Investigation of the effects of pneumatic tube transport system on routine biochemistry, hematology, and coagulation tests in Ankara City Hospital

被引:2
作者
Yurt, Emine F. [1 ]
Akbiyik, Filiz [2 ]
Bicer, Cemile [3 ,4 ]
机构
[1] Adilcevaz Oncol Hosp, Med Biochem, Bitlis, Turkey
[2] Ankara City Hosp, Siemens Healthineers, Med Biochem, Ankara, Turkey
[3] Ankara Yildirim Beyazit Univ, Fac Med, Med Biochem, Ankara, Turkey
[4] Ankara City Hosp, Med Biochem, Ankara, Turkey
关键词
pneumatic tube system (PTS); sample integrity; sample transportation; validation study; LACTATE-DEHYDROGENASE; DELIVERY-SYSTEM; BLOOD-SAMPLES; HEMOLYSIS; COUNT; VARIABILITY; PARAMETERS;
D O I
10.1515/cclm-2021-1235
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives Academics are far from a consensus regarding the effects of pneumatic tube system (PTS) delivery on sample integrity and laboratory test results. As for the reasons for conflicting opinions, each PTS is uniquely designed, sample tubes and patient characteristics differ among studies. This study aims to validate the PTS utilized in Ankara City Hospital for routine chemistry, coagulation, and hematology tests by comparing samples delivered via PTS and porter. Methods The study comprises 50 healthy volunteers. Blood samples were drawn into three biochemistry, two coagulation, and two hemogram tubes from each participant. Each of the duplicate samples was transferred to the emergency laboratory via Swiss log PTS (aka PTS-immediately) or by a porter. The last of the biochemistry tubes were delivered via the PTS, upon completion of coagulation of the blood (aka PTS-after). The results of the analysis in these groups were compared with multiple statistical analyses. Results The study did not reveal any correlation between the PTS and serum hemolysis index. There were statistically significant differences in several biochemistry tests. However, none of them reached the clinical significance threshold. Basophil and large unidentified cell (LUC) tests had poor correlations (r=0.47 and r=0.60; respectively) and reached clinical significance threshold (the average percentages of bias, 10.2%, and 15.4%, respectively). The remainder of the hematology and coagulation parameters did not reach clinical significance level either. Conclusions The modern PTS validated in this study is safe for sample transportation for routine chemistry, coagulation, and hematology tests frequently requested in healthy individuals except for basophil and LUC.
引用
收藏
页码:707 / 713
页数:7
相关论文
共 32 条
  • [1] The pneumatic tube system does not affect complete blood count results; a validation study at a tertiary care hospital
    Al-Riyami, A. Z.
    Al-Khabori, M.
    Al-Hadhrami, R. M.
    Al-Azwani, I. S.
    Davis, H. M.
    Al-Farsi, K. S.
    Alkindi, S. S.
    Daar, S. F.
    [J]. INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2014, 36 (05) : 514 - 520
  • [2] Pneumatic tube-transported blood samples in lithium heparinate gel separator tubes may be more susceptible to haemolysis than blood samples in serum tubes
    Boeckel-Frohnhoefer, Nicole
    Huebner, Ulrich
    Hummel, Bjoern
    Geisel, Juergen
    [J]. SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2014, 74 (07) : 599 - 602
  • [3] Analytical evaluation of the epoc® point-of-care blood analysis system in cardiopulmonary bypass patients
    Chen, Jianing
    Gorman, Monique
    O'Reilly, Bill
    Chen, Yu
    [J]. CLINICAL BIOCHEMISTRY, 2016, 49 (09) : 708 - 712
  • [4] Clinical Laboratory Improvement Amendments, NEW CLIA PROP RUL AC
  • [5] Changes of Serum Lactate Dehydrogenase and Potassium Levels Produced by a Pneumatic Tube System
    Cui, Ming
    Jing, Rongrong
    Wang, Haimin
    [J]. LABMEDICINE, 2009, 40 (12): : 728 - 731
  • [6] Effects of a pneumatic tube system on the hemolysis of blood samples: a PRISMA-compliant meta-analysis
    Ding, Xuemei
    Wen, Xiulin
    Wang, Liying
    Chen, Ting
    Zhou, Guangxia
    He, Hairong
    Xin, Xia
    [J]. SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2021, 81 (05) : 343 - 352
  • [7] Effect of Pneumatic Tube Delivery System Rate and Distance on Hemolysis of Blood Specimens
    Evliyaoglu, Osman
    Toprak, Gulten
    Tekin, Alicem
    Basarali, Mustafa Kemal
    Kilinc, Cumhur
    Colpan, Leyla
    [J]. JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2012, 26 (02) : 66 - 69
  • [8] Parameters for Validating a Hospital Pneumatic Tube System
    Farnsworth, Christopher W.
    Webber, Daniel M.
    Krekeler, James A.
    Budelier, Melissa M.
    Bartlett, Nancy L.
    Gronowski, Ann M.
    [J]. CLINICAL CHEMISTRY, 2019, 65 (05) : 694 - 702
  • [9] Pneumatic tube delivery system for blood samples reduces turnaround times without affecting sample quality
    Fernandes, CMB
    Worster, A
    Eva, K
    Hill, S
    McCallum, C
    [J]. JOURNAL OF EMERGENCY NURSING, 2006, 32 (02) : 139 - 143
  • [10] Validation of blood components transport through a pneumatic tube system
    Garcia, Lais Oliveira
    Speransa, Daniela Michelim Rodriguez
    Rodrigues, Cristiane Braga
    Benites, Rosimari Melgarejo
    Garcia, Mariana Timm
    Sekine, Leo
    Franz, Juliana Pires Marafon
    [J]. HEMATOLOGY TRANSFUSION AND CELL THERAPY, 2022, 44 (04) : 519 - 525