Laboratory and field evaluation of an improved glass honeycomb denuder filter pack sampler

被引:44
作者
Sioutas, C [1 ]
Wang, PY [1 ]
Ferguson, ST [1 ]
Koutrakis, P [1 ]
Mulik, JD [1 ]
机构
[1] US EPA,RES TRIANGLE PK,NC 27711
关键词
acid aerosols; honeycomb denuder; acid gases; ammonia; sampling losses;
D O I
10.1016/1352-2310(95)00324-X
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
A new and improved glass honeycomb denuder/filter pack sampler has been designed to measure atmospheric particles and gases. The new sampler combines the features of a previously developed glass honeycomb denuder sampler with a new inlet that minimizes loss of sampled gases on the inlet surfaces. Another novelty of the improved design is that the honeycomb denuders are entirely made of glass, instead of consisting of a large number of tubes sealed in an outer glass tube with epoxy resin. The new inlet was designed to reduce the residence time of the air sample as well as the overall surface area available for gas reaction. Although inlets of various materials were tested, the PTFE Teflon coated inlet was found to minimize loss of sampled gases. The honeycomb sampler with the improved inlet was characterized in two different field studies. Outdoor concentrations of nitric acid (HNO3), nitrous acid (HONO), ammonia (NH3), Sulfur dioxide (SO2), and fine-particle sulfate (SO42-), ammonium (NH4+), strong acidity (H+) and fine particulate mass determined using the honeycomb denuder sampler (HDS) were compared to those determined with a collocated Harvard/EPA annular denuder system (HEADS). The average collection of the HEADS sampler was in excellent agreement (e.g. within +/- 10% or less) with that of the honeycomb sampler for all the gas and particle phase species in the comparison studies. Results from the laboratory and field tests suggest that the new honeycomb denuder sampler with the improved inlet is suitable for sampling atmospheric particles and gases. In addition, it has the advantage of being more compact than annular denuder/filter pack systems, which makes the sampler more practical for large-scale monitoring studies.
引用
收藏
页码:885 / 895
页数:11
相关论文
共 18 条
[1]  
APPEL PB, 1988, ATMOS ENVIRON, V11, P2535
[2]   TUNGSTIC ACID FOR PRE-CONCENTRATION AND DETERMINATION OF GASEOUS AND PARTICULATE AMMONIA AND NITRIC-ACID IN AMBIENT AIR [J].
BRAMAN, RS ;
SHELLEY, TJ ;
MCCLENNY, WA .
ANALYTICAL CHEMISTRY, 1982, 54 (03) :358-364
[3]   EVALUATION OF THE GAS COLLECTION OF AN ANNULAR DENUDER SYSTEM UNDER SIMULATED ATMOSPHERIC CONDITIONS [J].
BRAUER, M ;
KOUTRAKIS, P ;
WOLFSON, JM ;
SPENGLER, JD .
ATMOSPHERIC ENVIRONMENT, 1989, 23 (09) :1981-1986
[4]   METHOD FOR DETERMINATION OF ATMOSPHERIC AMMONIA [J].
FERM, M .
ATMOSPHERIC ENVIRONMENT, 1979, 13 (10) :1385-1393
[5]  
FOREST J, 1982, ATMOS ENVIRON, V16, P1473
[6]   DEVELOPMENT AND EVALUATION OF A GLASS HONEYCOMB DENUDER FILTER PACK SYSTEM TO COLLECT ATMOSPHERIC GASES AND PARTICLES [J].
KOUTRAKIS, P ;
SIOUTAS, C ;
FERGUSON, ST ;
WOLFSON, JM ;
MULIK, JD ;
BURTON, RM .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1993, 27 (12) :2497-2501
[7]   EVALUATION OF AN ANNULAR DENUDER FILTER PACK SYSTEM TO COLLECT ACIDIC AEROSOLS AND GASES [J].
KOUTRAKIS, P ;
WOLFSON, JM ;
SLATER, JL ;
BRAUER, M ;
SPENGLER, JD ;
STEVENS, RK ;
STONE, CL .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1988, 22 (12) :1463-1468
[8]   DETERMINATION OF AEROSOL STRONG ACIDITY LOSSES DUE TO INTERACTIONS OF COLLECTED PARTICLES - RESULTS FROM LABORATORY AND FIELD STUDIES [J].
KOUTRAKIS, P ;
THOMPSON, KM ;
WOLFSON, JM ;
SPENGLER, JD ;
KEELER, GJ ;
SLATER, JL .
ATMOSPHERIC ENVIRONMENT PART A-GENERAL TOPICS, 1992, 26 (06) :987-995
[9]   DESIGN OF A GLASS IMPACTOR FOR AN ANNULAR DENUDER FILTER PACK SYSTEM [J].
KOUTRAKIS, P ;
WOLFSON, JM ;
BRAUER, M ;
SPENGLER, JD .
AEROSOL SCIENCE AND TECHNOLOGY, 1990, 12 (03) :607-612
[10]   CHARACTERISTICS OF LAMINAR JET IMPACTORS [J].
MARPLE, VA ;
LIU, BYH .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1974, 8 (07) :648-654