Recent developments in capillary isoelectric focusing with whole-column imaging detection

被引:26
作者
Fang, XH
Tragas, C
Wu, JQ
Mao, QL
Pawliszyn, J [1 ]
机构
[1] Univ Waterloo, Dept Chem, Waterloo, ON N2L 3G1, Canada
[2] Convergent Biosci Ltd, Etobicoke, ON, Canada
关键词
capillary isoelectric focusing; imaging detection; on-line desalting; tapered capillary; protein analysis;
D O I
10.1002/elps.1150191307
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Capillary isoelectric focusing (CIEF) is a high resolution technique for protein separation. The on-column single point detector requires a mobilization step which lengthens the analysis time and causes an uneven resolution along the separation column. The real time and whole column imaging detection has been developed for performing CIEF without mobilization. Three types of imaging detection systems have been developed: optical absorption, refractive index gradient, and laser induced fluorescence. This technique provides a fast analysis speed (about 6 min) and a good resolution of 0.03 pH unit level. Using the absorption imaging detector, ampholyte-free IEF in tapered capillary is being demonstrated, which eliminates the interference of the expensive carrier ampholytes for protein detection in UV region. Recent advancements in this imaged CIEF technique as well as its applications are reviewed.
引用
收藏
页码:2290 / 2295
页数:6