Glucose biosensing at carbon paste electrodes containing iron nanoparticles

被引:21
作者
Comba, Fausto N. [1 ]
Rubianes, Maria D. [1 ]
Herrasti, Pilar [2 ]
Rivas, Gustavo A. [1 ]
机构
[1] Univ Nacl Cordoba, Fac Ciencias Quim, Dept Quim Fis, INFIQC, RA-5000 Cordoba, Argentina
[2] Univ Autonoma Madrid, Fac Ciencias, Dept Quim Fis Aplicada, SSIC, E-28049 Madrid, Spain
关键词
Iron; Nanoparticles; Carbon paste electrode; Hydrogen peroxide; Glucose oxidase; Amperometric glucose biosensor; HYDROGEN-PEROXIDE; OXIDASE; IMMOBILIZATION; SENSORS; OXIDES; COPPER;
D O I
10.1016/j.snb.2010.06.020
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This work reports the advantages of incorporating a metal as common as iron in a carbon paste electrode containing glucose oxidase to develop highly selective, stable and sensitive amperometric glucose biosensors. The excellent catalytic activity of iron nanoparticles towards the reduction of hydrogen peroxide has made possible the quantification of glucose at very low potentials (-0.100V) avoiding, in this way, the interference of easily oxidizable compounds like ascorbic acid and uric acid without needing of redox mediators or permselective membranes. Linear relationship between current and glucose concentration was obtained up to 0.020M (3.6 g/L), with a detection limit of 0.2 mM, no interference of AA and UA, and very good correlation with the spectrophotometric methods when determining glucose in human blood serum. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:306 / 309
页数:4
相关论文
共 29 条
[1]   (Bio)sensors based on manganese dioxide-modified carbon substrates: retrospections, further improvements and applications [J].
Beyene, NW ;
Kotzian, P ;
Schachl, K ;
Alemu, H ;
Turkusic, E ;
Copra, A ;
Moderegger, H ;
Svancara, I ;
Vytras, K ;
Kalcher, K .
TALANTA, 2004, 64 (05) :1151-1159
[2]  
Celej MS, 1998, ELECTROANAL, V10, P771, DOI 10.1002/(SICI)1521-4109(199809)10:11<771::AID-ELAN771>3.0.CO
[3]  
2-P
[4]  
COMBA FN, ELECTROANAL IN PRESS
[5]   Amperometric biosensors based on carbon paste electrodes modified with nanostructured mixed-valence manganese oxides and glucose oxidase [J].
Cui, Xiaoli ;
Liu, Guodong ;
Lin, Yuehe .
NANOMEDICINE-NANOTECHNOLOGY BIOLOGY AND MEDICINE, 2005, 1 (02) :130-135
[6]  
Davis F, 2009, MICRO FUEL CELLS: PRINCIPLES AND APPLICATIONS, P243, DOI 10.1016/B978-0-12-374713-6.00006-8
[7]   Electrochemical glucose sensors and their applications in diabetes management [J].
Heller, Adam ;
Feldman, Ben .
CHEMICAL REVIEWS, 2008, 108 (07) :2482-2505
[8]   Carbon electrodes modified by nanoscopic iron(III) oxides to assemble chemical sensors for the hydrogen peroxide amperometric detection [J].
Hrbac, Jan ;
Halouzka, Vladimir ;
Zboril, Radek ;
Papadopoulos, Kyriakos ;
Triantis, Theodor .
ELECTROANALYSIS, 2007, 19 (17) :1850-1854
[9]   Super-washing does not leave single walled carbon nanotubes iron-free [J].
Jurkschat, Kerstin ;
Ji, Xiaobo ;
Crossley, Alison ;
Compton, Richard G. ;
Banks, Craig E. .
ANALYST, 2007, 132 (01) :21-23
[10]  