Comparison of Accuracy of Pedicle Screw Insertion Among 4 Guided Technologies in Spine Surgery

被引:63
作者
Fan, Yong [1 ]
Du, Jinpeng [1 ,2 ]
Zhang, Jianan [1 ]
Liu, Shichang [1 ]
Xue, Xukai [1 ]
Huang, Yunfei [1 ]
Zhang, Jing [1 ]
Hao, Dingjun [1 ,2 ]
机构
[1] Xi An Jiao Tong Univ, Affiliated Honghui Hosp, Dept Spine Surg, Xian, Shaanxi, Peoples R China
[2] Yanan Univ, Med Coll, Yanan, Shanxi, Peoples R China
来源
MEDICAL SCIENCE MONITOR | 2017年 / 23卷
关键词
Bone Screws; Fluoroscopy; Robotics; Surgery; Computer-Assisted; Templates; Genetic; PLACEMENT ACCURACY; LUMBAR SPINE; GUIDANCE;
D O I
10.12659/MSM.905713
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Background: As an available new tool for spinal surgery, robotic technology holds great potential and has been demonstrated to have better clinical outcomes compared with traditional techniques. However, it has not been compared with other assisted tools for the treatment of lumbar degenerative disease. This article focused on studying such variances. Material/Methods: A total of 176 pedicle screws were inserted in 39 patients using a spine robot (group 1), 134 screws were implanted in 28 patients using navigational template (group 2), 234 screws were implanted in 51 patients by O-arm-based navigation (group 3), and 346 screws were implanted in 72 patients by fluoroscopy-guided assistance (group 4). The screw position was evaluated using postoperative scans according to Rampersaud A to D classification, and other secondary data were also collected. Results: "Perfect" pedicle screw insertion (Grade A) was 90.34%, 91.79%, 84.19%, and 65.03% of groups 1-4, respectively. "Clinically acceptable" screw implantation (Grade A+B) was 94.32%, 95.52, 90.60%, and 78.03% in groups 1-4, respectively. Deviation sagittal (degrees) respectively was 3 +/- 9, 2 +/- 10, 4 +/- 7, and 10 +/- 8 degrees in groups 1-4, respectively. Deviation transversal (degrees) screw insertion was 3 +/- 8, 3 +/- 7, 4 +/- 9, and 8 +/- 13 degrees in groups 1-4, respectively. Statistical analysis showed group 1 had no significant difference in the accuracy of "Perfect and Clinical acceptable" as well as deviation sagittal or transversal, respectively, compared with groups 2 and 3 but not group 4. Conclusions: Robotic-assistance technology no clear advantage in terms of accuracy compared to the navigation template or O-arm systems for screw implantation, but it significantly reduced adverse events, fluoroscopy time per screw, postoperative stay, and blood loss.
引用
收藏
页码:5960 / 5968
页数:9
相关论文
共 19 条
  • [11] The accuracy of pedicle screw placement using intraoperative image guidance systems
    Mason, Alexander
    Paulsen, Renee
    Babuska, Jason M.
    Rajpal, Sharad
    Burneikiene, Sigita
    Nelson, E. Lee
    Villavicencio, Alan T.
    [J]. JOURNAL OF NEUROSURGERY-SPINE, 2014, 20 (02) : 196 - 203
  • [12] Robotics for spinal operations: reality or Alice in Wonderland?
    Matis, Georgios K.
    Silva, Danilo
    Chrysou, Olga I.
    Birbilis, Theodossios A.
    Bernardo, Antonio
    Stieg, Philip E.
    [J]. INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2012, 8 (01) : 125 - 126
  • [13] Lumbar and sacral pedicle screw placement using a template does not improve the midterm pain and disability outcome in comparison with free-hand method
    Merc M.
    Recnik G.
    Krajnc Z.
    [J]. European Journal of Orthopaedic Surgery & Traumatology, 2017, 27 (5) : 583 - 589
  • [14] A multi-level rapid prototyping drill guide template reduces the perforation risk of pedicle screw placement in the lumbar and sacral spine
    Merc, Matjaz
    Drstvensek, Igor
    Vogrin, Matjaz
    Brajlih, Tomaz
    Recnik, Gregor
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2013, 133 (07) : 893 - 899
  • [15] Electromagnetic Real Time Navigation in the Region of the Posterior Pelvic Ring: An Experimental In-Vitro Feasibility Study and Comparison of Image Guided Techniques
    Pishnamaz, Miguel
    Wilkmann, Christoph
    Na, Hong-Sik
    Pfeffer, Jochen
    Haenisch, Christoph
    Janssen, Max
    Bruners, Philipp
    Kobbe, Philipp
    Hildebrand, Frank
    Schmitz-Rode, Thomas
    Pape, Hans-Christoph
    [J]. PLOS ONE, 2016, 11 (02):
  • [16] Accuracy of Robot-Assisted Placement of Lumbar and Sacral Pedicle Screws A Prospective Randomized Comparison to Conventional Freehand Screw Implantation
    Ringel, Florian
    Stueer, Carsten
    Reinke, Andreas
    Preuss, Alexander
    Behr, Michael
    Auer, Florian
    Stoffel, Michael
    Meyer, Bernhard
    [J]. SPINE, 2012, 37 (08) : E496 - E501
  • [17] Safety and accuracy of robot-assisted versus fluoroscopy-guided pedicle screw insertion for degenerative diseases of the lumbar spine: a matched cohort comparison
    Schatlo, Bawarjan
    Molliqaj, Granit
    Cuvinciuc, Victor
    Kotowski, Marc
    Schaller, Karl
    Tessitore, Enrico
    [J]. JOURNAL OF NEUROSURGERY-SPINE, 2014, 20 (06) : 636 - 643
  • [18] Pedicle screw insertion accuracy with different assisted methods: a systematic review and meta-analysis of comparative studies
    Tian, Nai-Feng
    Huang, Qi-Shan
    Zhou, Ping
    Zhou, Yang
    Wu, Rui-Kai
    Lou, Yi
    Xu, Hua-Zi
    [J]. EUROPEAN SPINE JOURNAL, 2011, 20 (06) : 846 - 859
  • [19] Electromagnetic navigation for percutaneous guide-wire insertion: Accuracy and efficiency compared to conventional fluoroscopic guidance
    von Jako, R. A.
    Carrino, J. A.
    Yonemura, K. S.
    Noda, G. A.
    Zhue, W.
    Blaskiewicz, D.
    Rajue, M.
    Groszmann, D. E.
    Weber, G.
    [J]. NEUROIMAGE, 2009, 47 : T127 - T132