, Dept Anaesthesiol & Intens Care Med, Rudolf Buchheim Str, Giessen, Germany
[4] Klinikum Starnberg, Dept Internal Med, Osswaldstr, Starnberg, Germany
[5] Charite, Campus Charite Mitte, Ctr Musculoskeletal Surg, Charitepl, Berlin, Germany
关键词
Blood pressure; monitoring; non-invasive arterial pressure; invasive arterial pressure; Nexfin (R); UPPER-ARM; FINGER; MORTALITY; CUFF; VALIDATION; MANAGEMENT; AGREEMENT; ACCURACY; FOREARM; IMPACT;
D O I
10.1177/0300060516635383
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Objective Continuous invasive arterial blood pressure (IBP) monitoring remains the gold standard for BP measurement, but traditional oscillometric non-invasive intermittent pressure (NIBP) measurement is used in most low-to-moderate risk procedures. This study compared non-invasive continuous arterial BP measurement using a Nexfin (R) monitor with NIBP and IBP monitors. Methods This was a single-centre, prospective, pilot study in patients scheduled for elective orthopaedic surgery. Systolic BP, diastolic BP and mean arterial blood pressure (MAP) were measured by Nexfin (R), IBP and NIBP at five intraoperative time-points. Pearson correlation coefficients, Bland-Altman plots and trending ability of Nexfin (R) measurements were used as criteria for success in the investigation of measurement reliability. Results A total of 20 patients were enrolled in the study. For MAP, there was a sufficient correlation between IBP/Nexfin (R) (Pearson=0.75), which was better than the correlation between IBP/NIBP (Pearson=0.70). Bland-Altman analysis of the data showed that compared with IBP, there was a higher percentage error for MAP(NIBP) (30%) compared with MAP(Nexfin)(R) (27%). Nexfin (R) and NIBP underestimated systolic BP; NIBP also underestimated diastolic BP and MAP. Trending ability for MAP(Nexfin)(R) and MAP(NIBP) were comparable to IBP. Conclusion Non-invasive BP measurement with Nexfin (R) was comparable with IBP and tended to be more precise than NIBP.
引用
收藏
页码:832 / 843
页数:12
相关论文
共 43 条
  • [21] Models of brachial to finger pulse wave distortion and pressure decrement
    Gizdulich, P
    Prentza, A
    Wesseling, KH
    [J]. CARDIOVASCULAR RESEARCH, 1997, 33 (03) : 698 - 705
  • [22] Validation of a conical cuff on the forearm for estimating radial artery blood pressure
    Hersh, Lawrence T.
    Sesing, Jean C.
    Luczyk, William J.
    Friedman, Bruce A.
    Zhou, Shanggen
    Batchelder, Paul B.
    [J]. BLOOD PRESSURE MONITORING, 2014, 19 (01) : 38 - 45
  • [23] Assessment of agreement and trending between haemodynamic monitors is still challenging
    Hunsicker, Oliver
    Feldheiser, Aarne
    Wernecke, Klaus-Dieter
    Spies, Claudia
    [J]. INTENSIVE CARE MEDICINE, 2014, 40 (05) : 767 - 767
  • [24] Fifteen years experience with finger arterial pressure monitoring: assessment of the technology
    Imholz, BPM
    Wieling, W
    van Montfrans, GA
    Wesseling, KH
    [J]. CARDIOVASCULAR RESEARCH, 1998, 38 (03) : 605 - 616
  • [25] 2014 ESC/ESA Guidelines on non-cardiac surgery: cardiovascular assessment and management The Joint Task Force on non-cardiac surgery: cardiovascular assessment and management of the European Society of Cardiology (ESC) and the European Society of Anaesthesiology (ESA)
    Kristensen, Steen Dalby
    Knuuti, Juhani
    Saraste, Antti
    Anker, Stefan
    Botker, Hans Erik
    De Hert, Stefan
    Ford, Ian
    Gonzalez-Juanatey, Jose Ramon
    Gorenek, Bulent
    Heyndrickx, Guy Robert
    Hoeft, Andreas
    Huber, Kurt
    Iung, Bernard
    Kjeldsen, Keld Per
    Longrois, Dan
    Luescher, Thomas F.
    Pierard, Luc
    Pocock, Stuart
    Price, Susanna
    Roffi, Marco
    Sirnes, Per Anton
    Sousa-Uva, Miguel
    Voudris, Vasilis
    Funck-Brentano, Christian
    [J]. EUROPEAN HEART JOURNAL, 2014, 35 (35) : 2383 - 2431
  • [26] Noninvasive Continuous Arterial Blood Pressure Monitoring with Nexfin®
    Martina, Jerson R.
    Westerhof, Berend E.
    van Goudoever, Jeroen
    de Beaumont, Edouard M. F. H.
    Truijen, Jasper
    Kim, Yu-Sok
    Immink, Rogier V.
    Jobsis, Dorothea A.
    Hollmann, Markus W.
    Lahpor, Jaap R.
    de Mol, Bas A. J. M.
    van Lieshout, Johannes J.
    [J]. ANESTHESIOLOGY, 2012, 116 (05) : 1092 - 1103
  • [27] Low Blood Pressure Is Associated with Cognitive Impairment in Very Old People
    Molander, Lena
    Gustafson, Yngve
    Lovheim, Hugo
    [J]. DEMENTIA AND GERIATRIC COGNITIVE DISORDERS, 2010, 29 (04) : 335 - 341
  • [28] PENAZ J, 1992, PHYSIOL RES, V41, P5
  • [29] Predicting Hospital Mortality and Analysis of Long-Term Survival After Major Noncardiac Complications in Cardiac Surgery Patients
    Rahmanian, Parwis B.
    Adams, David H.
    Castillo, Javier G.
    Carpentier, Alain
    Filsoufi, Farzan
    [J]. ANNALS OF THORACIC SURGERY, 2010, 90 (04) : 1221 - 1229
  • [30] Association of intraoperative hypotension and pulmonary hypertension with adverse outcomes after orthotopic liver transplantation
    Reich, DL
    Wood, RK
    Emre, S
    Bodian, CA
    Hossain, S
    Krol, M
    Feierman, D
    [J]. JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2003, 17 (06) : 699 - 702