Scatter radiation dose profile evaluation in computed tomography using Monte Carlo simulation

被引:0
|
作者
Mohammadi, Gh R. Fallah [1 ]
Hesamnezhad, L. [2 ]
Mahdavi, M. [2 ]
机构
[1] Mazandaran Univ Med Sci, Fac Allied Med, Dept Radiol, Sari, Iran
[2] Univ Mazandaran, Dept Phys, Babolsar, Iran
来源
INTERNATIONAL JOURNAL OF RADIATION RESEARCH | 2021年 / 19卷 / 04期
关键词
Dose profile; CT scan; scatter radiation; CTDI phantom; MC simulation; CONE-BEAM CT; DOSIMETRY;
D O I
10.29242/ijrr.19.4.813
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Conventional radiation dosimetry methods in computed tomography (CT) are not able to measure the dose distribution along the patient's longitudinal axis. To calculate the dose index on a CT scan, the dose distribution from the center of the radiation field must be calculated. In this study, the most appropriate integral interval for calculating the CT dose index in the axial mode was determined using the Monte Carlo (MC) method based on X-ray photon energy and slice thickness. Materials and Methods: The computed tomography dose index (CTDI) phantom was simulated in the EGSnrc/BEAMnrcMC system and was irradiated with several X-ray energies and several slice thicknesses and dose profiles in phantom were investigated. The area under the dose profile and the scatter to primary radiation dose ratio (SPR) were calculated. Results: The range of scattered beams from the center of the radiation field reaches 450 mm in 140 kV and a 40 mm slice thickness. The SPR value for all levels of X-ray photon energy (between 80 and 140 kV) significantly decreases as slice thickness increases. CT scan imaging technical factors greater than 310 mm from the center of the slice thickness have no effect on the behavior of the scattered radiation. Conclusion: The primary beams are more affected by the energy of the photons, and the scatter beams are more strongly affected by the slice thickness. For 64-slice scanners, the polymethyl methacrylate (PMMA) phantom length should be between 700 mm and 900 mm to yield accurate CTDI estimations.
引用
收藏
页码:813 / 818
页数:6
相关论文
共 50 条
  • [1] Estimation of Computed Tomography Dose Index (CTDI) Using Monte Carlo Simulation
    Talbi, Mohammed
    Aabid, Mohammed
    El Mansouri, M'hamed
    Nhila, Oussama
    Semghouli, Slimane
    Sebihi, Rajaa
    Choukri, Abdelmajid
    Khalis, Mohammed
    MOSCOW UNIVERSITY PHYSICS BULLETIN, 2021, 76 (SUPPL 1) : S103 - S109
  • [2] Estimation of Computed Tomography Dose Index (CTDI) Using Monte Carlo Simulation
    Mohammed Talbi
    Mohammed Aabid
    M’hamed El Mansouri
    Oussama Nhila
    Slimane Semghouli
    Rajaa Sebihi
    Abdelmajid Choukri
    Mohammed Khalis
    Moscow University Physics Bulletin, 2021, 76 : S103 - S109
  • [3] Evaluation of the effect of patient dose from cone beam computed tomography on prostate IMRT using Monte Carlo simulation
    Chow, James C. L.
    Leung, Michael K. K.
    Islam, Mohammad K.
    Norrlinger, Bernhard D.
    Jaffray, David A.
    MEDICAL PHYSICS, 2008, 35 (01) : 52 - 60
  • [4] Monte Carlo dose index estimation in computed tomography
    Ati, Moncef
    Bouamrane, Rachid
    Addi, Djamel
    Maroc, Fatima Zohra
    Mecheret, Fatima Zohra
    AIMS BIOENGINEERING, 2020, 7 (04): : 224 - 235
  • [5] Thorax organ dose estimation in computed tomography based on patient CT data using Monte Carlo simulation
    Mohammadi, Gh. R. Fallah
    Alam, N. Riyahi
    Geraily, Gh.
    Paydar, R.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2016, 14 (04): : 313 - 321
  • [6] Monte Carlo simulation of single beam dose profile
    LIU Xiaowei
    WEN Guanghao
    LI Mianfeng(Zhongshan University
    NuclearScienceandTechniques, 1999, (02) : 87 - 90
  • [7] INVESTIGATION OF A PRACTICAL PATIENT DOSE INDEX FOR ASSESSMENT OF PATIENT ORGAN DOSE FROM CONE-BEAM COMPUTED TOMOGRAPHY IN RADIATION THERAPY USING A MONTE CARLO SIMULATION
    Kamezawa, H.
    Arimura, H.
    Arakawa, H.
    Kameda, N.
    RADIATION PROTECTION DOSIMETRY, 2018, 181 (04) : 333 - 342
  • [8] A new Monte Carlo tool for organ dose estimation in computed tomography
    Adrien, C.
    Le Loirec, C.
    Dreuil, S.
    Bordy, J. -M.
    RADIOPROTECTION, 2020, 55 (02) : 123 - 134
  • [9] Radiation doses in volume-of-interest breast computed tomography-A Monte Carlo simulation study
    Lai, Chao-Jen
    Zhong, Yuncheng
    Yi, Ying
    Wang, Tianpeng
    Shaw, Chris C.
    MEDICAL PHYSICS, 2015, 42 (06) : 3063 - 3075
  • [10] Evaluation of Organ Doses for Pediatric Computed Tomography Using a Newly Designed Radiophotoluminescence Glass Dosimeter and Comparison with a Monte Carlo Simulation-based Dose Calculator
    Chang, Weishan
    Koba, Yusuke
    HEALTH PHYSICS, 2021, 120 (03): : 288 - 295