Monte Carlo calculated TG-43 dosimetry parameters for the SeedLink™ 125iodine brachytherapy system

被引:2
作者
Medich, DC
Munro, JJ
机构
[1] Univ Massachusetts, Radiat Lab, Lowell, MA 01854 USA
[2] Implant Sci Corp, Wakefield, MA 01880 USA
关键词
brachytherapy; iodine-125; prostate; radiation dosimetry; Monte Carlo;
D O I
10.1118/1.1601914
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The SeedLink(TM) brachytherapy system is comprised from an assembly of I-Plant(TM) 3500 interstitial brachytherapy seeds and bioresorbable spacers joined together by a 6-mm-long titanium sleeve centered over each seed. This device is designed to maintain specified spacing between seeds during treatment thereby decreasing implant preparation time and reducing radionuclide migration within the prostate and periprostatic region. Reliable clinical treatment and planning applications necessitate accurate dosimetric data for source evaluation, therefore the authors report the results of a Monte Carlo study designed to calculate the AAPM Task Group Report No. 43 dosimetric parameters for the SeedLink brachytherapy source and compare these values against previously published Monte Carlo study results of the I-Plant 3500 brachytherapy seed. For this investigation, a total of 1 X 10(9) source photon histories were processed for each set of in-water and in-air calculations using the MCNP4C2 Monte Carlo radiation transport code (RSICC). Statistically, the radial dose function, g(r), and the dose-rate constant, Lambda, were identical to the values calculated previously for the Model 3500 with the dose-rate constant evaluated to be Lambda=1.000+/-0.026 cGyh(-1) U-1. The titanium sleeve used in SeedLink to bind together Model 3500 seeds and spacers resulted in slightly greater dosimetric anisotropy as exhibited in the anisotropy function, F(r, theta), the anisotropy factor, phi(an) (r), and the anisotropy constant, (φ) over bar (an), which was calculated to be (φ) over bar (an)=0.91+/-0.01, or roughly 2% lower than the value calculated previously for the Model 3500. These results indicate that the radiological characteristics of the SeedLink dosimetry system are comparable to those obtained for previously characterized single seeds such as the Implant Sciences Model 3500 I-Plant seed. (C) 2003 American Association of Physicists in Medicine.
引用
收藏
页码:2503 / 2508
页数:6
相关论文
共 50 条
  • [21] Clinical brachytherapy with neutron emitting 252Cf sources and adherence to AAPM TG-43 dosimetry protocol
    Rivard, MJ
    Wierzbicki, JG
    Van den Heuvel, F
    Martin, RC
    McMahon, RR
    MEDICAL PHYSICS, 1999, 26 (01) : 87 - 96
  • [22] Monte Carlo calculation of dosimetry parameters for a brachytherapy source
    Belousov, A. V.
    Kalachev, A. A.
    Osipov, A. S.
    MOSCOW UNIVERSITY PHYSICS BULLETIN, 2014, 69 (06) : 535 - 541
  • [23] Monte Carlo and experimental dosimetry of an 125I brachytherapy seed
    Dolan, James
    Li, Zuofeng
    Williamson, Jeffrey F.
    MEDICAL PHYSICS, 2006, 33 (12) : 4675 - 4684
  • [24] Evaluation of a new brachytherapy iodine-125 source by AAPM TG43 formalism
    Wallace, RE
    Fan, JJ
    MEDICAL PHYSICS, 1998, 25 (11) : 2190 - 2196
  • [25] A simple modification of TG-43 based brachytherapy dosimetry with improved fitting functions: Application to the selectSeed source
    Juan-Senabre, Xavier J.
    Porras, Ignacio
    Lallena, Antonio M.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2013, 29 (04): : 403 - 411
  • [26] AAPM TG-43U1 formalism adaptation and Monte Carlo dosimetry simulations of multiple-radionuclide brachytherapy sources
    Nuttens, VE
    Lucas, S
    MEDICAL PHYSICS, 2006, 33 (04) : 1101 - 1107
  • [27] Monte Carlo derivation of AAPM TG-43 dosimetric parameters for GZP6 Co-60 HDR sources
    Tabrizi, Sanaz Hariri
    Asl, Alireza Kamali
    Azma, Zohreh
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2012, 28 (02): : 153 - 160
  • [28] Experimental determination of the TG-43 dosimetric characteristics of EchoSeed™ model 6733 125I brachytherapy source
    Meigooni, AS
    Dini, SA
    Sowards, K
    Hayes, JL
    Al-Otoom, A
    MEDICAL PHYSICS, 2002, 29 (06) : 939 - 942
  • [29] Simulation of the shielding effects of an applicator on the AAPM TG-43 parameters of CS-137 Selectron LDR brachytherapy sources
    Sina, S.
    Faghihi, R.
    Meigooni, A. S.
    Mehdizadeh, S.
    Zehtabian, M.
    Mosleh-Shirazi, M. A.
    IRANIAN JOURNAL OF RADIATION RESEARCH, 2009, 7 (03): : 135 - 140
  • [30] Comparison of TG-43 dosimetric parameters of brachytherapy sources obtained by three different versions of MCNP codes
    Zaker, Neda
    Zehtabian, Mehdi
    Sina, Sedigheh
    Koontz, Craig
    Meigooni, Ali S.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (02): : 379 - 390