Failure of total hip arthroplasty with a precoated prosthesis - 4- to 11-year results

被引:0
|
作者
Dowd, JE
Cha, CW
Trakru, S
Kim, SY
Yang, CH
Rubash, HE
机构
[1] Massachusetts Gen Hosp, Dept Orthopaed Surg, Boston, MA 02114 USA
[2] Univ Pittsburgh, Med Ctr, Dept Orthopaed Surg, Pittsburgh, PA USA
[3] Univ Pittsburgh, Med Ctr, Div Adult Reconstruct Surg, Pittsburgh, PA USA
关键词
D O I
暂无
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
A series of early femoral component failures prompted a detailed retrospective clinical and radiographic review of 176 hybrid cemented total hip arthroplasties using a polymethylmethacrylate coated femoral prosthesis. All surgeries were performed using third generation cement techniques. Average length of followup was 6.3 years (range, 3-12 years). Twenty-one patients died, and one underwent revision surgery because of sepsis. Of the remaining 154 total hip arthroplasties, 23 (15%) of the femoral components failed (21 revised, two definitely loose). The average time to revision was 3.9 years. None of the acetabular components failed. Comparison between the failure and nonfailure groups revealed that poor cement mantles (Grades C or D) with distal cement mantle deficiencies were statistically significant predictors of femoral failure. The mast common mechanism of failure was progressive, circumferential cement-bone interface osteolysis with relative preservation of the cement-metal interface. Debonding of the cement column from the prosthesis was a late finding and occurred in only 45% of failed cases. Incorporating the techniques of centralization and centrifugation significantly improved clinical results. Strengthening of the cement-prosthesis interface may magnify the deleterious effects of a poor cement mantle and predisposes the cement-bone interface to failure.
引用
收藏
页码:123 / 136
页数:14
相关论文
共 50 条
  • [1] 6-YEAR TO 11-YEAR RESULTS OF TOTAL HIP-ARTHROPLASTY IN RHEUMATOID-ARTHRITIS
    POSS, R
    MALONEY, JP
    EWALD, FC
    THOMAS, WH
    BATTE, NJ
    HARTNESS, C
    SLEDGE, CB
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1984, (182) : 109 - 116
  • [2] Hybrid total hip, arthroplasty - A 7- to 11-year followup
    Goldberg, VM
    Ninomiya, J
    Kelly, G
    Kraay, M
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1996, (333) : 147 - 154
  • [3] TOTAL HIP-ARTHROPLASTY IN JUVENILE RHEUMATOID-ARTHRITIS - 2-YEAR TO 11-YEAR RESULTS
    LACHIEWICZ, PF
    MCCASKILL, B
    INGLIS, A
    RANAWAT, CS
    ROSENSTEIN, BD
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (04): : 502 - 508
  • [4] 9-to 11-year results of cemented titanium Mueller straight stem in total hip arthroplasty
    Baumann, Bernd
    Hendrich, Christian
    Barthel, Thomas
    Bockholt, Michael
    Walther, Markus
    Eulert, Jochen
    Rader, Christof P.
    ORTHOPEDICS, 2007, 30 (07) : 551 - 557
  • [5] Hybrid total hip arthroplasty with a precoated offset stem - Four to nine-year results
    Cannestra, VP
    Berger, RA
    Quigley, LR
    Jacobs, JJ
    Rosenberg, AG
    Galante, JO
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (09): : 1291 - 1299
  • [6] Fifteen-year Results of Precoated Femoral Stem in Primary Hybrid Total Hip Arthroplasty
    Suh, Dong Hun
    Yun, Ho Hyun
    Chun, Sung Kwang
    Shon, Won Yong
    CLINICS IN ORTHOPEDIC SURGERY, 2013, 5 (02) : 110 - 117
  • [7] Early failure of precoated femoral components in primary total hip arthroplasty
    Ong, A
    Wong, KL
    Lai, M
    Garino, JP
    Steinberg, ME
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A (05): : 786 - 792
  • [8] Precoated femoral component in total hip arthroplasty - Results of 5-to 9-year followup
    Brown, EC
    Lachiewicz, PF
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1999, (364) : 153 - 159
  • [9] Total hip arthroplasty with the porous-coated anatomic hip prosthesis - Results at 11 to 18 years
    Little, BS
    Wixson, RL
    Stulberg, SD
    JOURNAL OF ARTHROPLASTY, 2006, 21 (03): : 338 - 343
  • [10] Early failure of a roughened surface, precoated femoral component in total hip arthroplasty
    Sylvain, GM
    Kassab, S
    Coutts, R
    Santore, R
    JOURNAL OF ARTHROPLASTY, 2001, 16 (02): : 141 - 148