Monte Carlo calculated TG-43 dosimetry parameters for the SeedLink™ 125iodine brachytherapy system

被引:2
|
作者
Medich, DC
Munro, JJ
机构
[1] Univ Massachusetts, Radiat Lab, Lowell, MA 01854 USA
[2] Implant Sci Corp, Wakefield, MA 01880 USA
关键词
brachytherapy; iodine-125; prostate; radiation dosimetry; Monte Carlo;
D O I
10.1118/1.1601914
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The SeedLink(TM) brachytherapy system is comprised from an assembly of I-Plant(TM) 3500 interstitial brachytherapy seeds and bioresorbable spacers joined together by a 6-mm-long titanium sleeve centered over each seed. This device is designed to maintain specified spacing between seeds during treatment thereby decreasing implant preparation time and reducing radionuclide migration within the prostate and periprostatic region. Reliable clinical treatment and planning applications necessitate accurate dosimetric data for source evaluation, therefore the authors report the results of a Monte Carlo study designed to calculate the AAPM Task Group Report No. 43 dosimetric parameters for the SeedLink brachytherapy source and compare these values against previously published Monte Carlo study results of the I-Plant 3500 brachytherapy seed. For this investigation, a total of 1 X 10(9) source photon histories were processed for each set of in-water and in-air calculations using the MCNP4C2 Monte Carlo radiation transport code (RSICC). Statistically, the radial dose function, g(r), and the dose-rate constant, Lambda, were identical to the values calculated previously for the Model 3500 with the dose-rate constant evaluated to be Lambda=1.000+/-0.026 cGyh(-1) U-1. The titanium sleeve used in SeedLink to bind together Model 3500 seeds and spacers resulted in slightly greater dosimetric anisotropy as exhibited in the anisotropy function, F(r, theta), the anisotropy factor, phi(an) (r), and the anisotropy constant, (φ) over bar (an), which was calculated to be (φ) over bar (an)=0.91+/-0.01, or roughly 2% lower than the value calculated previously for the Model 3500. These results indicate that the radiological characteristics of the SeedLink dosimetry system are comparable to those obtained for previously characterized single seeds such as the Implant Sciences Model 3500 I-Plant seed. (C) 2003 American Association of Physicists in Medicine.
引用
收藏
页码:2503 / 2508
页数:6
相关论文
共 50 条
  • [11] Perturbation of TG-43 parameters of the brachytherapy sources under insufficient scattering materials
    Zehtabian, Mehdi
    Sina, Sedigheh
    Faghihi, Reza
    Meigooni, Ali
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (03): : 164 - 176
  • [12] Monte Carlo calculation of the TG-43 dosimetric parameters of a new BEBIG Ir-192HDR source
    Granero, D
    Pérez-Calatayud, J
    Ballester, F
    RADIOTHERAPY AND ONCOLOGY, 2005, 76 (01) : 79 - 85
  • [13] Dosimetric and radiobiological comparison of TG-43 and Monte Carlo calculations in 192Ir breast brachytherapy applications
    Peppa, V.
    Pappas, E. P.
    Karaiskos, P.
    Major, T.
    Polgar, C.
    Papagiannis, P.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (10): : 1245 - 1251
  • [14] RapidBrachyTG43: A Geant4-based TG-43 parameter and dose calculation module for brachytherapy dosimetry
    Kalinowski, Jonathan
    Enger, Shirin A.
    MEDICAL PHYSICS, 2024, 51 (05) : 3746 - 3757
  • [15] Comparison of dosimetry parameters of two commercially available Iodine brachytherapy seeds using Monte Carlo calculations
    Zehtabian, M.
    Faghihi, R.
    Sina, S.
    IRANIAN JOURNAL OF RADIATION RESEARCH, 2010, 7 (04): : 217 - 222
  • [16] EGSnrc Monte Carlo calculated dosimetry parameters for 192Ir and 169Yb brachytherapy sources
    Taylor, R. E. P.
    Rogers, D. W. O.
    MEDICAL PHYSICS, 2008, 35 (11) : 4933 - 4944
  • [17] Prostate cancer Monte Carlo dose model with 177Lutetium and 125Iodine treatments
    Juste, B.
    Miro, R.
    Morato, S.
    Verdu, G.
    Peris, S.
    RADIATION PHYSICS AND CHEMISTRY, 2020, 174 (174)
  • [18] Dosimetry of 192Ir wires for LDR interstitial brachytherapy following the AAPM TG-43 dosimetric formalism
    Karaiskos, P
    Papagiannis, P
    Angelopoulos, A
    Sakelliou, L
    Baltas, D
    Sandilos, P
    Vlachos, L
    MEDICAL PHYSICS, 2001, 28 (02) : 156 - 166
  • [19] Monte Carlo dosimetry of the selectSeed 125I interstitial brachytherapy seed
    Karaiskos, P
    Papagiannis, P
    Sakelliou, L
    Anagnostopoulos, G
    Baltas, D
    MEDICAL PHYSICS, 2001, 28 (08) : 1753 - 1760
  • [20] Monte Carlo calculation of dosimetry parameters for a brachytherapy source
    A. V. Belousov
    A. A. Kalachev
    A. S. Osipov
    Moscow University Physics Bulletin, 2014, 69 : 535 - 541