Evaluation of the reproducibility of the KT-1000 arthrometer

被引:42
作者
Sernert, N
Kartus, J
Köhler, K
Ejerhed, L
Karlsson, J
机构
[1] Norra Alvsborg Uddevalla Cty Hosp, Dept Orthopaed, Uddevalla, Sweden
[2] Sahlgrens Univ Hosp, Dept Orthopaed, Gothenburg, Sweden
关键词
anterior knee laxity; anterior cruciate ligament rupture; KT-1000; arthrometer; measurement; reproducibility;
D O I
10.1034/j.1600-0838.2001.011002120.x
中图分类号
G8 [体育];
学科分类号
04 ; 0403 ;
摘要
The aim of the study was to examine whether the KT-1000 arthrometer was reliable when it came to distinguishing between a group of patients,vith a chronic anterior cruciate ligament (ACL) rupture and a group of patients without an ACL rupture, and to examine the reproducibility of the examination between two experienced examiners. The aim was also to examine whether the KT-1000 measurements were dependent on whether the patients were awake or under anaesthesia. The study comprised 40 patients: Group a consisted of 20 patients who had a chronic unilateral ACL rupture and Group B consisted of 20 patients who were scheduled for arthroscopy due to knee problems other than an ACL rupture. The KT-1000 examination was performed before surgery by two experienced physiotherapists (PT I and PT II). PT II subsequently performed a retest of the patients under anaesthesia. The mean anterior side-to-side laxity difference between PT I and PT II was 0.2 mm in Group A and 1.8 mm in Group B (n.s., P=0.03). The anterior side-to-side measurements of knee laxity revealed significant differences between Group A and Group B, independent of who the measurements were made by when the patients were awake (PT I P=0.011, PT II P=0.001). However, no significant difference (P=0.063) was found when the patients were under anaesthesia. The interclass correlation coefficient (ICC) between PT I and PT II in Group A was 0.55 (P=0.005) for the anterior side-to-side laxity, while it was 0.60 (P= 0.002) in Group B. There were no significant differences within Group A or Group B between the measurements made when people were awake compared with those under anaesthesia. The conclusions of the study were that the KT-1000 arthrometer was able to distinguish a group of patients with an ACL rupture from a group without one. The reproducibility of the KT-1000 measurements of anterior knee laxity between two experienced examiners was considered as fair. Furthermore, the measurements were not dependent on whether the patients were awake or under anaesthesia.
引用
收藏
页码:120 / 125
页数:6
相关论文
共 18 条
  • [1] INSTRUMENTED EVALUATION OF KNEE LAXITY - A COMPARISON OF 5 ARTHROMETERS
    ANDERSON, AF
    SNYDER, RB
    FEDERSPIEL, CF
    LIPSCOMB, AB
    GRANA, WA
    [J]. AMERICAN JOURNAL OF SPORTS MEDICINE, 1992, 20 (02) : 135 - 140
  • [2] PREOPERATIVE INSTRUMENTED TESTING OF ANTERIOR AND POSTERIOR KNEE LAXITY
    ANDERSON, AF
    LIPSCOMB, AB
    [J]. AMERICAN JOURNAL OF SPORTS MEDICINE, 1989, 17 (03) : 387 - 392
  • [3] ARTHROMETRIC EVALUATION OF KNEES THAT HAVE A TORN ANTERIOR CRUCIATE LIGAMENT
    BACH, BR
    WARREN, RF
    FLYNN, WM
    KROLL, M
    WICKIEWIECZ, TL
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (09) : 1299 - 1306
  • [4] INFLUENCE OF EXAMINER EXPERIENCE AND GENDER ON INTERRATER RELIABILITY OF KT-1000 ARTHROMETER MEASUREMENTS
    BALLANTYNE, BT
    FRENCH, AK
    HEIMSOTH, SL
    KACHINGWE, AF
    LEE, JB
    SODERBERG, GL
    [J]. PHYSICAL THERAPY, 1995, 75 (10): : 898 - 906
  • [5] INSTRUMENTED MEASUREMENT OF ANTERIOR KNEE LAXITY IN PATIENTS WITH ACUTE ANTERIOR CRUCIATE LIGAMENT DISRUPTION
    DANIEL, DM
    STONE, ML
    SACHS, R
    MALCOM, L
    [J]. AMERICAN JOURNAL OF SPORTS MEDICINE, 1985, 13 (06) : 401 - 407
  • [6] IS THE KT1000 KNEE LIGAMENT ARTHROMETER RELIABLE
    FORSTER, IW
    WARRENSMITH, CD
    TEW, M
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1989, 71 (05): : 843 - 847
  • [7] Graham G P, 1991, Br J Sports Med, V25, P96
  • [8] RELIABILITY OF MEASURING ANTERIOR LAXITY OF THE KNEE-JOINT USING A KNEE LIGAMENT ARTHROMETER
    HANTEN, WP
    PACE, MB
    [J]. PHYSICAL THERAPY, 1987, 67 (03): : 357 - 359
  • [9] INSTRUMENTED LACHMAN TESTS FOR THE EVALUATION OF ANTERIOR LAXITY AFTER RECONSTRUCTION OF THE ANTERIOR CRUCIATE LIGAMENT
    HARTER, RA
    OSTERNIG, LR
    SINGER, KM
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1989, 71A (07) : 975 - 983
  • [10] KT-1000 ARTHROMETER - CONSCIOUS AND UNCONSCIOUS TEST-RESULTS USING 15, 20, AND 30 POUNDS OF FORCE
    HIGHGENBOTEN, CL
    JACKSON, AW
    JANSSON, KA
    MESKE, NB