Safety of 3 Tesla Magnetic Resonance Imaging in Patients With Auditory Brainstem Implants

被引:0
作者
Wichova, Helena [1 ]
Peng, Kevin [1 ]
Ledbetter, Luke [2 ]
Slattery, William [1 ]
Brackmann, Derald [1 ]
Lekovic, Gregory [1 ]
机构
[1] Univ Calif Los Angeles, David Geffen Sch Med, House Clin, Los Angeles, CA 90095 USA
[2] Univ Calif Los Angeles, David Geffen Sch Med, Dept Radiol Sci, Los Angeles, CA 90095 USA
关键词
Auditory brainstem implant; Magnetic resonance imaging; Neurofibromatosis-2; NEUROFIBROMATOSIS TYPE-2; COCHLEAR; MRI;
D O I
10.1097/MAO.0000000000003399
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objective: To evaluate the safety of 3 Tesla (T) magnetic resonance imaging (MRI) in patients with auditory brainstem implants (ABI) with the magnet removed at implantation and report incidence of complications. Study Design: Retrospective chart review. Setting: Tertiary neurotology ambulatory practice. Patients: Patients with diagnosis of Neurofibromatosis, type 2 (NF2) with functional ABIs. Intervention(s): Observational recordings. Main Outcome Measure(s): Of the 89 patients meeting inclusion criteria, 7 patients underwent 3T MRI, with a total of 39 scans done. Three patients had 1 scan each, one patient had 4 scans, one patient had 5 scans, one patient had 6 scans, and one patient had 21 scans. The mean time between ABI placement and first 3 T scan was 118 +/- 73 months. The most common indication for imaging was surveillance of NF2 lesions. The most frequent scans were MRI brain (25.6%), followed by MRI of cervical (15%), thoracic (15%) and lumbar (15%) spine, and MRI IAC (8%). There were no reported complications for any of the scans. No scans were interrupted due to patient discomfort. There were no device malfunctions. Conclusions: 3 T MRIs are safe in patients with ABIs as long as the magnet is removed. It is recommended that the magnet be removed at the time of implantation in all NF2 patients, who require frequent surveillance.
引用
收藏
页码:E263 / E267
页数:5
相关论文
共 21 条
  • [1] Current status and recommendations for imaging in neurofibromatosis type 1, neurofibromatosis type 2, and schwannomatosis
    Ahlawat, Shivani
    Blakeley, Jaishri O.
    Langmead, Shannon
    Belzberg, Allan J.
    Fayad, Laura M.
    [J]. SKELETAL RADIOLOGY, 2020, 49 (02) : 199 - 219
  • [2] Allison Jerry, 2015, AJR Am J Roentgenol, V205, pW140, DOI 10.2214/AJR.14.14173
  • [3] Auditory Brainstem Implant Array Position Varies Widely Among Adult and Pediatric Patients and Is Associated With Perception
    Barber, Samuel R.
    Kozin, Elliott D.
    Remenschneider, Aaron K.
    Puram, Sidharth V.
    Smith, Max
    Herrmann, Barbara S.
    Cunnane, Mary E.
    Brown, M. Christian
    Lee, Daniel J.
    [J]. EAR AND HEARING, 2017, 38 (06) : E343 - E351
  • [4] Magnetic Resonance Imaging With Cochlear Implant Magnet in Place: Safety and Imaging Quality
    Carlson, Matthew L.
    Neff, Brian A.
    Link, Michael J.
    Lane, John I.
    Watson, Robert E.
    McGee, Kiaran P.
    Bernstein, Matt A.
    Driscoll, Colin L. W.
    [J]. OTOLOGY & NEUROTOLOGY, 2015, 36 (06) : 965 - 971
  • [5] Cochlear&TRADE, MRI GUID PROF
  • [6] Complications in Auditory Brainstem Implant Surgery in Adults and Children
    Colletti, Vittorio
    Shannon, Robert V.
    Carner, Marco
    Veronese, Sheila
    Colletti, Liliana
    [J]. OTOLOGY & NEUROTOLOGY, 2010, 31 (04) : 558 - 564
  • [7] EDGERTON BJ, 1982, ANN OTO RHINOL LARYN, V91, P117
  • [8] Safety study of the Cochlear Nucleus® 24 Device with internal magnet in the 1.5 Tesla magnetic resonance imaging scanner
    Gubbels, SP
    McMenomey, SO
    [J]. LARYNGOSCOPE, 2006, 116 (06) : 865 - 871
  • [9] Conducting polymer electrodes for auditory brainstem implants
    Guex, Amelie A.
    Vachicouras, Nicolas
    Hight, Ariel Edward
    Brown, M. Christian
    Lee, Daniel J.
    Lacour, Stephanie P.
    [J]. JOURNAL OF MATERIALS CHEMISTRY B, 2015, 3 (25) : 5021 - 5027
  • [10] Heller JW, 1996, AM J OTOL, V17, P724