Detection of coronary stenoses at rest with myocardial contrast echocardiography

被引:28
|
作者
Wei, K
Tong, KL
Belcik, T
Rafter, P
Ragosta, M
Wang, XQ
Kaul, S
机构
[1] Univ Virginia, Cardivasc Imaging Ctr, Div Cardiovasc, Dept Med, Charlottesville, VA USA
[2] Univ Virginia, Dept Hlth Evaluat Sci, Div Biostat & Epidemiol, Charlottesville, VA USA
[3] Philips Ultrasound, Andover, MA USA
关键词
contrast media; coronary disease; diagnosis; echocardiography; imaging;
D O I
10.1161/CIRCULATIONAHA.104.513887
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background - We hypothesized that autoregulatory changes in arteriolar blood volume (aBV) that develop distal to a stenosis can be measured with myocardial contrast echocardiography, allowing coronary stenosis detection at rest without recourse to stress. Methods and Results - Patients with varying degrees of coronary artery stenosis on quantitative angiography underwent high-mechanical-index myocardial contrast echocardiography at 15 Hz to allow measurement of phasic changes in aBV in large intramyocardial vessels using either Definity (group 1; n = 22) or Imagent ( group 2; n = 22). Progressive increases in the background-subtracted systolic/diastolic aBV signal ratio were noted between each level ( none, mild [< 50%], moderate [50% to 75%], and severe [> 75%]) of stenosis severity for both group 1 (0.09 +/- 0.13, 0.13 +/- 0.08, 0.58 +/- 0.22, and 0.77 +/- 0.40; P < 0.001) and group 2 (0.10 +/- 0.05, 0.27 +/- 0.18, 0.39 +/- 0.28, and 0.74 +/- 0.37; P < 0.0001) patients. A systolic/diastolic aBV signal ratio of > 0.34 provided a sensitivity and specificity of 80% and 71%, respectively, for the detection of > 75% coronary stenosis in group 1 patients, whereas a ratio of > 0.43 provided a sensitivity and specificity of 89% and 74%, respectively, for the detection of > 75% stenosis in group 2 patients. Conclusions - Both the presence and severity of a physiologically significant coronary stenosis can be detected at rest by measuring the increase in aBV on myocardial contrast echocardiography that occurs distally to the stenosis without recourse to any form of stress.
引用
收藏
页码:1154 / 1160
页数:7
相关论文
共 50 条
  • [1] Stress echo without the stress - Detection of coronary stenosis at rest by myocardial contrast echocardiography
    Grayburn, PA
    CIRCULATION, 2005, 112 (08) : 1085 - 1087
  • [2] DETECTION OF CORONARY STENOSES AND QUANTIFICATION OF THE DEGREE AND SPATIAL EXTENT OF BLOOD-FLOW MISMATCH DURING CORONARY HYPEREMIA WITH MYOCARDIAL CONTRAST ECHOCARDIOGRAPHY
    ISMAIL, S
    JAYAWEERA, AR
    GOODMAN, NC
    CAMARANO, GP
    SKYBA, DM
    KAUL, S
    CIRCULATION, 1995, 91 (03) : 821 - 830
  • [3] Comparative value of dobutamine and adenosine stress in the detection of coronary stenosis with myocardial contrast echocardiography
    Lafitte, S
    Matsugata, H
    Peters, B
    Togni, M
    Strachan, M
    Kwan, OL
    DeMaria, AN
    CIRCULATION, 2001, 103 (22) : 2724 - 2730
  • [4] Usefulness of dobutamine stress myocardial contrast echocardiography for assessing coronary artery disease
    Guo, SZ
    Shu, XH
    Pan, CZ
    Li, YL
    Ge, JB
    Chen, HZ
    CHINESE MEDICAL JOURNAL, 2005, 118 (21) : 1766 - 1772
  • [5] Myocardial contrast echocardiography - A 25-year retrospective
    Kaul, Sanjiv
    CIRCULATION, 2008, 118 (03) : 291 - 308
  • [6] Assessment of Coronary Microcirculation with Myocardial Contrast Echocardiography
    Danijela, Trifunovic
    Jelena, Dudic
    Olga, Petrovic
    Zorana, Vasiljevic-Pokrajcic
    CURRENT PHARMACEUTICAL DESIGN, 2018, 24 (25) : 2943 - 2949
  • [7] Adenosine myocardial contrast echo in intermediate severity coronary stenoses: a prospective two-center study
    Firschke, Christian
    Andrassy, Peter
    Linka, Andre Z.
    Busch, Raymonde
    Martinoff, Stefan
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2007, 23 (03): : 311 - 321
  • [8] Adenosine myocardial contrast echo in intermediate severity coronary stenoses: a prospective two-center study
    Christian Firschke
    Peter Andrássy
    Andre Z. Linka
    Raymonde Busch
    Stefan Martinoff
    The International Journal of Cardiovascular Imaging, 2007, 23 : 311 - 321
  • [9] Effect of coronary stenosis on adjacent bed flow reserve - Assessment of microvascular mechanisms using myocardial contrast echocardiography
    Pacella, John J.
    Villanueva, Flordeliza S.
    CIRCULATION, 2006, 114 (18) : 1940 - 1947
  • [10] Noninvasive quantification of coronary blood flow reserve in humans using myocardial contrast echocardiography
    Wei, K
    Ragosta, M
    Thorpe, J
    Coggins, M
    Moos, S
    Kaul, S
    CIRCULATION, 2001, 103 (21) : 2560 - 2565