Determination of ATP via the photochemical generation of hydrogen peroxide using flow injection luminol chemiluminescence detection

被引:97
作者
Pérez-Ruiz, T [1 ]
Martínez-Lozano, C [1 ]
Tomás, V [1 ]
Martín, J [1 ]
机构
[1] Univ Murcia, Fac Chem, Dept Analyt Chem, E-30071 Murcia, Spain
关键词
ATP; 9,10-anthraquinone-2,6-disulfonate; photochemical reaction; flow injection; chemiluminescence detection;
D O I
10.1007/s00216-003-2071-2
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The determination of ATP using the coupling between a photochemical reaction and a chemiluminescence reaction in a flow injection (FI) system is described. The method is based on the reaction of glucose with ATP catalyzed by hexokinase and Mg2+ ions. The glucose that is not consumed by ATP is subsequently photooxidized using 9,10-anthraquinone-2,6-disulfonate as a sensitizer. The hydrogen peroxide produced in the photochemical reaction is monitored through the chemiluminescence reaction with luminol catalyzed by hematine. There is a linear relationship between the decrease in the chemiluminescence response and the ATP concentration at a constant concentration of glucose. Under the optimum conditions, the calibration graph is linear in the range 0.20-50.5 mg L-1 with a throughput of 25 samples per hour and relative standard deviations between +/-0.62 and +/-1.42%. The limit of detection is 0.07 mg L-1. The method was used for the determination of ATP in pharmaceuticals, milk, and soils.
引用
收藏
页码:189 / 194
页数:6
相关论文
共 50 条
  • [21] DETERMINATION OF PROMETHAZINE BY ITS INHIBITION OF THE CHEMILUMINESCENCE OF THE LUMINOL HYDROGEN-PEROXIDE CHROMIUM(III) SYSTEM
    ALWARTHAN, AA
    ALTAMRAH, SA
    AKEL, AA
    ANALYTICA CHIMICA ACTA, 1993, 282 (01) : 169 - 174
  • [22] Flow-injection chemiluminescence determination of DNA based on a photochemical reaction
    Pérez-Ruiz, T
    Martínez-Lozano, C
    Tomás, V
    Martín, J
    ANALYTICA CHIMICA ACTA, 1999, 402 (1-2) : 13 - 19
  • [23] Investigation of iron(III) reduction and trace metal interferences in the determination of dissolved iron in seawater using flow injection with luminol chemiluminescence detection
    Ussher, Simon J.
    Milne, Angela
    Landing, William M.
    Attiq-ur-Rehman, Kakar
    Seguret, Marie J. M.
    Holland, Toby
    Achterberg, Eric P.
    Nabi, Abdul
    Worsfold, Paul J.
    ANALYTICA CHIMICA ACTA, 2009, 652 (1-2) : 259 - 265
  • [24] FLOW-INJECTION DETERMINATION OF BENZIMIDAZOLE FUNGICIDES IN NATURAL WATERS WITH COPPER(II)-HYDROGEN PEROXIDE CHEMILUMINESCENCE
    Waseem, Amir
    Yaqoob, Mohammad
    Nabi, Abdul
    ANALYTICAL LETTERS, 2010, 43 (04) : 603 - 617
  • [25] A highly sensitive and selective assay for cysteine using the chemiluminescence reaction of luminol and hydrogen peroxide
    Li, YX
    Zhu, CQ
    Wang, L
    MICROCHIMICA ACTA, 2005, 150 (01) : 95 - 99
  • [26] Chemiluminescence detection for Co (II) based on luminol-hydrogen peroxide reaction on a microfluidic chip
    Xueye Chen
    Fuqiang Chang
    Xing Wei
    Geng Li
    Journal of Food Science and Technology, 2015, 52 : 601 - 604
  • [27] A Novel Flow Injection Chemiluminescence Method with Ferricyanide and Luminol for the Determination of Ractopamine in Biological Samples
    Feng Ting
    Hu Yu-fei
    Li Gong-ke
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2012, 32 (01) : 74 - 77
  • [28] Chemiluminescence detection for Co (II) based on luminol-hydrogen peroxide reaction on a microfluidic chip
    Chen, Xueye
    Chang, Fuqiang
    Wei, Xing
    Li, Geng
    JOURNAL OF FOOD SCIENCE AND TECHNOLOGY-MYSORE, 2015, 52 (01): : 601 - 604
  • [29] Flow injection chemiluminescence method for the determination of phenol with N-bromosuccinimide-luminol system
    Du, LY
    Ji, NN
    Wang, SH
    Zhuang, HS
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2005, 25 (12) : 1943 - 1946
  • [30] Determination of norfloxacin by flow injection analysis with chemiluminescence detection
    Sun, Han-wen
    Li, Li-qing
    Chen, Xue-yan
    Shi, Hong-mei
    Lue, Yun-kai
    CANADIAN JOURNAL OF ANALYTICAL SCIENCES AND SPECTROSCOPY, 2006, 51 (02): : 100 - 107