Factors affecting short-term bone density precision assessment and the effect on patient monitoring

被引:47
作者
Leslie, William D. [1 ]
机构
[1] Univ Manitoba, Dept Med, Winnipeg, MB, Canada
关键词
bone densitometry; DXA; osteoporosis; precision;
D O I
10.1359/JBMR.071019
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The most widely used procedure for performing a BMD reproducibility assessment (same-technologist with simple repositioning on the same day) systematically underestimates precision error and will lead to over categorization of change in a large fraction of monitored patients. Introduction: The most common procedure for establishing the least significant change (LSC) to monitor bone mineral density (BMD) with DXA is for the same technologist to perform repeat subject scans on the same day with simple repositioning. The objective of the current report is to determine how the reproducibility scanning procedure impacts on the precision assessment and categorization of change in routine clinical practice. Materials and Methods: The study population was drawn from the database of the Manitoba Bone Density Program which includes all clinical DXA test results for the Province of Manitoba, Canada. All patients who had baseline and follow up total spine (L1-4) and the total hip BMD measurements on the same instrument up to March 31, 2007 were included as the 'clinical monitoring population' (N = 5048 scan-pairs). BMD precision was assessed in a convenience sample of patients who were agreeable to undergoing a repeat assessment (50% performed on the same day with repositioning, 68% by different technologists) (N = 331 spine and 328 hip scan-pairs). Results: Precision error was greater when the scan-pairs were acquired on different days than on the same day for both the total spine (p < .001) and total hip (p < .01). No other factor was consistently associated with precision error. The reference LSC (different days and different technologists) categorized the smallest fraction of the monitored population with change, whereas other combinations gave a significant rate of over categorization (up to 19.3% for the lumbar spine and up to 18.3% for the total hip). Conclusions: The most widely procedure for performing a BMD reproducibility assessment (same-technologist with simple repositioning on the same day) systematically underestimates precision error and will lead to over categorization of change in a large fraction of monitored patients.
引用
收藏
页码:199 / 204
页数:6
相关论文
共 50 条
  • [1] The impact of bone area on short-term bone density precision
    Leslie, William D.
    JOURNAL OF CLINICAL DENSITOMETRY, 2006, 9 (02) : 150 - 153
  • [2] Factors affecting short-term precision of musculoskeletal measures using peripheral quantitative computed tomography (pQCT)
    Swinford, R. R.
    Warden, S. J.
    OSTEOPOROSIS INTERNATIONAL, 2010, 21 (11) : 1863 - 1870
  • [3] Factors affecting short-term precision of musculoskeletal measures using peripheral quantitative computed tomography (pQCT)
    R. R. Swinford
    S. J. Warden
    Osteoporosis International, 2010, 21 : 1863 - 1870
  • [4] FACTORS INFLUENCING SHORT-TERM PRECISION OF DUAL X-RAY BONE ABSORPTIOMETRY (DXA) OF SPINE AND FEMUR
    ENGELKE, K
    GLUER, CC
    GENANT, HK
    CALCIFIED TISSUE INTERNATIONAL, 1995, 56 (01) : 19 - 25
  • [5] Minimum sample size requirements for bone density precision assessment produce inconsistency in clinical monitoring
    Leslie, W. D.
    Moayyeri, A.
    OSTEOPOROSIS INTERNATIONAL, 2006, 17 (11) : 1673 - 1680
  • [6] Minimum sample size requirements for bone density precision assessment produce inconsistency in clinical monitoring
    W. D. Leslie
    A. Moayyeri
    Osteoporosis International, 2006, 17 : 1673 - 1680
  • [7] The Short-Term Effect of Prunes in Improving Bone in Men
    George, Kelli S.
    Munoz, Joseph
    Ormsbee, Lauren T.
    Akhavan, Neda S.
    Foley, Elizabeth M.
    Siebert, Shalom C.
    Kim, Jeong-Su
    Hickner, Robert C.
    Arjmandi, Bahram H.
    NUTRIENTS, 2022, 14 (02)
  • [8] Sample size requirements for bone density precision assessments and effect on patient categorization: A Monte Carlo simulation study
    Moayyeri, Alireza
    Sadatsafavi, Molisen
    Leslie, William D.
    BONE, 2007, 41 (04) : 679 - 684
  • [9] The Factors Affecting Bone Density in Cirrhosis
    Hajiabbasi, Asghar
    Shafaghi, Afshin
    Fayazi, Haniyeh Sadat
    Masooleh, Irandokht Shenavar
    Emami, Mohammad Hassan Hedayati
    Parsa, Pooneh Ghavidel
    Maafi, Alireza Amir
    HEPATITIS MONTHLY, 2015, 15 (04)
  • [10] Short-term precision for morphometric X-ray absorptiometry
    Ferrar, L
    Jiang, G
    Eastell, R
    OSTEOPOROSIS INTERNATIONAL, 2001, 12 (09) : 710 - 715