Prospective clinical and joint simulator studies of a new total hip arthroplasty using alumina ceramic heads and cross-linked polyethylene cups

被引:188
作者
Wroblewski, BM [1 ]
Siney, PD [1 ]
Dowson, D [1 ]
Collins, SN [1 ]
机构
[1] UNIV LEEDS,DEPT MECH ENGN,LEEDS LS2 9JT,W YORKSHIRE,ENGLAND
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1996年 / 78B卷 / 02期
关键词
D O I
10.1302/0301-620X.78B2.0780280
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We report the findings from independent prospective clinical and laboratory-based joint-simulator studies of the performance of ceramic femoral heads of 22.225 mm diameter in cross-linked polyethylene (XLP) acetabular cups. We found remarkable qualitative and quantitative agreement between the clinical and simulator results for the wear characteristics with time, and confirmed that ceramic femoral heads penetrate the XLP cups at only about half the rate of otherwise comparable metal heads. In the clinical study, 19 hips in 17 patients were followed for an average of 77 months, In the hip-joint simulator a similar prosthesis was tested for 7.3 million cycles. Both clinical and simulator results showed relatively high rates of penetration over the first 18 months or 1.5 million cycles, followed by a very much lower wear thereafter, Once an initial bedding-in of 0.2 mm to 0.4 mm had taken place the subsequent rates of penetration were very small, The initial clinical wear during bedding-in averaged 0.29 mm/year; subsequent progression was an order of magnitude lower at about 0.022 mm/year, lower than the 0.07 mm/year in metal-to-UHMWP Charnley LFAs. Our results show the excellent tribological features of alumina-ceramic-to-XLP implants, and also confirm the value of well-designed joint simulators for the evaluation of total joint replacements.
引用
收藏
页码:280 / 285
页数:6
相关论文
共 26 条
  • [1] Charnley J., 1979, Low friction arthroplasty of the hip: theory and practice, P20
  • [2] COLLINS SN, 1960, THESIS U LEEDS
  • [3] THE ROLE OF COUNTERFACE IMPERFECTIONS IN THE WEAR OF POLYETHYLENE
    DOWSON, D
    TAHERI, S
    WALLBRIDGE, NC
    [J]. WEAR, 1987, 119 (03) : 277 - 293
  • [4] Dowson D, 1988, Eng Med, V17, P111, DOI 10.1243/EMED_JOUR_1988_017_031_02
  • [5] DOWSON D, 1994, WEAR, V170, P171
  • [6] DOWSON D, 1993, WEAR, V162, P88
  • [7] EGLI A, 1990, ULTRAHIGH MOL WEIGHT, P154
  • [8] GRIFFITH MJ, 1978, CLIN ORTHOP RELAT R, V137, P37
  • [9] ISAAC GH, 1992, CLIN ORTHOP RELAT R, V276, P115
  • [10] Ohashi T, 1989, BIOCERAMICS, V1, P278