Design and performance of collimated coincidence point sources for simultaneous transmission measurements in 3-D PET

被引:15
作者
Watson, CC [1 ]
Eriksson, L
Casey, ME
Jones, WF
Moyers, JC
Miller, S
Hamill, J
Van Lingen, A
Bendriem, B
Nutt, R
机构
[1] CTI PET Syst Inc, Knoxville, TN 37932 USA
[2] Free Univ Amsterdam Hosp, NL-1007 MB Amsterdam, Netherlands
关键词
attentuation correction; collimated coincidence point sources; positron emission tomography (PET); reference detector; simultaneous transmission measurement;
D O I
10.1109/23.940145
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We have implemented a simultaneous emission-transmission measurement for three-dimensional positron emission tomography (3-D PET) using a collimated coincidence point source design employing a fast, dedicated, reference detector close to the transmission source. This design reduces the effects of randoms, scatter, dead time, and sensitivity loss on the emission data compared to previous implementations. It also greatly reduces the effect of emission contamination of the transmission data compared to the use of rod sources. In this paper, we present performance characterizations of this measurement technique on both the Siemens/CTI ECAT ART and PET/SPECT tomographs. The main effect of the transmission sources on the emission measurement is an increased randoms rate, which leads to a 10-25% reduction in NECR at specific activities >2 kBq/mL in a 21-cm-diameter phantom on the PET/SPECT. Emission contamination effects on the transmission measurement are estimated to be less than I % for up to 20 kBq/mL in a 21-cm phantom on the PET/SPECT. Both the emission and transmission NECR are dominated by the effects of randoms. Considering the effects of both emission and transmission noise on the final corrected image, it appears that 3-6 kBq/mL of emitter concentration is an optimal imaging range for simultaneous acquisitions. We present the first images of a normal volunteer using this system on a Siemens/CTI PET/SPECT tomograph.
引用
收藏
页码:673 / 679
页数:7
相关论文
共 10 条
[1]   PROBLEMS IN SIGNAL-TO-NOISE RATIO FOR ATTENUATION CORRECTION IN HIGH-RESOLUTION PET [J].
DAHLBOM, M ;
HOFFMAN, EJ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1987, 34 (01) :288-293
[2]   OPTIMIZING ROD WINDOW WIDTH IN POSITRON EMISSION TOMOGRAPHY [J].
JONES, WF ;
DIGBY, WM ;
LUK, WK ;
CASEY, ME ;
BYARS, LG .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1995, 14 (02) :266-270
[3]   Fast-channel LSO detectors and fiber-optic encoding for excellent dual photon transmission measurements in PET [J].
Jones, WF ;
Moyers, JC ;
Casey, ME ;
Watson, CC ;
Nutt, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1999, 46 (04) :979-984
[4]  
MEIKLE SR, 1995, J NUCL MED, V36, P1680
[5]  
MICHEL C, 1999, 1998 IEEE NSS MIC, V2, P1323
[6]   EVALUATING EFFICIENCY OF IMAGING PROCESSES [J].
SHAW, R .
REPORTS ON PROGRESS IN PHYSICS, 1978, 41 (07) :1103-1155
[7]   SIMULTANEOUS TRANSMISSION AND EMISSION SCANS IN POSITRON EMISSION TOMOGRAPHY [J].
THOMPSON, CJ ;
RANGER, NT ;
EVANS, AC .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1989, 36 (01) :1011-1016
[8]  
Townsend David W., 1999, Clin Positron Imaging, V2, P5, DOI 10.1016/S1095-0397(98)00057-0
[9]  
Watson CC, 1998, 1997 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1 & 2, P1366
[10]  
WATSON CC, COINCIDENCE TRANSMIS