Dual Inductive Link Coil Design for a Neural Recording System

被引:0
作者
Rush, Alexander [1 ]
Troyk, Philip R. [1 ]
机构
[1] IIT, Chicago, IL 60616 USA
来源
2011 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC) | 2011年
关键词
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper reports an approach to the physical design of the coils used in a dual inductive link to provide two-way wireless communication and power for a neural recording system. The design approach makes use of an analytic model of the link performance in terms of the physical parameters of the link, which allows physical parameters to be iterated on a computer rather than on the bench to find the optimal design within the physical restrictions imposed. In particular, this approach was used to choose the optimal implant data coil sizing to maximize the difference between the contributions of the
引用
收藏
页码:6397 / 6400
页数:4
相关论文
共 11 条
[1]   EFFECTS OF EDDY CURRENTS IN TRANSFORMER WINDINGS [J].
DOWELL, PL .
PROCEEDINGS OF THE INSTITUTION OF ELECTRICAL ENGINEERS-LONDON, 1966, 113 (08) :1387-&
[2]  
JOW UM, 2009, P 31 IEEE EMBS ANN I, P6387
[3]   Self-capacitance of inductors [J].
Massarini, A ;
Kazimierczuk, MK .
IEEE TRANSACTIONS ON POWER ELECTRONICS, 1997, 12 (04) :671-676
[4]  
Nan X, 2003, IEEE POWER ELECTRON, P853
[5]  
REUBEN L, 1947, ELECT TRANSFORMERS C
[6]  
ROSA EB, 1911, SCI PAPERS BUREAU ST
[7]   Inductive Link Design for Miniature Implants [J].
Troyk, Philip R. ;
Rush, Alexander D. .
2009 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-20, 2009, :204-209
[8]  
TROYK PR, 2011, P 33 ANN INT C IEEE
[9]  
Wang G., 2006, P INT C IEEE ENG MED, P4392
[10]   Inductor modeling in wireless links for implantable electronics [J].
Yang, Zhi ;
Liu, Wentai ;
Basham, Eric .