Thresholds determined using the monotic and dichotic multiple auditory steady-state response technique in normal-hearing subjects

被引:107
作者
Herdman, AT [1 ]
Stapells, DR [1 ]
机构
[1] Univ British Columbia, Sch Audiol & Speech Sci, Vancouver, BC V6T 1Z3, Canada
来源
SCANDINAVIAN AUDIOLOGY | 2001年 / 30卷 / 01期
关键词
auditory steady-state response; evoked potential audiometry; multiple simultaneous stimuli; threshold;
D O I
10.1080/010503901750069563
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Auditory steady-state responses (ASSRs) were elicited by presenting single or multiple, 77-105 Hz amplitude-modulated 0.5, 1, 2, and 4 kHz tones to one or both ears. Objectives of this study were to (i) replicate and extend previous multiple ASSR studies in a quiet double-walled sound booth, and (ii) discover differences (if any) between thresholds assessed in monotic and dichotic conditions, which ranged between 15 and 22dB SPL. The present study's behavioural and ASSR thresholds are 0-10 dB lower (better) than results of previous monotic studies. Further, there are no significant differences in ASSR thresholds between dichotic and monotic stimulus conditions. Therefore, dichotic multiple AM tone stimulation does not produce a change in the ASSR that affects threshold estimation in a clinically significant manner. Thus, at least for detecting normal hearing, the dichotic multiple ASSR technique is a feasible method for estimating hearing thresholds that would substantially reduce recording time compared to conventional single-stimulus techniques.
引用
收藏
页码:41 / 49
页数:9
相关论文
共 50 条