Comparison of two point-of-care international normalized ratio devices and laboratory method

被引:5
作者
Vazquez, Sara R. [1 ]
Fleming, Ryan P. [1 ]
Johnson, Stacy A. [2 ]
机构
[1] Univ Utah, Hlth Thrombosis Serv, Dept Pharm Serv, 50 N Med Dr,Room 1R211, Salt Lake City, UT 84132 USA
[2] Univ Utah, Sch Med, Dept Gen Internal Med, Salt Lake City, UT 84112 USA
关键词
anticoagulation; Coag-Sense; CoaguChek; correction factor; international normalized ratio; laboratory; Neoplastine; point-of-care; Stago; warfarin; ORAL ANTICOAGULANT-THERAPY; VITAMIN-K-ANTAGONISTS; ANTIPHOSPHOLIPID SYNDROME; SELF-MANAGEMENT; COAGULOMETER; INSTRUMENTS; COLLEGE; XS;
D O I
10.1097/MBC.0000000000000633
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A number of factors contribute to interlaboratory variation of international normalized ratio (INR) results. This variability can lead to differences in clinical decision-making. This prospective cohort study evaluated the accuracy, precision, and clinical decision-making variability of two point-of-care (POC) devices (CoaguChek XS and the Coag-Sense) to a clinical laboratory method of INR measurement as the 'reference' and by using the laboratory's accuracy standard (+/- 25% of the reference INR). Study subjects included adults taking warfarin undergoing a POC INR with each device and a laboratory INR on the same day and evaluated differences in clinical decision-making using a warfarin nomogram. A correction factor was derived for each POC method to estimate the corresponding laboratory INR for POC INRs more than 3.0. A total of 100 patient encounters (76 unique patients) included INR testing by all three methods. Both POC devices demonstrated excellent precision across the entire range (r=0.98), but poor accuracy relative to the laboratory standard for POC INRs greater than 3.0 (accuracy CoaguChek 19% and CoagSense 35%). A correction factor applied to POC INRs greater than 3.0 improved accuracy to 99% for both devices. Applying the correction factor also significantly reduced differences in clinical decision-making (49-28% for CoaguChek and 56-40% for Coag-Sense, P<0.001). Both POC devices demonstrated poor accuracy for INRs over 3.0. A device-specific, institution-specific correction factor applied to POC INRs greater than 3.0 can provide better accuracy of INR measurement and can significantly reduce variability in warfarin dosing decisions. More study is needed to assess impact on clinical adverse events. Copyright (C) 2017 Wolters Kluwer Health, Inc. All rights reserved.
引用
收藏
页码:534 / 539
页数:6
相关论文
共 24 条
[1]   Oral Anticoagulant Therapy Antithrombotic Therapy and Prevention of Thrombosis, 9th ed: American College of Chest Physicians Evidence-Based Clinical Practice Guidelines [J].
Ageno, Walter ;
Gallus, Alexander S. ;
Wittkowsky, Ann ;
Crowther, Mark ;
Hylek, Elaine M. ;
Palareti, Gualtiero .
CHEST, 2012, 141 (02) :E44S-E88S
[2]  
[Anonymous], 2016, FDA REGULATORY OVERS
[3]   Evolving models of warfarin management: Anticoagulation clinics, patient self-monitoring, and patient self-management [J].
Ansell, JE ;
Hughes, R .
AMERICAN HEART JOURNAL, 1996, 132 (05) :1095-1100
[4]   Antiphospholipid syndrome patients: The performance of Coagucheck XS in the monitoring of Vitamin K-Antagonists [J].
Barcellona, Doris ;
Fenu, Lara ;
Vannini, Maria Luigia ;
Piras, Maria ;
Marongiu, Francesco .
THROMBOSIS RESEARCH, 2012, 129 (04) :E168-E170
[5]   Effects of the lupus anticoagulant on the control of oral anticoagulant treatment [J].
Carrasco, C ;
Campbell, B ;
Shiach, C .
BRITISH JOURNAL OF HAEMATOLOGY, 2004, 124 (04) :562-563
[6]  
CoaguSense Inc, 2012, COAG SENS
[7]   Manufacturer failed to disclose faulty device in rivaroxaban trial [J].
Cohen, Deborah .
BMJ-BRITISH MEDICAL JOURNAL, 2016, 354
[8]   Comparison of International Normalized Ratios provided by two point-of-care devices and laboratory-based venipuncture in a pharmacist-managed anticoagulation clinic [J].
Donaldson, Mark ;
Sullivan, John ;
Norbeck, Amber .
AMERICAN JOURNAL OF HEALTH-SYSTEM PHARMACY, 2010, 67 (19) :1616-1622
[9]  
Fairweather RB, 1998, ARCH PATHOL LAB MED, V122, P768
[10]   Clinical impact of interlaboratory variation in International Normalized Ratio determinations [J].
Finck, KM ;
Doetkott, C ;
Miller, DR .
AMERICAN JOURNAL OF HEALTH-SYSTEM PHARMACY, 2001, 58 (08) :684-688