Do patient-specific instruments (PSI) for UKA allow non-expert surgeons to achieve the same saw cut accuracy as expert surgeons?

被引:20
作者
Jones, Gareth G. [1 ]
Logishetty, K. [1 ]
Clarke, S. [1 ]
Collins, R. [1 ]
Jaere, M. [1 ]
Harris, S. [1 ]
Cobb, J. P. [1 ]
机构
[1] Imperial Coll London, MSk Lab, Charing Cross Hosp, 7th Floor Lab Block, London W6 8RF, England
关键词
Unicompartmental knee arthroplasty; Partial knee replacement; Patient-specific instrumentation; PSI; Patient-specific guides; 3D printing; UNICOMPARTMENTAL KNEE ARTHROPLASTY; NATIONAL JOINT REGISTRY; COMPONENT POSITION; MATCHED PATIENTS; OXFORD UKA; REPLACEMENT; OUTCOMES; ENGLAND; WALES;
D O I
10.1007/s00402-018-3031-9
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction High-volume unicompartmental knee arthroplasty (UKA) surgeons have lower revision rates, in part due to improved intra-operative component alignment. This study set out to determine whether PSI might allow non-expert surgeons to achieve the same level of accuracy as expert surgeons. Materials and methods Thirty-four surgical trainees with no prior experience of UKA, and four high-volume UKA surgeons were asked to perform the tibial saw cuts for a medial UKA in a sawbone model using both conventional and patient-specific instrumentation (PSI) with the aim of achieving a specified pre-operative plan. Half the participants in each group started with conventional instrumentation, and half with PSI. CT scans of the 76 cut sawbones were then segmented and reliably orientated in space, before saw cut position in the sagittal, coronal and axial planes was measured, and compared to the pre-operative plan. Results The compound error (absolute error in the coronal, sagittal and axial planes combined) for experts using conventional instruments was significantly less than that of the trainees (11.6 degrees +/- 4.0 degrees v 7.7 degrees +/- 2.3 degrees, p = 0.029). PSI improved trainee accuracy to the same level as experts using conventional instruments (compound error 5.5 degrees +/- 3.4 degrees v 7.7 degrees +/- 2.3 degrees, p = 0.396) and patient-specific instruments (compound error 5.5 degrees +/- 3.4 degrees v 7.3 degrees +/- 4.1 degrees, p = 0.3). PSI did not improve the accuracy of high-volume surgeons (p = 0.3). Conclusions In a sawbone model, PSI allowed inexperienced surgeons to achieve more accurate saw cuts, equivalent to expert surgeons, and thus has the potential to reduce revision rates. The next test will be to determine whether these results can be replicated in a clinical trial.
引用
收藏
页码:1601 / 1608
页数:8
相关论文
共 37 条
  • [31] Three-dimensional finite element analysis of unicompartmental knee arthroplasty - the influence of tibial component inclination
    Sawatari, T
    Tsumura, H
    Iesaka, K
    Furushiro, Y
    Torisu, T
    [J]. JOURNAL OF ORTHOPAEDIC RESEARCH, 2005, 23 (03) : 549 - 554
  • [32] Tibial Loading After UKA: Evaluation of Tibial Slope, Resection Depth, Medial Shift and Component Rotation
    Small, Scott R.
    Berend, Michael E.
    Rogge, Renee D.
    Archer, Derek B.
    Kingman, Amanda L.
    Ritter, Merrill A.
    [J]. JOURNAL OF ARTHROPLASTY, 2013, 28 (09) : 179 - 183
  • [33] Benefits of using customized instrumentation in total knee arthroplasty: results from an activity-based costing model
    Tibesku, Carsten O.
    Hofer, Pamela
    Portegies, Wesley
    Ruys, C. J. M.
    Fennema, Peter
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2013, 133 (03) : 405 - 411
  • [34] The influence of different sets of surgical instrumentation in Oxford UKA on bearing size and component position
    Walker, Tilman
    Heinemann, Pascal
    Bruckner, Thomas
    Streit, Marcus R.
    Kinkel, Stefan
    Gotterbarm, Tobias
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2017, 137 (07) : 895 - 902
  • [35] Unicondylar knee arthroplasty in the UK National Health Service: An analysis of candidacy, outcome and cost efficacy
    Willis-Owen, Charles A.
    Brust, Klaus
    Alsop, Helen
    Miraldo, Marisa
    Cobb, Justin P.
    [J]. KNEE, 2009, 16 (06) : 473 - 478
  • [36] Surgeon's experience influences UKA survivorship: a comparative study between all-poly and metal back designs
    Zambianchi, F.
    Digennaro, V.
    Giorgini, A.
    Grandi, G.
    Fiacchi, F.
    Mugnai, R.
    Catani, F.
    [J]. KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2015, 23 (07) : 2074 - 2080
  • [37] The learning curve for minimally invasive Oxford phase 3 unicompartmental knee arthroplasty: cumulative summation test for learning curve (LC-CUSUM)
    Zhang, Qidong
    Zhang, Qian
    Guo, Wanshou
    Liu, Zhaohui
    Cheng, Liming
    Yue, Debo
    Zhang, Nianfei
    [J]. JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2014, 9