Major adverse clinical events associated with implantation of a leadless intracardiac pacemaker

被引:48
作者
Hauser, Robert G. [1 ]
Gornick, Charles C. [1 ]
Abdelhadi, Raed H. [1 ]
Tang, Chuen Y. [1 ]
Casey, Susan A. [1 ]
Sengupta, Jay D. [1 ]
机构
[1] Minneapolis Heart Inst Fdn, Minneapolis, MN USA
关键词
Cardiac pacing; Complications; Leadless; Lead; Perforation; Tamponade; Transcatheter;
D O I
10.1016/j.hrthm.2021.03.015
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND Leadless intracardiac pacemakers were developed to avoid the complications of transvenous pacing systems. The Medtronic Micra (TM) transcatheter pacemaker is one such system. We found an unexpected number of major adverse clinical events (MACE) in the Food and Drug Administration's Manufacturers and User Facility Device Experience (MAUDE) database associated with Micra implantation. OBJECTIVE The purpose of this study was to describe these MACE and compare them to implant procedure MACE in MAUDE for Medtronic CapSureFix (TM) active-fixation transvenous pacing leads. METHODS During January 2021, we queried the MAUDE database for reports of MACE for Micra pacemakers and CapSureFix leads using the simple search terms " death," "tamponade," and "perforation." Reports from 2016-2020 were included. RESULTS The search identified 363 MACE for Micra and 960 MACE for CapSureFix leads, including 96 Micra deaths (26.4%) vs 23 CapSureFix deaths (2.4%) (P<.001); 287 Micra tamponades (79.1%) vs 225 tamponades for CapSureFix (23.4%) (P<.001); and 99 rescue thoracotomies for Micra (27.3%) vs 50 rescue thoracotomies for CapSureFix (5.2%) (P<.001). More Micra patients required cardiopulmonary resuscitation (21.8% vs 1.1%) and suffered hypotension or shock (22.0% vs 5.8%) than CapSureFix recipients (P<.001). Micra patients were more likely to survive a myocardial perforation or tear if they had surgical repair (P=.014). CONCLUSION Micra leadless pacemaker implantation may be complicated by myocardial and vascular perforations and tears that result in cardiac tamponade and death. We estimate the incidence is low (<1%). Rescue surgery to repair perforations may be lifesaving. MACE are significantly less for implantation of CapSureFix transvenous ventricular pacing leads.
引用
收藏
页码:1132 / 1139
页数:8
相关论文
共 16 条
  • [1] Updated performance of the Micra transcatheter pacemaker in the real-world setting: A comparison to the investigational study and a transvenous historical control
    El-Chami, Mikhael F.
    Al-Samadi, Faisal
    Clementy, Nicolas
    Garweg, Christophe
    Luis Martinez-Sande, Jose
    Piccini, Jonathan P.
    Iacopino, Saverio
    Lloyd, Michael
    Vinolas Prat, Xavier
    Jacobsen, Michael Dilou
    Ritter, Philippe
    Johansen, Jens Brock
    Tondo, Claudio
    Liu, Fang
    Fagan, Dedra H.
    Eakley, Alyssa K.
    Roberts, Paul R.
    [J]. HEART RHYTHM, 2018, 15 (12) : 1800 - 1807
  • [2] Garg A., 2020, HEART RHYTHM, V17, P2056, DOI [10.1016/j.hrthm.2020.07.035, DOI 10.1016/j.hrthm.2020.07.035]
  • [3] Deaths caused by the failure of Riata and Riata ST implantable cardioverter-defibrillator leads
    Hauser, Robert G.
    Abdelhadi, Raed
    McGriff, Deepa
    Retel, Linda Kallinen
    [J]. HEART RHYTHM, 2012, 9 (08) : 1227 - 1235
  • [4] Deaths and cardiovascular injuries due to device-assisted implantable cardioverter-defibrillator and pacemaker lead extraction
    Hauser, Robert G.
    Katsiyiannis, William T.
    Gornick, Charles C.
    Almquist, Adrian K.
    Kallinen, Linda M.
    [J]. EUROPACE, 2010, 12 (03): : 395 - 401
  • [5] Leadless Cardiac Pacemakers Back to the Future
    Miller, Marc A.
    Neuzil, Petr
    Dukkipati, Srinivas R.
    Reddy, Vivek Y.
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2015, 66 (10) : 1179 - 1189
  • [6] Troubleshooting in PM Leadless: How to manage an indissoluble knot
    Morani, Giovanni
    Bolzan, Bruna
    Tomasi, Luca
    Zimelli, Emma
    Marini, Paola
    Ribichini, Flavio Luciano
    [J]. JOURNAL OF ARRHYTHMIA, 2019, 35 (04) : 676 - 678
  • [7] Percutaneous Implantation of an Entirely Intracardiac Leadless Pacemaker
    Reddy, Vivek Y.
    Exner, Derek V.
    Cantillon, Daniel J.
    Doshi, Rahul
    Bunch, T. Jared
    Tomassoni, Gery F.
    Friedman, Paul A.
    Estes, N. A. Mark, III
    Ip, John
    Niazi, Imran
    Plunkitt, Kenneth
    Banker, Rajesh
    Porterfield, James
    Ip, James E.
    Dukkipati, Srinivas R.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2015, 373 (12) : 1125 - 1135
  • [8] Permanent Leadless Cardiac Pacing Results of the LEADLESS Trial
    Reddy, Vivek Y.
    Knops, Reinoud E.
    Sperzel, Johannes
    Miller, Marc A.
    Petru, Jan
    Simon, Jaroslav
    Sediva, Lucie
    de Groot, Joris R.
    Tjong, Fleur V. Y.
    Jacobson, Peter
    Ostrosff, Alan
    Dukkipati, Srinivas R.
    Koruth, Jacob S.
    Wilde, Arthur A. M.
    Kautzner, Josef
    Neuzil, Petr
    [J]. CIRCULATION, 2014, 129 (14) : 1466 - 1471
  • [9] A Leadless Intracardiac Transcatheter Pacing System
    Reynolds, Dwight
    Duray, Gabor Z.
    Omar, Razali
    Soejima, Kyoko
    Neuzil, Petr
    Zhang, Shu
    Narasimhan, Calambur
    Steinwender, Clemens
    Brugada, Josep
    Lloyd, Michael
    Roberts, Paul R.
    Sagi, Venkata
    Hummel, John
    Bongiorni, Maria Grazia
    Knops, Reinoud E.
    Ellis, Christopher R.
    Gornick, Charles C.
    Bernabei, Matthew A.
    Laager, Verla
    Stromberg, Kurt
    Williams, Eric R.
    Hudnall, J. Harrison
    Ritter, Philippe
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2016, 374 (06) : 533 - 541
  • [10] The rationale and design of the Micra Transcatheter Pacing Study: safety and efficacy of a novel miniaturized pacemaker
    Ritter, Philippe
    Duray, Gabor Z.
    Zhang, Shu
    Narasimhan, Calambur
    Soejima, Kyoko
    Omar, Razali
    Laager, Verla
    Stromberg, Kurt
    Williams, Eric
    Reynolds, Dwight
    [J]. EUROPACE, 2015, 17 (05): : 807 - 813