Double-AOTF-based aberration-free spectral imaging endoscopic system for biomedical applications

被引:26
作者
Machikhin, Alexander [1 ]
Pozhar, Vitold [1 ]
机构
[1] Sci & Technol Ctr Unique Instrumentat RAS, Moscow 117342, Russia
关键词
Spectral imaging; spectroscopy; double monochromatization; acousto-optical tunable filter; photoluminescence diagnostics; endoscopy; SPECTROMETER; FLUORESCENCE; DIFFRACTION;
D O I
10.1142/S1793545815410096
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The problem of in vivo photoluminescence diagnostics of the tissues accessible by endoscopes is discussed. The spectral imaging module attachable to conventional rigid and flexible medical endoscopes is developed and described. It is based on a double acousto-optical tunable filter (AOTF) and a specialized optical coupling system. The module provides wide field of view (FOV), absence of image distortions, random spectral access, fast spectral image acquisition at any wavelength in the visible range and accurate measurement of reflectance spectrum in each pixel of the image. Images of typical biomedical samples are presented and discussed. Their spectra are compared to the reference data.
引用
收藏
页数:6
相关论文
共 17 条
[1]   Fluorescence imaging spectroscopy utilising acousto-optic tuneable filters [J].
Bouhifd, M ;
Whelan, MP ;
Aprahamian, M .
OPTO-IRELAND 2005: OPTICAL SENSING AND SPECTROSCOPY, 2005, 5826 :185-193
[2]  
Gupta N., 2008, P SOC PHOTO-OPT INS, V6842, P1
[3]   AOTF-based Imaging spectrometer for research of small-size biological objects [J].
Kutuza, IB ;
Pozhar, VE ;
Pustovoit, VI .
NOVEL OPTICAL INSTRUMENTATION FOR BIOMEDICAL APPLICATIONS, 2003, 5143 :165-169
[4]   Spatial and spectral image distortions caused by diffraction of an ordinary polarised light beam by an ultrasonic wave [J].
Machihin, A. S. ;
Pozhar, V. E. .
QUANTUM ELECTRONICS, 2015, 45 (02) :161-165
[5]   A spectral distortion correction method for an imaging spectrometer [J].
Machihin, A. S. ;
Pozhar, V. E. .
INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 2009, 52 (06) :847-853
[6]   An optical system for coupling double acousto-optic monochromators and eyepieces of visual optical instruments [J].
Machikhin, A. S. ;
Batshev, V. I. .
INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 2014, 57 (06) :736-741
[7]   An acousto-optic endoscopic imaging spectrometer [J].
Machikhin, A. S. ;
Pozhar, V. E. ;
Batshev, V. I. .
INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 2013, 56 (04) :477-481
[8]  
Machikhin A. S., 2015, P SOC PHOTO-OPT INS, V9448, P1
[9]   Dual modality fluorescence and reflectance hyperspectral imaging: principle and applications [J].
Martin, ME ;
Wabuyele, MB ;
Panjehpour, M ;
Phan, MN ;
Overholt, BF ;
DeNovo, RC ;
Moyers, T ;
Song, SG ;
Vo-Dinh, T .
Advanced Biomedical and Clinical Diagnostic Systems III, 2005, 5692 :133-139
[10]  
Mazur M., 2006, ACHIEVEMENTS MOD RAD, V10, P19