Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients

被引:147
作者
Frasca, Denis [1 ,2 ,3 ]
Dahyot-Fizelier, Claire [1 ,2 ,3 ]
Catherine, Karen [1 ]
Levrat, Quentin [1 ]
Debaene, Bertrand [1 ,2 ,3 ]
Mimoz, Olivier [1 ,2 ,3 ]
机构
[1] Ctr Hosp Univ Poitiers, Poitiers, France
[2] INSERM, ERI 23, F-75654 Paris 13, France
[3] Univ Poitiers, UFR Med Pharm, Poitiers, France
关键词
anemia; blood transfusion; hemoglobinometry methods; intensive care; monitoring; point-of-care; BLOOD-TRANSFUSION; ANEMIA; ANALYZER; SCORE;
D O I
10.1097/CCM.0b013e3182227e2d
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To determine whether noninvasive hemoglobin measurement by Pulse CO-Oximetry could provide clinically acceptable absolute and trend accuracy in critically ill patients, compared to other invasive methods of hemoglobin assessment available at bedside and the gold standard, the laboratory analyzer. Design: Prospective study. Setting: Surgical intensive care unit of a university teaching hospital. Patients: Sixty-two patients continuously monitored with Pulse CO-Oximetry (Masimo Radical-7). Interventions: None. Measurements and Results: Four hundred seventy-one blood samples were analyzed by a point-of-care device (HemoCue 301), a satellite lab CO-Oximeter (Siemens RapidPoint 405), and a laboratory hematology analyzer (Sysmex XT-2000i), which was considered the reference device. Hemoglobin values reported from the invasive methods were compared to the values reported by the Pulse CO-Oximeter at the time of blood draw. When the case-to-case variation was assessed, the bias and limits of agreement were 0.0 +/- 1.0 g/dL for the Pulse CO-Oximeter, 0.3 +/- 1.3g/dL for the point-of-care device, and 0.9 +/- 0.6 g/dL for the satellite lab CO-Oximeter compared to the reference method. Pulse CO-Oximetry showed similar trend accuracy as satellite lab CO-Oximetry, whereas the point-of-care device did not appear to follow the trend of the laboratory analyzer as well as the other test devices. Conclusion: When compared to laboratory reference values, hemoglobin measurement with Pulse CO-Oximetry has absolute accuracy and trending accuracy similar to widely used, invasive methods of hemoglobin measurement at bedside. Hemoglobin measurement with pulse CO-Oximetry has the additional advantages of providing continuous measurements, noninvasively, which may facilitate hemoglobin monitoring in the intensive care unit. (Crit Care Med 2011; 39:2277-2282)
引用
收藏
页码:2277 / 2282
页数:6
相关论文
共 23 条
  • [1] Anemia, blood loss, and blood transfusions in North American children in the intensive care unit
    Bateman, Scot T.
    Lacroix, Jacques
    Boven, Katia
    Forbes, Peter
    Barton, Roger
    Thomas, Neal J.
    Jacobs, Brian
    Markovitz, Barry
    Goldstein, Brahm
    Hanson, James H.
    Li, H. Agnes
    Randolph, Adrienne G.
    [J]. AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2008, 178 (01) : 26 - 33
  • [2] Agreement between methods of measurement with multiple observations per individual
    Bland, J. Martin
    Altman, Douglas G.
    [J]. JOURNAL OF BIOPHARMACEUTICAL STATISTICS, 2007, 17 (04) : 571 - 582
  • [3] Burdick R.K., 1992, Confidence intervals on variance components
  • [4] The CRIT Study: Anemia and blood transfusion in the critically ill - Current clinical practice in the United States
    Corwin, HL
    Gettinger, A
    Pearl, RG
    Fink, MP
    Levy, MM
    Abraham, E
    MacIntyre, NR
    Shabot, M
    Duh, MS
    Shapiro, MJ
    [J]. CRITICAL CARE MEDICINE, 2004, 32 (01) : 39 - 52
  • [5] Hemoximetry as the "Gold Standard"? Error Assessment Based on Differences Among Identical Blood Gas Analyzer Devices of Five Manufacturers
    Gehring, Hartmut
    Duembgen, Lutz
    Peterlein, Mareike
    Hagelberg, Soehnke
    Dibbelt, Leif
    [J]. ANESTHESIA AND ANALGESIA, 2007, 105 : S24 - S30
  • [6] Lamhaut L, 2010, EUR J ANAESTH, V27, p3AP7
  • [7] A NEW SIMPLIFIED ACUTE PHYSIOLOGY SCORE (SAPS-II) BASED ON A EUROPEAN NORTH-AMERICAN MULTICENTER STUDY
    LEGALL, JR
    LEMESHOW, S
    SAULNIER, F
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1993, 270 (24): : 2957 - 2963
  • [8] Macknet M, 2007, ANESTH ANALG, V104, pSCA1
  • [9] The Accuracy of Noninvasive and Continuous Total Hemoglobin Measurement by Pulse CO-Oximetry in Human Subjects Undergoing Hemodilution
    Macknet, Mark R.
    Allard, Martin
    Applegate, Richard L.
    Rook, James
    [J]. ANESTHESIA AND ANALGESIA, 2010, 111 (06) : 1424 - 1426
  • [10] EVALUATION OF 24-SECOND CYANIDE-CONTAINING AND CYANIDE-FREE METHODS FOR WHOLE-BLOOD HEMOGLOBIN ON THE TECHNICON H-STAR-1TM ANALYZER WITH NORMAL AND ABNORMAL-BLOOD SAMPLES
    MALIN, MJ
    SCLAFANI, LD
    WYATT, JL
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1989, 92 (03) : 286 - 294