breast augmentation;
breast implants;
implant rupture;
silicone gel;
adverse effects;
D O I:
10.1097/01.sap.0000164470.76432.4f
中图分类号:
R61 [外科手术学];
学科分类号:
摘要:
The objective was to evaluate the usefulness of clinical examination in the evaluation of breast-implant integrity, using the diagnosis at magnetic resonance imaging (MRI) as the "gold standard." Fifty-five women with 109 implants underwent a breast examination either just before or shortly after an M RI examination. Twenty-four of 109 implants were clinically diagnosed with possible rupture or rupture. Eighteen of the 24 implants were ruptured according to the MRI examination (75%). Eighty-five implants were clinically classified as intact, and 43 of these were actually ruptured at MRI (51%). The sensitivity of the clinical examination for diagnosing rupture was thus 30% and the specificity 88%. The positive predictive value of a clinical diagnosis of rupture was 75%, and the negative predictive value was 49%. In this study, we found that when a clinical examination is used as the sole diagnostic tool to identify implant rupture, neither the sensitivity nor the specificity is acceptable.