The measurement of Pb isotope ratios in sub-ng quantities by fast scanning single collector sector field-ICP-MS

被引:23
作者
Newman, Karla [1 ]
Georg, R. Bastian [1 ]
机构
[1] Trent Univ, Water Qual Ctr, Peterborough, ON K9J 8E5, Canada
关键词
Lead isotopes; Pb-depleted; Single collector ICP-MS; Matrix effects; Multiple ion counters; MC-ICP-MS; PLASMA-MASS SPECTROMETRY; LEAD ISOTOPES; ENVIRONMENTAL-SAMPLES; DOUBLE SPIKE; PRECISION; ACCURATE; GROUNDWATER; THALLIUM; PB-206; TIMS;
D O I
10.1016/j.chemgeo.2012.02.010
中图分类号
P3 [地球物理学]; P59 [地球化学];
学科分类号
0708 ; 070902 ;
摘要
This work describes an analytical protocol for the measurement of isotope ratios in samples containing small amounts of lead (i.e. <= 0.2 ng total Pb) using the Nu Attom single collector-ICP-MS. The method was evaluated using the NISI' SRM 981 isotopic standard and standard-sample bracketing for correction of the instrumental mass fractionation. The long-term precision (n = 32, +/- 2SD) was 0.11% for the Pb-208/Pb-206 and Pb-207/Pb-206 ratios and 021-0.23% for the Pb-29x/Pb-204 ratios (where x = 208, 207, 206). The precision of the measured Pb isotope ratios reported here (for equivalent total Pb quantities) is comparable to multi-collector (MC) methodologies, using either an array of ion counters (Cocherie and Robert, 2007, Chem. Geol. 243,90-104) or conventional Faraday collectors coupled with a high efficiency sample cone (Makishima and Nakamura, 2010, J. Anal. At. Spectrom. 25, 1712-1716). The accuracy of the measured Pb isotope ratios and the effect of the sample matrix were evaluated for a series of USGS rock standard digests and doped solutions of the NIST SRM 981 standard. For total cation concentrations < 10 mu g mL(-1) and concomitant low concentrations (< 1.5 mu g mL(-1)) of easily ionisable elements (i.e. Na and K), matrix effects were not resolvable at the level of precision reported. However, significant matrix effects were associated with sample compositions typical of natural waters, and removal of (at least) the major cations is required for accurate isotope ratio measurements. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:151 / 157
页数:7
相关论文
共 35 条
[1]   Isotopic analysis of lead in sub-nanogram quantities by TIMS using a 202Pb-205Pb spike [J].
Amelin, Y. ;
Davis, W. J. .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2006, 21 (10) :1053-1061
[2]   Pb isotopic analysis of standards and samples using a 207Pb-204Pb double spike and thallium to correct for mass bias with a double-focusing MC-ICP-MS [J].
Baker, J ;
Peate, D ;
Waight, T ;
Meyzen, C .
CHEMICAL GEOLOGY, 2004, 211 (3-4) :275-303
[3]   Influence of non-spectral matrix effects on the accuracy of Pb isotope ratio measurement by MC-ICP-MS: implications for the external normalization method of instrumental mass bias correction [J].
Barling, Jane ;
Weis, Dominique .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2008, 23 (07) :1017-1025
[4]   OCCURRENCE AND REDUCTION OF NOISE IN INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY FOR ENHANCED PRECISION IN ISOTOPE RATIO MEASUREMENT [J].
BEGLEY, IS ;
SHARP, BL .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (03) :171-176
[5]  
Bouman C., 2009, 30187 THERM FISH SCI
[6]   Direct measurement of lead isotope ratios in low concentration environmental samples by MC-ICP-MS and multi-ion counting [J].
Cocherie, Alain ;
Robert, Michele .
CHEMICAL GEOLOGY, 2007, 243 (1-2) :90-104
[7]   Applications of accurate, high-precision Pb isotope ratio measurement by multi-collector ICP-MS [J].
Collerson, KD ;
Kamber, BS ;
Schoenberg, R .
CHEMICAL GEOLOGY, 2002, 188 (1-2) :65-83
[8]  
de Laeter JR, 1998, MASS SPECTROM REV, V17, P97, DOI 10.1002/(SICI)1098-2787(1998)17:2<97::AID-MAS2>3.0.CO
[9]  
2-J
[10]   Development of a new method for Pb isotopic analysis of archaeological artefacts using single-collector ICP-dynamic reaction cell-MS [J].
De Muynck, David ;
Cloquet, Christophe ;
Vanhaecke, Frank .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2008, 23 (01) :62-71