The Impact of Cataract, and Its Surgical Removal, on Measures of Macular Pigment Using the Heidelberg Spectralis HRA plus OCT MultiColor Device

被引:23
作者
Akuffo, Kwadwo Owusu [1 ]
Nolan, John M. [1 ]
Stack, Jim [1 ]
Power, Rebecca [1 ]
Kirwan, Clare [1 ,2 ,3 ]
Moran, Rachel [1 ]
Corcoran, Laura [1 ]
Owens, Niamh [1 ]
Beatty, Stephen [1 ,2 ,3 ]
机构
[1] Waterford Inst Technol, Sch Hlth Sci, Nutr Res Ctr Ireland, Macular Pigment Res Grp, Waterford, Ireland
[2] Whitfield Clin, Inst Eye Surg, Waterford, Ireland
[3] Whitfield Clin, Inst Vis Res, Waterford, Ireland
基金
欧洲研究理事会;
关键词
macular pigment; macular pigment optical density; fundus autofluorescence; concordance correlation coefficient; agreement; spectralis; cataract; cataract surgery; phacoemulsification; Tecnis IOL; lutein; zeaxanthin; meso-zeaxanthin; visual acuity; optical coherence tomography; HETEROCHROMATIC FLICKER PHOTOMETRY; VISUAL FUNCTION; ALZHEIMERS-DISEASE; IN-VIVO; ZEAXANTHIN; LUTEIN; SUPPLEMENTATION; AUTOFLUORESCENCE; CONCORDANCE; THICKNESS;
D O I
10.1167/iovs.16-19141
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. To investigate the effect of cataract (and cataract surgery) on macular pigment (MP) measurements using the Heidelberg Spectralis HRA+OCT MultiColor device. METHODS. Thirty-six patients (age, 54-87 years) scheduled for cataract surgery at the Institute of Eye Surgery, Ireland, were enrolled in this study. Cataracts were graded using the Lens Opacities Classification System (LOCS) III, and surgery was performed using standard phacoemulsification technique with implantation of a Tecnis ZCB00 or Tecnis ZCT intraocular lens. Macular pigment was measured before and after cataract surgery in the operated (study) eye and in the fellow (control) eye. RESULTS. In the study eye, there was statistically significant disagreement in measures of MP taken before and after surgery. At all eccentricities, and also for MP volume, the postsurgery measurements were significantly (P < 0.05) greater, ranging from an average 16% greater at 1.728 to an average 35% greater at 0.23 degrees eccentricity. Eyes exhibiting large disagreement between pre- and postsurgery measurements at a given eccentricity also generally exhibited substantial disagreement at other eccentricities. Overall severity of cataract contributed to greater disagreement between pre- and postoperative measures of MP, as did grade of nuclear opalescence, nuclear color, and posterior subcapsular cataract. In control eyes, there was no statistically significant disagreement in terms of measures of MP taken before and after cataract surgery (P > 0.05 for all; 1-sample t-test). CONCLUSIONS. Macular pigment measurements using the Spectralis are affected by cataract. Accordingly, we recommend that cataract be graded when measuring MP with a device that utilizes dual-wavelength fundus autofluorescence and propose the employment of a correction factor to compensate for cataract when measuring MP.
引用
收藏
页码:2552 / 2563
页数:12
相关论文
共 37 条
  • [1] Sustained supplementation and monitored response with differing carotenoid formulations in early age-related macular degeneration
    Akuffo, K. O.
    Nolan, J. M.
    Howard, A. N.
    Moran, R.
    Stack, J.
    Klein, R.
    Klein, B. E.
    Meuer, S. M.
    Sabour-Pickett, S.
    Thurnham, D. I.
    Beatty, S.
    [J]. EYE, 2015, 29 (07) : 902 - 912
  • [2] Concordance of Macular Pigment Measurement Using Customized Heterochromatic Flicker Photometry and Fundus Autofluorescence in Age-Related Macular Degeneration
    Akuffo, Kwadwo Owusu
    Beatty, Stephen
    Stack, Jim
    Peto, Tunde
    Leung, Irene
    Corcoran, Laura
    Power, Rebecca
    Nolan, John M.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2015, 56 (13) : 8207 - 8214
  • [3] Central Retinal Enrichment Supplementation Trials (CREST): Design and Methodology of the CREST Randomized Controlled Trials
    Akuffo, Kwadwo Owusu
    Beatty, Stephen
    Stack, Jim
    Dennison, Jessica
    O'Regan, Sarah
    Meagher, Katherine A.
    Peto, Tunde
    Nolan, John
    [J]. OPHTHALMIC EPIDEMIOLOGY, 2014, 21 (02) : 111 - 123
  • [4] OPTICAL-DENSITY SPECTRA OF THE MACULAR PIGMENT INVIVO AND INVITRO
    BONE, RA
    LANDRUM, JT
    CAINS, A
    [J]. VISION RESEARCH, 1992, 32 (01) : 105 - 110
  • [5] Heterochromatic flicker photometry
    Bone, RA
    Landrum, JT
    [J]. ARCHIVES OF BIOCHEMISTRY AND BIOPHYSICS, 2004, 430 (02) : 137 - 142
  • [6] Prognostic Indicators and Outcome Measures for Surgical Removal of Symptomatic Nonadvanced Cataract
    Charalampidou, Sofia
    Loughman, James
    Nolan, John
    Stack, Jim
    Cassidy, Lorraine
    Pesudovs, Konrad
    Beatty, Stephen
    [J]. ARCHIVES OF OPHTHALMOLOGY, 2011, 129 (09) : 1155 - 1161
  • [7] Lutein plus Zeaxanthin and Omega-3 Fatty Acids for Age-Related Macular Degeneration The Age-Related Eye Disease Study 2 (AREDS2) Randomized Clinical Trial
    Chew, Emily Y.
    Clemons, Traci E.
    SanGiovanni, John Paul
    Danis, Ronald
    Ferris, Frederick L., III
    Elman, Michael
    Antoszyk, Andrew
    Ruby, Alan
    Orth, David
    Bressler, Susan
    Fish, Gary
    Hubbard, Baker
    Klein, Michael
    Chandra, Suresh
    Blodi, Barbara
    Domalpally, Amitha
    Friberg, Thomas
    Wong, Wai
    Rosenfeld, Philip
    Agron, Elvira
    Toth, Cynthia
    Bernstein, Paul
    Sperduto, Robert
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2013, 309 (19): : 2005 - 2015
  • [8] THE LENS OPACITIES CLASSIFICATION SYSTEM-III
    CHYLACK, LT
    WOLFE, JK
    SINGER, DM
    LESKE, MC
    BULLIMORE, MA
    BAILEY, IL
    FRIEND, J
    MCCARTHY, D
    WU, SY
    [J]. ARCHIVES OF OPHTHALMOLOGY, 1993, 111 (06) : 831 - 836
  • [9] Autofluorescence method to measure macular pigment optical densities fluorometry and autofluorescence imaging
    Delori, FC
    [J]. ARCHIVES OF BIOCHEMISTRY AND BIOPHYSICS, 2004, 430 (02) : 156 - 162
  • [10] Concordance of macular pigment measurements obtained using customized heterochromatic flicker photometry, dual-wavelength autofluorescence, and single-wavelength reflectance
    Dennison, Jessica L.
    Stack, Jim
    Beatty, Stephen
    Nolan, John M.
    [J]. EXPERIMENTAL EYE RESEARCH, 2013, 116 : 190 - 198