The Orbitrap: a new mass spectrometer

被引:825
作者
Hu, QZ
Noll, RJ
Li, HY
Makarov, A
Hardman, M
Cooks, RG
机构
[1] Purdue Univ, Dept Chem, W Lafayette, IN 47907 USA
[2] Thermo Electron Bremem, D-28199 Bremen, Germany
[3] Thermo Electron San Jose, San Jose, CA 95134 USA
来源
JOURNAL OF MASS SPECTROMETRY | 2005年 / 40卷 / 04期
关键词
high resolution; Fourier transform mass spectrometry (FT-MS); proteomics; metabotomics; novel mass analyzers; electrospray ionization; serine octamer;
D O I
10.1002/jms.856
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Research areas such as proteomics and metabolomics are driving the demand for mass spectrometers that have high performance but modest power requirements, size, and cost. This paper describes such an instrument, the Orbitrap, based on a new type of mass analyzer invented by Makarov. The Orbitrap operates by radially trapping ions about a central spindle electrode. An outer barrel-like electrode is coaxial with the inner spindlelike electrode and mass/charge values are measured from the frequency of harmonic ion oscillations, along the axis of the electric field, undergone by the orbitally trapped ions. This axial frequency is independent of the energy and spatial spread of the ions. Ion frequencies are measured non-destructively by acquisition of time-domain image current transients, with subsequent fast Fourier transforms (FFTs) being used to obtain the mass spectra. In addition to describing the Orbitrap mass analyzer, this paper also describes a complete Orbitrap-based mass spectrometer, equipped with an electrospray ionization source (ESI). Ions are transferred from the ESI source through three stages of differential pumping using RF guide quadrupoles. The third quadrupole, pressurized to less than 10(-3) Torr with collision gas, acts as an ion accumulator; ion/neutral collisions slow the ions and cause them to pool in an axial potential well at the end of the quadrupole. Ion bunches are injected from this pool into the Orbitrap analyzer for mass analysis. The ion injection process is described in a simplified way, including a description of electrodynamic squeezing, field compensation for the effects of the ion injection slit, and criteria for orbital stability. Features of the Orbitrap at its present stage of development include high mass resolution (up to 150000), large space charge capacity, high mass accuracy (2-5 ppm), a mass/charge range of at least 6000, and dynamic range greater than 10.(3) Applications based on electrospray ionization are described, including characterization of transition-metal complexes, oligosaccharides, peptides, and proteins. Use is also made of the high-resolution capabilities of the Orbitrap to confirm the presence of metaclusters of serine octamers in ESI mass spectra and to perform H/D exchange experiments on these ions in the storage quadrupole. Copyright (c) 2005 John Wiley & Sons, Ltd.
引用
收藏
页码:430 / 443
页数:14
相关论文
共 83 条
[1]  
Amster IJ, 1996, J MASS SPECTROM, V31, P1325
[2]   Monitoring structural changes of proteins in an ion trap over ∼10-200 ms:: Unfolding transitions in cytochrome c ions [J].
Badman, ER ;
Hoaglund-Hyzer, CS ;
Clemmer, DE .
ANALYTICAL CHEMISTRY, 2001, 73 (24) :6000-6007
[3]  
BLUEMEL R, 1995, PHYS REV A, V51, pR30
[4]   Intercomparison study on accurate mass measurement of small molecules in mass spectrometry [J].
Bristow, AWT ;
Webb, KS .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2003, 14 (10) :1086-1098
[5]   KINGDON CAGE AS MOLECULAR BEAM DETECTOR [J].
BROOKS, PR ;
HERSCHBACH, DR .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1964, 35 (11) :1528-&
[6]   STRUCTURAL AND ENERGETIC CONSTRAINTS ON GAS-PHASE HYDROGEN-DEUTERIUM EXCHANGE-REACTIONS OF PROTONATED PEPTIDES WITH D2O, CD3OD, CD3CO2D, AND ND3 [J].
CAMPBELL, S ;
RODGERS, MT ;
MARZLUFF, EM ;
BEAUCHAMP, JL .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1994, 116 (21) :9765-9766
[7]   Deuterium exchange reactions as a probe of biomolecule structure. Fundamental studies of cas phase H/D exchange reactions of protonated glycine oligomers with D2O, CD3OD, CD3CO2D, and ND3 [J].
Campbell, S ;
Rodgers, MT ;
Marzluff, EM ;
Beauchamp, JL .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1995, 117 (51) :12840-12854
[8]   Rapid bioanalysis of vancomycin in serum and urine by high-performance liquid chromatography tandem mass spectrometry using on-line sample extraction and parallel analytical columns [J].
Cass, RT ;
Villa, JS ;
Karr, DE ;
Schmidt, DE .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2001, 15 (06) :406-412
[9]   An interface with a linear quadrupole ion guide for an electrospray-ion trap mass spectrometer system [J].
Cha, BC ;
Blades, M ;
Douglas, DJ .
ANALYTICAL CHEMISTRY, 2000, 72 (22) :5647-5654
[10]   Ketalization of phosphonium ions by 1,4-dioxane: Selective detection of the chemical warfare agent simulant DMMP in mixtures using ion/molecule reactions [J].
Chen, H ;
Zheng, XB ;
Cooks, RG .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2003, 14 (03) :182-188