Proposals for the type tests criteria and calibration conditions of passive eye lens dosemeters to be used in interventional cardiology and radiology workplaces

被引:17
作者
Bordy, J. M. [1 ]
Daures, J. [1 ]
Denoziere, M. [1 ]
Gualdrini, G. [2 ]
Ginjaume, M. [3 ]
Carinou, E. [4 ]
Vanhavere, F. [5 ]
机构
[1] Inst CEA LIST, Ctr Saclay, DRT, LIST,LNHB, F-91191 Gif Sur Yvette, France
[2] ENEA BAS ION IRP Radiat Protect Inst, I-40136 Bologna, BO, Italy
[3] Univ Politecn Cataluna, E-08028 Barcelona, Spain
[4] GAEC, Athens 15310, Greece
[5] CEN SCK, Belgian Nucl Res Ctr, B-2400 Mol, Belgium
关键词
Type test; Calibration; Conversion coefficient; Interventional radiology; Interventional cardiology; Eye lens dosimetry; THERMOLUMINESCENCE; EXPOSURE; CATARACT;
D O I
10.1016/j.radmeas.2011.07.035
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
The paper is aimed at making a proposal for the type test and calibration of eye lens passive dosemeters especially used in the interventional cardiology/radiology (IC/IR). Starting from the only existing standard dealing with eye lens dosimetry using TLDs (ISO 12794), parameters such as, detection threshold, energy and angle dependence of response criteria have been reviewed and it has been tried to harmonise them as much as possible with the IEC 62387 requirements, taking into account the particular use at IC/IR workplaces. Conversion coefficients from air kerma to dose equivalent at 3 mm depth for RQR and ISO radiation qualities, employed for type test and calibration purposes, have been calculated in a new phantom introduced within the ORAMED (Optimization of RAdiation protection for MEDical staff) project. This phantom is more representative of the head so that the estimation of H-lens by H-P(3) is more accurate. (C) 2011 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1235 / 1238
页数:4
相关论文
共 22 条
  • [1] [Anonymous], 2011, Statement on tissue reactions
  • [2] [Anonymous], 12794 ISO
  • [3] Dose conversion coefficients for electron exposure of the human eye lens
    Behrens, R.
    Dietze, G.
    Zankl, M.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (13) : 4069 - 4087
  • [4] BORDY JM, 2011, RAD PROT DO IN PRESS
  • [5] BORDY JM, 2008, P INT WORKSH UNC ASS
  • [6] Recommendations to reduce extremity and eye lens doses in interventional radiology and cardiology
    Carinou, E.
    Brodecki, M.
    Domienik, J.
    Donadille, L.
    Koukorava, C.
    Krim, S.
    Nikodemova, D.
    Ruiz-Lopez, N.
    Sans-Merce, M.
    Struelens, L.
    Vanhavere, F.
    [J]. RADIATION MEASUREMENTS, 2011, 46 (11) : 1324 - 1329
  • [7] From films to thermoluminescence dosemeters:: The Greek Atomic Energy Commission experience
    Carinou, E
    Drikos, G
    Hourdakis, C
    Hyvönen, H
    Kamenopoulou, V
    [J]. RADIATION PROTECTION DOSIMETRY, 2001, 96 (1-3) : 205 - 208
  • [8] Risk of cataract after exposure to low doses of ionizing radiation: A 20-year prospective cohort study among US radiologic technologists
    Chodick, Gabriel
    Bekiroglu, Nural
    Hauptmann, Michael
    Alexander, Bruce H.
    Freedman, D. Michal
    Doody, Michele Morin
    Cheung, Li C.
    Simon, Steven L.
    Weinstock, Robert M.
    Bouville, Andre
    Sigurdson, Alice J.
    [J]. AMERICAN JOURNAL OF EPIDEMIOLOGY, 2008, 168 (06) : 620 - 631
  • [9] Daures J., 2009, CEA SACLAY
  • [10] European Commission, 2009, RAD PROT, P160, DOI [10.2768/19817, DOI 10.2768/19817]