Corneal topography with an aberrometry-topography system

被引:6
作者
Muelhaupt, Michael [1 ]
Dietzko, Sven [1 ]
Wolffsohn, James [2 ]
Bandlitz, Stefan [1 ,2 ]
机构
[1] Cologne Sch Optometry, Hohere Fachschule Augenopt Koln, Bayenthalgurtel 6-8, D-50968 Cologne, Germany
[2] Aston Univ, Ophthalm Res Grp, Life & Hlth Sci, Birmingham, W Midlands, England
关键词
Corneal topography; Placido-based; Corneal radius; Corneal eccentricity; Aberrometry-topography; REPEATABILITY; KERATOCONUS; AGREEMENT; REPRODUCIBILITY; PRECISION;
D O I
10.1016/j.clae.2018.05.001
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To investigate the agreement between the central corneal radii and corneal eccentricity measurements generated by the new Wave Analyzer 700 Medica (WAV) compared to the Keratograph 4 (KER) and to test the repeatability of the instruments. Methods: 20 subjects (10 male, mean age 29.1 years, range 21-50 years) were recruited from the students and staff of the Cologne School of Optometry. Central corneal radii for the flat (r(c/f1)) and steep (r(c/st)) meridian as well as corneal eccentricity for the nasal (e(nas)), temporal (e(temp)), inferior (e(inf)) and superior (e(sup)) directions were measured using WAV and KER by one examiner in a randomized order. Results: Central radii of the flat (r(c/f1)) and steep (r(c/st)) meridian measured with both instruments were statically significantly correlated (r = 0.945 and r = 0.951; p < 0.001). Comparison between the WAV and KER showed that r(c/f1) and r(c/st) measured with WAV were significantly steeper than those measured with KER (p < 0.001). Corneal eccentricities were statistically significantly correlated in all meridians (p < 0.05). Compared to KER, etemp and e s measured with WAV were greater (p < 0.05), while there were no statistically significant differences for eas and e mf (p = 0.350 and p = 0.083). For the central radii, repeated measurements were not significantly different for the KER or WAV (p > 0.05). Limits of agreement (LoA) indicate a better repeatability for the KER compared to WAV. Conclusions: Corneal topography measurements captured with the WAV were strongly correlated with the KER. However, due to the differences in measured corneal radii and eccentricities, the devices cannot be used interchangeably. For corneal topography the KER demonstrated better repeatability.
引用
收藏
页码:436 / 441
页数:6
相关论文
共 21 条
[1]   Optical models of the human eye [J].
Atchison, David A. ;
Thibos, Larry N. .
CLINICAL AND EXPERIMENTAL OPTOMETRY, 2016, 99 (02) :99-106
[2]   Clinical evaluation of the Oculus Keratograph [J].
Best, N. ;
Drury, L. ;
Wolffsohn, J. S. .
CONTACT LENS & ANTERIOR EYE, 2012, 35 (04) :171-174
[3]   The performance of four different corneal topographers on normal human corneas and its impact on orthokeratology lens fitting [J].
Cho, P ;
Lam, AKC ;
Mountford, J ;
Ng, L .
OPTOMETRY AND VISION SCIENCE, 2002, 79 (03) :175-183
[4]   Interchangeability between Placido disc and Scheimpflug system: quantitative and qualitative analysis [J].
De Stefano, Vinicius Silbiger ;
Soares Melo Junior, Luiz Alberto ;
Mallmann, Felipe ;
Schor, Paulo .
ARQUIVOS BRASILEIROS DE OFTALMOLOGIA, 2010, 73 (04) :363-366
[5]   EYESYS CORNEAL TOPOGRAPHY MEASUREMENT APPLIED TO CALIBRATED ELLIPSOIDAL CONVEX SURFACES [J].
DOUTHWAITE, WA .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1995, 79 (09) :797-801
[6]   Cornea measurement comparison with Orbscan II and EyeSys videokeratoscope [J].
Douthwaite, William A. ;
Mallen, Edward A. H. .
OPTOMETRY AND VISION SCIENCE, 2007, 84 (07) :598-604
[7]  
Fan R., 2017, CLIN EXP OPHTHALMOL
[8]  
Fung M. W., 2016, CORNEAL TOPOGRAPHY I
[9]  
Gokul A., 2017, CLIN EXP OPHTHALMOL
[10]   Comparison of reliability and repeatability of corneal curvature assessment with six keratometers [J].
Hamer, Catriona A. ;
Buckhurst, Hetal ;
Purslow, Christine ;
Shum, Gary L. ;
Habib, Nabil E. ;
Buckhurst, Phillip J. .
CLINICAL AND EXPERIMENTAL OPTOMETRY, 2016, 99 (06) :583-589