Multiparameter Correction Equation for Goldmann Applanation Tonometry

被引:37
作者
Elsheikh, Ahmed [2 ]
Alhasso, Daad [2 ]
Gunvant, Pinakin [1 ,3 ,4 ]
Garway-Heath, David [5 ,6 ,7 ]
机构
[1] So Coll Optometry, Dept Res, Memphis, TN 38117 USA
[2] Univ Dundee, Div Civil Engn, Dundee, Scotland
[3] Univ Louisville, Dept Psychol & Brain Sci, Louisville, KY 40292 USA
[4] Univ Memphis, Dept Elect & Comp Engn, Memphis, TN 38152 USA
[5] Moorfields Eye Hosp NHS Fdn Trust, NIHR Biomed Res Ctr Ophthalmol, London, England
[6] UCL, Inst Ophthalmol, London, England
[7] IRCCS, GB Bietti Fdn Res Ophthalmol, Rome, Italy
基金
美国国家卫生研究院;
关键词
tonometry; intraocular pressure; cornea; numerical modeling; CENTRAL CORNEAL THICKNESS; INTRAOCULAR-PRESSURE MEASUREMENTS; DYNAMIC CONTOUR TONOMETRY; OPEN-ANGLE GLAUCOMA; BIOMECHANICAL PROPERTIES; OCULAR HYPERTENSION; KERATOCONUS CORNEAS; DIURNAL-VARIATION; TONO-PEN; CURVATURE;
D O I
10.1097/OPX.0b013e3181fc3453
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose. To develop a correction factor to improve the accuracy of intraocular pressure (IOP) measurements made by the Goldmann applanation tonometer (GAT), which considers the combined effects of variations in central corneal thickness (CCT), central anterior curvature (R), age, and the IOP level itself. Methods. Nonlinear numerical simulations based on the finite element method were used to represent corneal behavior under the effect of IOP and external tonometric pressure. The simulations considered various biomechanical corneal properties including the cornea's nonuniform thickness, elliptical topography, weak stromal interlamellar cohesion, low epithelial and endothelial stiffness, and hyperelastic and hysteretic material behavior. The simulations were used to model the GAT procedure on corneas to obtain a correction equation based on the values of CCT, R, age, and IOP measured using GAT (IOPG). The efficiency of the equation in reducing the effects of corneal parameters on IOPG measurements was also assessed using an independent clinical database. Results. The individual effects of variations in CCT, R, and age were estimated at 1.66 mm Hg/100 mu of CCT, 0.89 mm Hg/1 mm of R, and 0.12 mm Hg/decade of age. The correction equation reduced the association between clinical IOP measurements and corneal parameters with r(2) reducing from 11.8 to 0.02%. Conclusions. The GAT correction factor can consider the combined effect of variations in corneal thickness, curvature, age, and IOP. The factor could significantly reduce the reliance of IOPG measurements on corneal stiffness parameters. (Optom Vis Sci 2011; 88: E102-E112)
引用
收藏
页码:E102 / E112
页数:11
相关论文
共 56 条
[1]   BIOMECHANICAL PROPERTIES OF KERATOCONUS AND NORMAL CORNEAS [J].
ANDREASSEN, TT ;
SIMONSEN, AH ;
OXLUND, H .
EXPERIMENTAL EYE RESEARCH, 1980, 31 (04) :435-441
[2]  
Bhan A, 2002, INVEST OPHTH VIS SCI, V43, P1389
[3]   Constitutive laws for biomechanical modeling of refractive surgery [J].
Bryant, MR ;
McDonnell, PJ .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1996, 118 (04) :473-481
[4]  
BRYANT MR, 1994, INVEST OPHTH VIS SCI, V35, P3022
[5]   Assessment of true intraocular pressure: The gap between theory and practical data [J].
Chihara, Etsuo .
SURVEY OF OPHTHALMOLOGY, 2008, 53 (03) :203-218
[6]  
Coster D., 2002, CORNEA
[7]   Radius and asphericity of the posterior corneal surface determined by corrected Scheimpflug photography [J].
Dubbelman, M ;
Weeber, HA ;
van der Heijde, RGL ;
Völker-Dieben, HJ .
ACTA OPHTHALMOLOGICA SCANDINAVICA, 2002, 80 (04) :379-383
[8]   Biomechanics and wound healing in the cornea [J].
Dupps, William J., Jr. ;
Wilson, Steven E. .
EXPERIMENTAL EYE RESEARCH, 2006, 83 (04) :709-720
[9]  
EHLERS N, 1975, ACTA OPHTHALMOL, V53, P34
[10]   Numerical modelling of corneal biomechanical behaviour [J].
Elsheikh, Ahmed ;
Wang, Defu .
Computer Methods in Biomechanics and Biomedical Engineering, 2007, 10 (02) :85-95