Analysis of the in-vivo failure of the adhesive interlayer for a DLC coated articulating metatarsophalangeal joint

被引:40
作者
Hauert, R. [1 ]
Thorwarth, G. [1 ]
Mueller, U. [1 ]
Stiefel, M. [1 ]
Falub, C. V. [1 ]
Thorwarth, K. [1 ]
Joyce, T. J. [2 ]
机构
[1] Empa, Swiss Fed Labs Mat Sci & Technol, CH-8600 Dubendorf, Switzerland
[2] Newcastle Univ, Sch Mech & Syst Engn, Newcastle Upon Tyne NE1 7RU, Tyne & Wear, England
关键词
Adhesion; Interface; Silicon; Corrosion; DLC; Joint replacement; DIAMOND-LIKE-CARBON; TRIBOLOGICAL BEHAVIOR; OXIDE; BIOSTABILITY; SILICON;
D O I
10.1016/j.diamond.2012.02.001
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Unwanted physiological reactions, caused by particle release in articulating joints, are still a major concern in total hip arthroplasty. Diamond-like carbon coatings (DLC) show very low wear values in many technical applications. Attempts to transfer these outstanding tribological properties to in-vivo load bearing articulating joints turned out to be extremely problematic. A DLC coated metallic metatarsophalangeal (MTP) prosthesis, with both articulating surfaces coated, was explanted after approximately 4 years. The DLC-coating was removed from the entire face of the phalangeal component and from most of the face of the metatarsal component. It appeared that the coating was scratched and then delaminated parallel to the scratches. To retrospectively investigate the reason for the in-vivo failure, the explanted prosthesis was analyzed by X-ray photoelectron spectroscopy (XPS) depth profiling and cross sectional scanning electron microscope (SEM) analysis. The coating consisted of an approximately 180 nm thick DLC film and a circa 120 nm thick silicon based adhesion promoting interlayer on a CoCrMo base material. SEM observation of transversal cuts showed that the delamination of the DLC-coatings was mainly caused by a slow in-vivo corrosion of the whole Si interlayer due to crevice corrosion. Additionally, a weak mechanical bond between the CoCrMo substrate and the Si-based interlayer further facilitated the access of body liquid below the Si interlayer. The instability of Si toward crevice corrosion was also demonstrated by a separate crevice experiment using silicon single crystals. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:34 / 39
页数:6
相关论文
共 18 条
[1]  
[Anonymous], 2008, TRIBOL DIAM LIKE CAR, DOI DOI 10.1007/978-0-387-49891-1_20
[2]  
Canham LT, 1999, ADV MATER, V11, P1505, DOI 10.1002/(SICI)1521-4095(199912)11:18<1505::AID-ADMA1505>3.0.CO
[3]  
2-C
[4]   THE EFFECT OF EXPOSURE TO BIOLOGICAL-FLUIDS ON THE SPALLATION RESISTANCE OF DIAMOND-LIKE CARBON COATINGS ON METALLIC SUBSTRATES [J].
CHANDRA, L ;
ALLEN, M ;
BUTTER, R ;
RUSHTON, N ;
LETTINGTON, AH ;
CLYNE, TW .
JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 1995, 6 (10) :581-589
[5]   In vitro studies of the adhesion of diamond-like carbon thin films on CoCrMo biomedical implant alloy [J].
Falub, C. V. ;
Mueller, U. ;
Thorwarth, G. ;
Parlinska-Wojtan, M. ;
Voisard, C. ;
Hauert, R. .
ACTA MATERIALIA, 2011, 59 (11) :4678-4689
[6]   Biostability of micro-photodiode arrays for subretinal implantation [J].
Hämmerle, H ;
Kobuch, K ;
Kohler, K ;
Nisch, W ;
Sachs, H ;
Stelzle, M .
BIOMATERIALS, 2002, 23 (03) :797-804
[7]   An overview on the tribological behavior of diamond-like carbon in technical and medical applications [J].
Hauert, R .
TRIBOLOGY INTERNATIONAL, 2004, 37 (11-12) :991-1003
[8]   Surface analysis and bioreactions of F and Si containing a-C:H [J].
Hauert, R ;
Muller, U ;
Francz, G ;
Birchler, F ;
Schroeder, A ;
Mayer, J ;
Wintermantel, E .
THIN SOLID FILMS, 1997, 308 :191-194
[9]  
Hauert R., ACTA BIOMATER UNPUB
[10]   Examination of failed ex vivo metal-on-metal metatarsophalangeal prosthesis and comparison with theoretically determined lubrication regimes [J].
Joyce, T. J. .
WEAR, 2007, 263 :1050-1054