A New Paradigm of Dielectric Relaxation Spectroscopy for Non-invasive Detection of Breast Abnormalities A - preliminary feasibility analysis

被引:5
作者
Dhurjaty, Sreeram [1 ]
Qiu, Yuchen [2 ]
Tan, Maxine [2 ]
Qian, Wei [3 ]
Zheng, Bin [2 ]
机构
[1] Dhurjaty Elect Consulting LLC, Rochester, NY 14618 USA
[2] Univ Oklahoma, Sch Elect & Comp Engn, Norman, OK 73019 USA
[3] Univ Texas El Paso, Dept Elect & Comp Engn, El Paso, TX 79968 USA
来源
MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING | 2016年 / 9783卷
关键词
A new breast cancer prescreening method; Detection of breast abnormality; Dielectric Relaxation Spectroscopy (DRS); Electrical Impedance Spectroscopy (EIS); CANCER RISK; MAMMOGRAPHY; WOMEN;
D O I
10.1117/12.2217196
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
In order to improve efficacy of screening mammography, in recent years, we have been investigating the feasibility of applying a resonance-frequency based electrical impedance spectroscopy (REIS) technology to non-invasively detect breast abnormalities that may lead to the development of cancer in the near-term. Despite promising study-results, we found that REIS suffered from relatively poor reproducibility due to perturbations in electrode placement, contact pressure variation on the breast, as well as variation of the resonating inductor. To overcome this limitation, in this study, we propose and analyze a new paradigm of Dielectric Relaxation Spectroscopy (DRS) that measures polarization-lag of dielectric signals in breast-capacitance when excited by the pulses or sine waves. Unlike conventional DRS that operates using the signals at very high frequencies (GHz) to examine changes in polarization, our new method detects and characterizes the dielectric properties of tissue at low frequencies (<= 10 MHz) due to the advent of inexpensive oscillators that are accurate to 1 pico-second (used in GPS receivers) as well as measurement of amplitudes of 1 ppm or better. From theoretical analysis, we have proved that the sensitivity of new DRS in detecting permittivity of water increased by >= 80 times as compared to conventional DRS, which operates at frequencies around 4GHz. By analyzing and comparing the relationship between the new DRS and REIS, we found that this DRS has potential advantages in enhancing repeatability from various readings, including temperature-insensitive detection, and yielding higher resolution or sensitivity (up to 100 Femtofarads).
引用
收藏
页数:7
相关论文
共 15 条
[1]   Detection of Breast Cancer With Addition of Annual Screening Ultrasound or a Single Screening MRI to Mammography in Women With Elevated Breast Cancer Risk [J].
Berg, Wendie A. ;
Zhang, Zheng ;
Lehrer, Daniel ;
Jong, Roberta A. ;
Pisano, Etta D. ;
Barr, Richard G. ;
Boehm-Velez, Marcela ;
Mahoney, Mary C. ;
Evans, W. Phil, III ;
Larsen, Linda H. ;
Morton, Marilyn J. ;
Mendelson, Ellen B. ;
Farria, Dione M. ;
Cormack, Jean B. ;
Marques, Helga S. ;
Adams, Amanda ;
Yeh, Nolin M. ;
Gabrielli, Glenna .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2012, 307 (13) :1394-1404
[2]   More Mammography Muddle: Emotions, Politics, Science, Costs, and Polarization [J].
Berlin, Leonard ;
Hall, Ferris M. .
RADIOLOGY, 2010, 255 (02) :311-316
[3]   Risk-Based Mammography Screening: An Effort to Maximize the Benefits and Minimize the Harms [J].
Brawley, Otis W. .
ANNALS OF INTERNAL MEDICINE, 2012, 156 (09) :662-663
[4]  
Dhurjaty S., 2014, Patent No. [US2015/212027 A1, 2015212027]
[5]   Resonance-frequency based Electrical Impedance Spectroscopy and Its Detection Sensitivity to Breast Lesions [J].
Dhurjaty, Sreeram ;
Zheng, Bin ;
Gur, David .
MEDICAL IMAGING 2013: PHYSICS OF MEDICAL IMAGING, 2013, 8668
[6]   A Support Vector Machine Designed to Identify Breasts at High Risk Using Multi-probe Generated REIS Signals: A Preliminary Assessment [J].
Gur, David ;
Zheng, Bin ;
Lederman, Dror ;
Dhurjaty, Sreeram ;
Sumkin, Jules ;
Zuley, Margarita .
MEDICAL IMAGING 2010: IMAGE PERCEPTION, OBSERVER PERFORMANCE, AND TECHNOLOGY ASSESSMENT, 2010, 7627
[7]   Technology review: The use of electrical impedance scanning in the detection of breast cancer [J].
Hope, TA ;
Iles, SE .
BREAST CANCER RESEARCH, 2004, 6 (02) :69-74
[8]   Cumulative Probability of False-Positive Recall or Biopsy Recommendation After 10 Years of Screening Mammography A Cohort Study [J].
Hubbard, Rebecca A. ;
Kerlikowske, Karla ;
Flowers, Chris I. ;
Yankaskas, Bonnie C. ;
Zhu, Weiwei ;
Miglioretti, Diana L. .
ANNALS OF INTERNAL MEDICINE, 2011, 155 (08) :481-U46
[9]   Variability of impedivity in normal and pathological breast tissue [J].
Jossinet, J .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1996, 34 (05) :346-350
[10]   A GMM-based breast cancer risk stratification using a resonance-frequency electrical impedance spectroscopy [J].
Lederman, Dror ;
Zheng, Bin ;
Wang, Xingwei ;
Sumkin, Jules H. ;
Gur, David .
MEDICAL PHYSICS, 2011, 38 (03) :1649-1659