A technical solution to avoid partial scan artifacts in cardiac MDCT

被引:17
作者
Primak, A. N. [1 ]
Dong, Y. [2 ]
Dzyubak, O. P. [1 ]
Jorgensen, S. M. [2 ]
McCollough, C. H. [1 ]
Ritman, E. L. [2 ]
机构
[1] Mayo Clin, Coll Med, Dept Radiol, CT Clin Innovat Ctr, Rochester, MN 55905 USA
[2] Mayo Clin, Coll Med, Dept Physiol & Biomed Engn, Rochester, MN 55905 USA
关键词
multidetector row CT; CT perfusion; partial scan reconstruction; cardiac CT; CT artifacts;
D O I
10.1118/1.2805476
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Quantitative evaluation of cardiac image data obtained using multidetector row computed tomography (CT) is compromised by partial scan reconstructions, which improve the temporal resolution but significantly increase image-to-image CT number variations for a fixed region of interest compared to full reconstruction images. The feasibility of a new approach to solve this problem is assessed. An anthropomorphic cardiac phantom and an anesthetized pig were scanned on a dual-source CT scanner using both full and partial scan acquisition modes under different conditions. Additional scans were conducted with the electrocardiogram (ECG) signal being in synchrony with the gantry rotation. In the animal study, a simple x-ray detector was used to generate a signal once per gantry rotation. This signal was then used to pace the pig's heart. Phantom studies demonstrated that partial scan artifacts are strongly dependent on the rotational symmetry of angular projections, which is determined by the object shape and composition and its position with respect to the isocenter. The degree of partial scan artifacts also depends on the location of the region of interest with respect to highly attenuating materials (bones, iodine, etc.) within the object. Single-source partial scan images (165 ms temporal resolution) were significantly less affected by partial scan artifacts compared to dual-source partial scan images (82 ms temporal resolution). When the ECG signal was in synchrony with the gantry rotation, the same cardiac phase always corresponded to the same positions of the x-ray tube(s) and, hence, the same scattering and beam hardening geometry. As a result, the range of image-to-image CT number variations for partial scan reconstruction images acquired in synchronized mode was decreased to that achieved using full reconstruction image data. The success of the new approach, which synchronizes the ECG signal with the position of the x-ray tube(s), was demonstrated both in the phantom and animal experiments. (C) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:4726 / 4737
页数:12
相关论文
共 35 条
  • [1] Detection of calcified and noncalcified coronary atherosclerotic plaque by contrast-enhanced, submillimeter multidetector spiral computed tomography - A segment-based comparison with intravascular ultrasound
    Achenbach, S
    Moselewski, F
    Ropers, D
    Ferencik, M
    Hoffmann, U
    MacNeill, B
    Pohle, K
    Baum, U
    Anders, K
    Jang, I
    Daniel, WG
    Brady, TJ
    [J]. CIRCULATION, 2004, 109 (01) : 14 - 17
  • [2] Achenbach S, 2007, HERZ, V32, P97, DOI 10.1007/s00059-007-2961-8
  • [3] Validation of minimally invasive measurement of myocardial perfusion using electron beam computed tomography and application in human volunteers
    Bell, MR
    Lerman, LO
    Rumberger, JA
    [J]. HEART, 1999, 81 (06) : 628 - 635
  • [4] PROPOSED DYNAMIC CARDIAC 3-D DENSITOMETER FOR EARLY DETECTION AND EVALUATION OF HEART-DISEASE
    BOYD, DP
    GOULD, RG
    QUINN, JR
    SPARKS, R
    STANLEY, JH
    HERRMANNSFELDT, WB
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1979, 26 (02) : 2724 - 2726
  • [5] Usefulness of electron beam computed tomography scanning for distinguishing ischemic from nonischemic cardiomyopathy
    Budoff, MJ
    Shavelle, DM
    Lamont, DH
    Kim, HT
    Akinwale, P
    Kennedy, JM
    Brundage, BH
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 32 (05) : 1173 - 1178
  • [6] Accuracy of 16-slice multi-detector CT to quantify the degree of coronary artery stenosis: Assessment of cross-sectional and longitudinal vessel reconstructions
    Cury, RC
    Ferencik, M
    Achenbach, S
    Pomerantsev, E
    Nieman, K
    Moselewski, F
    Abbara, S
    Jang, IK
    Brady, TJ
    Hoffmann, U
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2006, 57 (03) : 345 - 350
  • [7] Comprehensive cardiac CT study: Evaluation of coronary arteries, left ventricular function, and myocardial perfusion - Is it possible?
    Cury, Ricardo C.
    Nieman, Koen
    Shapiro, Michael D.
    Nasir, Khurram
    Cury, Roberto C.
    Brady, Thomas J.
    [J]. JOURNAL OF NUCLEAR CARDIOLOGY, 2007, 14 (02) : 229 - 243
  • [8] Evaluation of porcine myocardial microvascular permeability and fractional vascular volume using 64-slice helical computed tomography (CT)
    Daghini, Elena
    Primak, Andrew N.
    Chade, Alejandro R.
    Zhu, Xiangyang
    Ritman, Erik L.
    McCollough, Cynthia H.
    Lerman, Lilach O.
    [J]. INVESTIGATIVE RADIOLOGY, 2007, 42 (05) : 274 - 282
  • [9] Assessment of myocardial microvascular function: New opportunities in fast computed tomography
    Daghini, Elena
    Lerman, Lilach O.
    [J]. TRENDS IN CARDIOVASCULAR MEDICINE, 2007, 17 (01) : 14 - 19
  • [10] Heart rate adaptive optimization of spatial and temporal resolution for electrocardiogram-gated multislice spiral CT of the heart
    Flohr, T
    Ohnesorge, B
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2001, 25 (06) : 907 - 923