Automated applicator digitization for high-dose-rate cervix brachytherapy using image thresholding and density-based clustering

被引:14
作者
Deufel, Christopher L. [1 ]
Tian, Shulan [2 ]
Yan, Benjamin B. [1 ]
Vaishnav, Birjoo D. [1 ]
Haddock, Michael G. [1 ]
Petersen, Ivy A. [1 ]
机构
[1] Mayo Clin, Dept Radiat Oncol, Rochester, MN 55905 USA
[2] Mayo Clin, Div Biomed Stat & Informat, Dept Hlth Sci Res, Rochester, MN USA
关键词
Automation; Digitization; Segmentation; Brachytherapy; HDR; Tandem and ovoids; Cervix; Threshold; Clustering; RECONSTRUCTION-UNCERTAINTIES; AUTORECONSTRUCTION;
D O I
10.1016/j.brachy.2019.09.002
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
PURPOSE: The purpose of the study was to develop and evaluate an automated digitization algorithm for high-dose-rate cervix brachytherapy, with the goal of reducing the duration of treatment planning, staff resources, variability, and potential for human error. METHODS: An automated digitization algorithm was developed and retrospectively evaluated using treatment planning data from 10 patients with cervix cancer who were treated with a titanium tandem and ovoids applicator set. Applicators were segmented, without human interaction, by thresholding CT images to isolate high-density voxels and assigning the voxels to applicator and nonapplicator structures using HDBSCAN, a density-based linkage clustering algorithm. The applicator contours were determined from the centroid of the clustered voxels on each image slice and written to a treatment plan file. Automated contours were evaluated against manual digitization using distance and dosimetric metrics. RESULTS: A close agreement between automatic and manual digitization was observed. The mean magnitude of contour disagreement for 10 patients equaled 0.3 mm. Hausdorff distances were <= 1.0 mm. The applicator tip coordinates had submillimeter agreement. The median and mean dose volume histogram parameter differences were less than or equal to 1% for high-risk clinical target volume D-90, high-risk clinical target volume D-95, bladder D-2cc , rectum D-2cc, large bowel D-2cc, and small bowel D-2cc. The average execution time for the automated algorithm was less than 30 s. CONCLUSION: The digitization of titanium tandem and ovoids applicators for high-dose-rate brachytherapy treatment planning can be automated using straightforward thresholding and clustering algorithms. The adoption of automated digitization is expected to improve the consistency of treatment plans and reduce the duration of treatment planning. (C) 2020 The Authors. Published by Elsevier Inc. on behalf of American Brachytherapy Society.
引用
收藏
页码:111 / 118
页数:8
相关论文
共 17 条
[1]  
[Anonymous], 2019, PYTH SCI LECT NOT
[2]   Investigating the use of image thresholding in brachytherapy catheter reconstruction [J].
Binny, Diana ;
Lancaster, Craig M. ;
Kairn, Tanya ;
Trapp, Jamie V. ;
Back, Phil ;
Cheuk, Robyn ;
Crowe, Scott B. .
AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2016, 39 (04) :913-919
[3]   Automated construction of an intraoperative high-dose-rate treatment plan library for the Varian brachytherapy treatment planning system [J].
Deufel, Christopher L. ;
Furutani, Keith M. ;
Dahl, Robert A. ;
Haddock, Michael G. .
BRACHYTHERAPY, 2016, 15 (04) :531-536
[4]   Development and evaluation of an automatic interstitial catheter digitization tool for adaptive high-dose-rate brachytherapy [J].
Dise, Joseph ;
Liang, Xing ;
Scheuermann, Joshua ;
Anamalayil, Shibu ;
Mesina, Carmen ;
Lin, Lilie L. ;
Teo, Boon-Keng Kevin .
BRACHYTHERAPY, 2015, 14 (05) :619-625
[5]  
Ford NL, 2014, J BIOMED SCI ENG, V7, P343
[6]   Automated High-Dose-Rate Surface Brachytherapy Treatment Planning for Complex Head and Neck Cases with 3D-Printable Masks [J].
Guthier, C. V. ;
O'Farrell, D. A. ;
Bhagwat, M. S. ;
Devlin, P. M. ;
Cormack, R. A. ;
Buzurovic, I. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2018, 102 (03) :S54-S55
[7]   Automated calculation of point A coordinates for CT-based high-dose-rate brachytherapy of cervical cancer [J].
Kang, Hyejoo ;
Padilla, Laura ;
Hasan, Yasmin ;
Ai-Hallaq, Hania .
JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2017, 9 (04) :354-358
[8]   Understanding human management of automation errors [J].
McBride, Sara E. ;
Rogers, Wendy A. ;
Fisk, Arthur D. .
THEORETICAL ISSUES IN ERGONOMICS SCIENCE, 2014, 15 (06) :545-577
[9]   Workflow efficiency for the treatment planning process in CT-guided high-dose-rate brachytherapy for cervical cancer [J].
Michaud, Anthony L. ;
Benedict, Stanley ;
Montemayor, Eliseo ;
Hunt, Jon Paul ;
Wright, Cari ;
Mathai, Mathew ;
Mayadev, Jyoti S. .
BRACHYTHERAPY, 2016, 15 (05) :578-583
[10]   Catheter autoreconstruction in computed tomography based brachytherapy treatment planning [J].
Milickovic, N ;
Giannouli, S ;
Baltas, D ;
Lahanas, M ;
Kolotas, C ;
Zamboglou, N ;
Uzunoglu, N .
MEDICAL PHYSICS, 2000, 27 (05) :1047-1057