Gas-liquid separator for automated hydride generation and atomic absorption spectrometry

被引:14
作者
Laborda, F [1 ]
Gomez, MT [1 ]
Jimenez, MS [1 ]
Mir, JM [1 ]
Castillo, JR [1 ]
机构
[1] UNIV ZARAGOZA,FAC SCI,DEPT ANALYT CHEM,ANALYT SPECT & SENSOR GRP,E-50009 ZARAGOZA,SPAIN
关键词
gas-liquid separator; automated hydride generation; selenium; atomic absorption spectrometry;
D O I
10.1039/ja9961101121
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A straightforward phase separator for continuous hydride generation was developed. The separator uses a small plastic syringe that minimizes the typical dead volume of conventional separators and avoids analyte losses by adsorption on glass, The device is simple and easy to fit to any FI system. It was used for the AAS determination of selenium, with a detection limit of 6.5 mu g 1(-1). The method has a reproducibility of 2-3% expressed as the relative standard deviation.
引用
收藏
页码:1121 / 1122
页数:2
相关论文
共 8 条
[1]   BISMUTH(III) HYDRIDE GENERATION, ITS SEPARATION AND THE DETERMINATION OF BISMUTH(III) BY ATOMIC-ABSORPTION SPECTROMETRY USING FLOW-INJECTION [J].
CHAN, WF ;
HON, PK .
ANALYST, 1990, 115 (05) :567-569
[2]  
DAVIDOWSKI L, 1994, ATOM SPECTROSC, V15, P140
[3]   DETERMINATION OF TOTAL SELENIUM IN WATER BY ATOMIC-ABSORPTION SPECTROMETRY AFTER HYDRIDE GENERATION AND PRECONCENTRATION IN A COLD TRAP SYSTEM [J].
ORNEMARK, U ;
PETTERSSON, J ;
OLIN, A .
TALANTA, 1992, 39 (09) :1089-1096
[4]   DETERMINATION OF ULTRA-TRACE AMOUNTS OF SELENIUM(IV) BY FLOW-INJECTION HYDRIDE GENERATION ATOMIC-ABSORPTION SPECTROMETRY WITH ONLINE PRECONCENTRATION BY COPRECIPITATION WITH LANTHANUM HYDROXIDE [J].
TAO, GH ;
HANSEN, EH .
ANALYST, 1994, 119 (02) :333-337
[5]   RADIOTRACER EXAMINATION OF GAS-LIQUID SEPARATORS USED IN ARSENIC SPECIATION BY HYDRIDE GENERATION - AAS [J].
VANELTEREN, JT ;
DAS, HA ;
BAX, D .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-ARTICLES, 1993, 174 (01) :133-144
[6]   DETERMINATION OF SELENIUM AND ARSENIC IN MINERAL WATERS WITH HYDRIDE GENERATION ATOMIC-ABSORPTION SPECTROMETRY [J].
VEBER, M ;
CUJES, K ;
GOMISCEK, S .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (03) :285-290
[7]  
Vijan P. N., 1974, Atomic Absorption Newsletter, V13, P33
[8]   MEMBRANE GAS-LIQUID SEPARATOR FOR FLOW-INJECTION HYDRIDE-GENERATION ATOMIC-ABSORPTION SPECTROMETRY [J].
YAMAMOTO, M ;
TAKADA, K ;
KUMAMARU, T ;
YASUDA, M ;
YOKOYAMA, S ;
YAMAMOTO, Y .
ANALYTICAL CHEMISTRY, 1987, 59 (19) :2446-2448