Design and evaluation of perception reconstruction with sensor system for artificial anal sphincter based on vector similarity

被引:4
作者
Han, Ding [1 ,2 ]
Yan, Guozheng [1 ,2 ]
Hua, Fangfang [1 ,2 ]
Wang, Lichao [1 ,2 ]
Wang, Zhiwu [1 ,2 ]
Jiang, Pingping [1 ,2 ]
Kuang, Shuai [1 ,2 ]
机构
[1] Shanghai Jiao Tong Univ, Sch Elect Informat & Elect Engn, Shanghai 200240, Peoples R China
[2] Shanghai Engn Res Ctr Intelligent Addict Treatmen, Shanghai, Peoples R China
基金
中国国家自然科学基金;
关键词
artificial anal sphincter; perception reconstruction; self-packaging strain gauge sensor; vector similarity; weight optimization; FECAL INCONTINENCE; AUGMENTATION; IMPLANTATION;
D O I
10.1111/aor.14369
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background Artificial organs are playing an increasingly important role in medical field. Artificial anal sphincter, as an example, is a widely used medical device for fecal incontinence. Though it could help patients maintain continence, the issue of perception reconstruction is still unsolved, which means that patients cannot control defecation as desired. Methods In this paper, a novel sensor system based on vector similarity has been analyzed and evaluated. The sensor system is a self-packaged strain gauge sensor with a small size 14.5 mm x 6 mm, especially for medical device implanted in body to reconstruct perception function. In order to overcome the disadvantages of single threshold judgment, a more accurate and reliable judgment standard for content pressure detection in rectum is developed by using vector similarity method. Therefore, in the novel sensor system, standard multi-parameters are comprehensive considered. Furthermore, weight-optimization of sensors are investigated since each sensor has played a different role in detection. Finally, sensor calibration, in vitro and in vivo experiments are established. Results In sensor calibration, the result R-2 > 0.99 presents the strain gauge pressure sensor has a good linearity. After that, a series of in vitro experiments have been conducted. The optimized sensor system shows a high accuracy rate in perception reconstruction, which is 87.5% on early warning and 95% on the alarm. In the following in vivo experiments. The results indicate that the average accuracy of the proposed perception reconstruction module has increased by 17.7%, compared with that without optimized sensor system. Conclusions All of these have proven that the novel perception reconstruction module with sensor system based on vector similarity is more acceptable and reliable.
引用
收藏
页码:2391 / 2399
页数:9
相关论文
共 23 条
[1]   IMPLANTATION OF ARTIFICIAL SPHINCTER FOR ANAL INCONTINENCE - REPORT OF 5 CASES [J].
CHRISTIANSEN, J ;
LORENTZEN, M .
DISEASES OF THE COLON & RECTUM, 1989, 32 (05) :432-436
[2]   A Wearable All-Gel Multimodal Cutaneous Sensor Enabling Simultaneous Single-Site Monitoring of Cardiac-Related Biophysical Signals [J].
Chun, Kyoung-Yong ;
Seo, Seunghwan ;
Han, Chang-Soo .
ADVANCED MATERIALS, 2022, 34 (16)
[3]  
Fan Z, 2018, CHIN CONTR CONF, P8548, DOI 10.23919/ChiCC.2018.8484127
[4]   Artificial Muscle Devices: Innovations and Prospects for Fecal Incontinence Treatment [J].
Fattorini, Elisa ;
Brusa, Tobia ;
Gingert, Christian ;
Hieber, Simone E. ;
Leung, Vanessa ;
Osmani, Bekim ;
Dominietto, Marco D. ;
Buechler, Philippe ;
Hetzer, Franc ;
Mueller, Bert .
ANNALS OF BIOMEDICAL ENGINEERING, 2016, 44 (05) :1355-1369
[5]   Outcome after implantation of a novel prosthetic anal sphincter in humans [J].
Finlay, IG ;
Richardson, W ;
Hajivassiliou, CA .
BRITISH JOURNAL OF SURGERY, 2004, 91 (11) :1485-1492
[6]   Gatekeeper Improves Voluntary Contractility in Patients With Fecal Incontinence [J].
Grossi, Ugo ;
De Simone, Veronica ;
Parello, Angelo ;
Litta, Francesco ;
Donisi, Lorenza ;
Luca Di Tanna, Gian ;
Goglia, Marta ;
Ratto, Carlo .
SURGICAL INNOVATION, 2019, 26 (03) :321-327
[7]   An artificial anal sphincter based on a novel clamping mechanism: Design, analysis, and testing [J].
Han, Ding ;
Yan, Guozheng ;
Wang, Zhiwu ;
Jiang, Pingping ;
Liu, Dasheng ;
Zhao, Kai ;
Ma, Jin .
ARTIFICIAL ORGANS, 2021, 45 (08) :E293-E303
[8]   Preliminary Study of a Novel Puborectalis-Like Artificial Anal Sphincter [J].
Jin, Wentian ;
Yan, Guozheng ;
Wu, Hao ;
Lu, Shan ;
Zhou, Zerun .
ARTIFICIAL ORGANS, 2017, 41 (09) :845-851
[9]  
Ke Lei, 2014, Journal of Shanghai Jiaotong University, V48, P193
[10]   Robotic Localization Based on Planar Cable Robot and Hall Sensor Array Applied to Magnetic Capsule Endoscope [J].
Kim, Min-Cheol ;
Kim, Eui-Sun ;
Park, Jong-Oh ;
Choi, Eunpyo ;
Kim, Chang-Sei .
SENSORS, 2020, 20 (20) :1-18