Comparison of central corneal thickness measurements using optical low-coherence reflectometry, Fourier domain optical coherence tomography, and Scheimpflug camera

被引:9
作者
Gonul, Saban [1 ]
Koktekir, Bengu Ekinci [1 ]
Bakbak, Berker [1 ]
Gedik, Sansal [1 ]
机构
[1] Selcuk Univ, Fac Med, Dept Ophthalmol, Konya, Turkey
关键词
Cornea/anatomy & histology; Corneal topography/methods; Tomography; optical coherence/methods; Fourier analysis; Comparative study; Observer variation; REPEATABILITY; PACHYMETRY; REPRODUCIBILITY; AGREEMENT; PENTACAM; INTRAOBSERVER; PHOTOGRAPHY; BIOMETRY; ORBSCAN; SYSTEM;
D O I
10.5935/0004-2749.20140087
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To compare the results of central corneal thickness (CCT) measurements obtained using optical low-coherence reflectometry (OLCR), Fourier domain optical coherence tomography (FD-OCT), and a Scheimpflug camera (SC), combined with Placido corneal topography. Methods: A total of 25 healthy subjects were enrolled in the present study, and one eye of each subject was included. A detailed ophthalmic examination was performed in all cases following CCT measurements with OLCR, FD-OCT, and SC. The results were compared using an ANOVA test. Bland-Altman analysis was used to demonstrate agreement between methods. Intra-examiner repeatability was assessed by using intraclass correlation coefficients (ICCs). Results: Statistically significant differences were observed between the results of the CCT measurements obtained using the three different devices (p=0.009). Significant correlations were found between OLCR and FD-OCT (r=0.97; p<0.0001), FD-OCT and SC (r=0.91; p<0.0001), and OLCR and SC (r=0.95; p<0.0001). The 95% limits of agreement (LOA) obtained from Bland-Altman plots were from -7.2 mu m to 28.7 mu m for OLCR versus FD-OCT, from -19.2 mu m to 30.4 mu m for OLCR versus SC, and from -42.6 to 32.3 mu m for FD-OCT versus SC. Intra-examiner repeatability was excellent for each method, with ICCs >0.98. Conclusions: Although the results of CCT measurements obtained from these three devices were highly correlated with one another and the mean differences between instruments were comparable with the reported diurnal CCT fluctuation, the measurements are not directly interchangeable in clinical practice because of the wide LOA values.
引用
收藏
页码:345 / 350
页数:6
相关论文
共 28 条
[1]   Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry [J].
Barkana, Y ;
Gerber, Y ;
Elbaz, U ;
Schwartz, S ;
Ken-Dror, G ;
Avni, I ;
Zadok, D .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (09) :1729-1735
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   Comparison of Oculus Pentacam and Stratus Optical Coherence Tomography for Measurement of Central Corneal Thickness [J].
Ceylan, Osman Melih ;
Turk, Adem ;
Erdurman, Cuneyt ;
Mumcuoglu, Tarkan ;
Erdem, Uzeyir ;
Gokce, Gokcen ;
Dagli, Selim .
CORNEA, 2011, 30 (06) :670-674
[4]   Measurement of central corneal thickness by high-resolution Scheimpflug imaging, Fourier-domain optical coherence tomography and ultrasound pachymetry [J].
Chen, Shihao ;
Huang, Jinhai ;
Wen, Daizong ;
Chen, Wuhe ;
Huang, David ;
Wang, Qinmei .
ACTA OPHTHALMOLOGICA, 2012, 90 (05) :449-455
[5]   Scheimpflug-Placido topographer and optical low-coherence reflectometry biometer: Repeatability and agreement [J].
Chen, Wuhe ;
McAlinden, Colm ;
Pesudovs, Konrad ;
Wang, Qinmei ;
Lu, Fan ;
Feng, Yifan ;
Chen, Jie ;
Huang, Jinhai .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (09) :1626-1632
[6]   Evaluation of the Lenstar LS 900 non-contact biometer [J].
Cruysberg, L. P. J. ;
Doors, M. ;
Verbakel, F. ;
Berendschot, T. T. J. M. ;
De Brabander, J. ;
Nuijts, R. M. M. A. .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2010, 94 (01) :106-110
[7]   Comparison of central corneal thickness and anterior chamber depth measurements using three imaging technologies in normal eyes and after phakic intraocular lens implantation [J].
Doors, Muriel ;
Cruysberg, Lars P. J. ;
Berendschot, Tos T. J. M. ;
de Brabander, John ;
Verbakel, Frenne ;
Webers, Carroll A. B. ;
Nuijts, Rudy M. M. A. .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2009, 247 (08) :1139-1146
[8]  
Drexler W, 1997, INVEST OPHTH VIS SCI, V38, P1304
[9]   Partial coherence interferometry: A novel approach to biometry in cataract surgery [J].
Drexler, W ;
Findl, O ;
Menapace, R ;
Rainer, G ;
Vass, C ;
Hitzenberger, CK ;
Fercher, AF .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1998, 126 (04) :524-534
[10]   Reproducibility of RTVue Retinal Nerve Fiber Layer Thickness and Optic Disc Measurements and Agreement with Stratus Optical Coherence Tomography Measurements [J].
Gonzalez-Garcia, Alberto O. ;
Vizzeri, Gianmarco ;
Bowd, Christopher ;
Medeiros, Felipe A. ;
Zangwill, Linda M. ;
Weinreb, Robert N. .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2009, 147 (06) :1067-1074