Comparison of Two Software Packages for Perfusion Imaging: Ischemic Core and Penumbra Estimation and Patient Triage in Acute Ischemic Stroke

被引:6
|
作者
Zhou, Xiang [1 ]
Nan, Yashi [2 ]
Ju, Jieyang [3 ]
Zhou, Jingyu [2 ]
Xiao, Huanhui [2 ]
Wang, Silun [2 ]
机构
[1] Tongji Univ, Tongji Hosp, Sch Med, Dept Radiol, 389 Xincun Rd, Shanghai 200065, Peoples R China
[2] YIWEI Med Technol Co Ltd, Room 1001,MAI KE LONG Bldg, Shenzhen 518000, Peoples R China
[3] Nanjing Med Univ, Affiliated Hosp 2, 121 Jiangjiayuan Rd, Nanjing 210011, Peoples R China
基金
中国国家自然科学基金;
关键词
RealNow; RAPID; ischemic core; penumbra; acute ischemic stroke; FINAL INFARCT VOLUME; PREDICT; MAPS; THROMBECTOMY; THRESHOLDS; SELECTION; TIME;
D O I
10.3390/cells11162547
中图分类号
Q2 [细胞生物学];
学科分类号
071009 ; 090102 ;
摘要
Purpose: Automated postprocessing packages have been developed for managing acute ischemic stroke (AIS). These packages identify ischemic core and penumbra using either computed tomographic perfusion imaging (CTP) data or magnetic resonance imaging (MRI) data. Measurements of abnormal tissues and treatment decisions derived from different vendors can vary. The purpose of this study is to investigate the agreement of volumetric and decision-making outcomes derived from two software packages. Methods: A total of 594 AIS patients (174 underwent CTP and 420 underwent MRI) were included. Imaging data were accordingly postprocessed by two software packages: RAPID and RealNow. Volumetric outputs were compared between packages by performing intraclass correlation coefficient (ICC), Wilcoxon paired test and Bland-Altman analysis. Concordance of selecting patients eligible for mechanical thrombectomy (MT) was assessed based on neuroimaging criteria proposed in DEFUSE3. Results: In the group with CTP data, mean ischemic core volume (ICV)/penumbral volume (PV) was 14.9/81.1 mL via RAPID and 12.6/83.2 mL via RealNow. Meanwhile, in the MRI group, mean ICV/PV were 52.4/68.4 mL and 48.9/61.6 mL via RAPID and RealNow, respectively. Reliability, which was measured by ICC of ICV and PV in CTP and MRI groups, ranged from 0.87 to 0.99. The bias remained small between measurements (CTP ICV: 0.89 mL, CTP PV: -2 mL, MRI ICV: 3.5 mL and MRI PV: 6.8 mL). In comparison with CTP ICV with follow-up DWI, the ICC was 0.92 and 0.94 for RAPID and Realnow, respectively. The bias remained small between CTP ICV and follow-up DWI measurements (Rapid: -4.65 mL, RealNow: -3.65 mL). Wilcoxon paired test showed no significant difference between measurements. The results of patient triage were concordant in 159/174 cases (91%, ICC: 0.90) for CTP and 400/420 cases (95%, ICC: 0.93) for MRI. Conclusion: The CTP ICV derived from RealNow was more accurate than RAPID. The similarity in volumetric measurement between packages did not necessarily relate to equivalent patient triage. In this study, RealNow showed excellent agreement with RAPID in measuring ICV and PV as well as patient triage.
引用
收藏
页数:17
相关论文
共 50 条
  • [1] Agreement and Accuracy of Ischemic Core Volume Evaluated by Three CT Perfusion Software Packages in Acute Ischemic Stroke
    Liu, Qin Chen
    Jia, Zhen Yu
    Zhao, Lin Bo
    Cao, Yue Zhou
    Ma, Gao
    Shi, Hai Bin
    Liu, Sheng
    JOURNAL OF STROKE & CEREBROVASCULAR DISEASES, 2021, 30 (08)
  • [2] Comparison of three commonly used CT perfusion software packages in patients with acute ischemic stroke
    Koopman, Miou S.
    Berkhemer, Olvert A.
    Geuskens, Ralph R. E. G.
    Emmer, Bart J.
    van Walderveen, Marianne A. A.
    Jenniskens, Sjoerd F. M.
    van Zwam, Wim H.
    van Oostenbrugge, Robert J.
    van der Lugt, Aad
    Dippel, Diederik W. J.
    Beenen, Ludo F.
    Roos, Yvo B. W. E. M.
    Marquering, Henk A.
    Majoie, Charles B. L. M.
    JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2019, 11 (12) : 1249 - 1256
  • [3] Agreement of three CT perfusion software packages in patients with acute ischemic stroke: A comparison with RAPID
    Lu, Qingqing
    Fu, Junyan
    Lv, Kun
    Han, Yan
    Pan, Yuning
    Xu, Yiren
    Zhang, Jun
    Geng, Daoying
    EUROPEAN JOURNAL OF RADIOLOGY, 2022, 156
  • [4] Comparison of 3 CT Perfusion Software Packages in Estimation of Ischemic Lesions in Acute Ischemic Stroke Patients
    Li, Xiang
    Zeng, Chen
    Li, Yige
    Liu, Huan
    Liu, Ling
    Zeng, Wenbing
    Yang, Ran
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2023, 47 (03) : 500 - 506
  • [5] Does imaging of the ischemic penumbra have value in acute ischemic stroke with large vessel occlusion?
    Seners, Pierre
    Baron, Jean-Claude
    Olivot, Jean-Marc
    Albers, Gregory W.
    CURRENT OPINION IN NEUROLOGY, 2024, 37 (01) : 1 - 7
  • [6] Perfusion Imaging and Clinical Outcome in Acute Ischemic Stroke with Large Core
    Seners, Pierre
    Oppenheim, Catherine
    Turc, Guillaume
    Albucher, Jean-Francois
    Guenego, Adrien
    Raposo, Nicolas
    Christensen, Soren
    Calviere, Lionel
    Viguier, Alain
    Darcourt, Jean
    Januel, Anne-Christine
    Mlynash, Michael
    Sommet, Agnes
    Thalamas, Claire
    Sibon, Igor
    Rousseau, Vanessa
    Tourdias, Thomas
    Menegon, Patrice
    Bonneville, Fabrice
    Mazighi, Mikael
    Charron, Sylvain
    Legrand, Laurence
    Cognard, Christophe
    Albers, Gregory W.
    Baron, Jean-Claude
    Olivot, Jean-Marc
    ANNALS OF NEUROLOGY, 2021, 90 (03) : 417 - 427
  • [7] Collaterals are a major determinant of the core but not the penumbra volume in acute ischemic stroke
    Nannoni, Stefania
    Cereda, Carlo W.
    Sirimarco, Gaia
    Lambrou, Dimitris
    Strambo, Davide
    Eskandari, Ashraf
    Dunet, Vincent
    Wintermark, Max
    Michel, Patrik
    NEURORADIOLOGY, 2019, 61 (09) : 971 - 978
  • [8] Higher agreement in endovascular treatment decision-making than in parametric quantifications among automated CT perfusion software packages in acute ischemic stroke
    Perez-Pelegri, Manuel
    Biarnes, Carles
    Thio-Henestrosa, Santiago
    Remollo, Sebastia
    Gimeno, Alfredo
    Cuba, Victor
    Teceno, Mikel
    Marti-Navas, Marian
    Serena, Joaquin
    Pedraza, Salvador
    Ruiz-Constantino, Juan Sebastian
    Puig, Josep
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2021, 29 (05) : 823 - 834
  • [9] Improving the Accuracy of Perfusion Imaging in Acute Ischemic Stroke
    Albers, Gregory W.
    Fisher, Marc
    ANNALS OF NEUROLOGY, 2011, 70 (03) : 347 - 349
  • [10] Value of Perfusion Computed Tomography in Acute Ischemic Stroke: Diagnosis of Infarct Core and Penumbra
    Pan, Jiawei
    Zhang, Jun
    Huang, Weiyuan
    Cheng, Xin
    Ling, Yifeng
    Dong, Qiang
    Geng, Daoying
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2013, 37 (05) : 645 - 649