Determination of ATP via the photochemical generation of hydrogen peroxide using flow injection luminol chemiluminescence detection

被引:97
|
作者
Pérez-Ruiz, T [1 ]
Martínez-Lozano, C [1 ]
Tomás, V [1 ]
Martín, J [1 ]
机构
[1] Univ Murcia, Fac Chem, Dept Analyt Chem, E-30071 Murcia, Spain
关键词
ATP; 9,10-anthraquinone-2,6-disulfonate; photochemical reaction; flow injection; chemiluminescence detection;
D O I
10.1007/s00216-003-2071-2
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The determination of ATP using the coupling between a photochemical reaction and a chemiluminescence reaction in a flow injection (FI) system is described. The method is based on the reaction of glucose with ATP catalyzed by hexokinase and Mg2+ ions. The glucose that is not consumed by ATP is subsequently photooxidized using 9,10-anthraquinone-2,6-disulfonate as a sensitizer. The hydrogen peroxide produced in the photochemical reaction is monitored through the chemiluminescence reaction with luminol catalyzed by hematine. There is a linear relationship between the decrease in the chemiluminescence response and the ATP concentration at a constant concentration of glucose. Under the optimum conditions, the calibration graph is linear in the range 0.20-50.5 mg L-1 with a throughput of 25 samples per hour and relative standard deviations between +/-0.62 and +/-1.42%. The limit of detection is 0.07 mg L-1. The method was used for the determination of ATP in pharmaceuticals, milk, and soils.
引用
收藏
页码:189 / 194
页数:6