Evaluation of a 3D-Printed Transoral Robotic Surgery Simulator Utilizing Artificial Tissue

被引:2
|
作者
Murr, Alexander T. [1 ]
Lumley, Catherine J. [1 ]
Feins, Richard H. [2 ]
Hackman, Trevor G. [1 ]
机构
[1] Univ N Carolina, Dept Otolaryngol Head & Neck Surg, Chapel Hill, NC 27515 USA
[2] Univ N Carolina, Dept Surg, Chapel Hill, NC 27515 USA
基金
美国国家卫生研究院;
关键词
Transoral robotic surgery; surgical simulation; 3D printing; artificial tissue; TRAINING CURRICULUM; LEARNING-CURVE; SKILLS; VALIDATION; TOOL;
D O I
10.1002/lary.29981
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Objectives/Hypothesis Transoral robotic surgery (TORS) poses challenges for operators in training, with limited robot access on a platform requiring distinct surgical skills. Few simulators exist, and current virtual reality training modules exclude head and neck simulations. This study evaluates the construct validity for a novel low-cost TORS simulator. Study Design Single institution prospective observational study. Methods Using 3D-printed oral cavity structures and replaceable artificial tissue components, a modular TORS simulator was constructed for short-duration hands-on simulations with the da Vinci SI robot. Sixteen surgeons of differing robotic skill levels, no experience (novice), prior experience, and formal robot training, participated in simulated tonsil and tongue base tumor resections. Video recordings of each participant were graded by a blinded robotically trained surgeon using a 35-point Global Evaluative Assessment of Robotic Surgery (GEARS) criterion adapted for the TORS simulator. Results Operators reporting formal robotic training or prior robot experience achieved significantly higher mean total GEARS scores compared to novice operators (32 vs. 20.5; P < .001). Overall, mean total GEARS scores correlated with reported experience level; novice operators scored 54% of total points at 19 (4.5), operators with prior experience scored 82.3% of total points at 28.8 (2.6), and robotically trained operators scored 97.1% of total points at 34 (1.7). Conclusion With a GEARS criterion, our simulator successfully differentiated novice from experienced and robotically trained operators of the da Vinci SI robot during simulated tonsillectomy and base of tongue resections. These findings support the construct validity of this prototype simulator and offer a foundation for further testing of predictive validity. Level of Evidence 2 Laryngoscope, 2021
引用
收藏
页码:1588 / 1593
页数:6
相关论文
共 50 条
  • [1] 3D-printed pediatric endoscopic ear surgery simulator for surgical training
    Barber, Samuel R.
    Kozin, Elliott D.
    Dedmon, Matthew
    Lin, Brian M.
    Lee, Kyuwon
    Sinha, Sumi
    Black, Nicole
    Remenschneider, Aaron K.
    Lee, Daniel J.
    INTERNATIONAL JOURNAL OF PEDIATRIC OTORHINOLARYNGOLOGY, 2016, 90 : 113 - 118
  • [2] A novel 3D-printed simulator for hands-on training in orthognathic surgery
    Lukas Benedikt Seifert
    Christopher Groepper
    Rosa Rohin
    Daniel Thiem
    Philipp Becker
    Florian Markus Thieringer
    Robert Alexander Sader
    Maxillofacial Plastic and Reconstructive Surgery, 47 (1)
  • [3] 3D-Printed Artificial Microfish
    Zhu, Wei
    Li, Jinxing
    Leong, Yew J.
    Rozen, Isaac
    Qu, Xin
    Dong, Renfeng
    Wu, Zhiguang
    Gao, Wei
    Chung, Peter H.
    Wang, Joseph
    Chen, Shaochen
    ADVANCED MATERIALS, 2015, 27 (30) : 4411 - 4417
  • [4] Validation of a 3D-Printed Percutaneous Injection Laryngoplasty Simulator: A Randomized Controlled Trial
    Kostas, Julianna C.
    Lee, Andrew S. S.
    Arunkumar, Amit
    Han, Catherine
    Lee, Mark
    Goel, Alexander N.
    Alrassi, James
    Crosby, Tyler
    Clark, Christine M.
    Amin, Milan
    Abu-Ghanem, Sara
    Kirke, Diana
    Rameau, Anais
    LARYNGOSCOPE, 2024, 134 (01) : 318 - 323
  • [5] Toward the improvement of 3D-printed vessels' anatomical models for robotic surgery training
    Marconi, S.
    Negrello, E.
    Mauri, V
    Pugliese, L.
    Peri, A.
    Argenti, F.
    Auricchio, F.
    Pietrabissa, A.
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2019, 42 (10) : 558 - 565
  • [6] Evaluation of long-term stability of monolithic 3D-printed robotic manipulator structures for minimally invasive surgery
    Krieger, Yannick S.
    Ostler, Daniel
    Rzepka, Korbinian
    Meining, Alexander
    Feussner, Hubertus
    Wilhelm, Dirk
    Lueth, Tim C.
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2020, 15 (10) : 1693 - 1697
  • [7] 3D-printed tracheoesophageal puncture and prosthesis placement simulator
    Barber, Samuel R.
    Kozin, Elliott D.
    Naunheim, Matthew R.
    Sethi, Rosh
    Remenschneider, Aaron K.
    Deschler, Daniel G.
    AMERICAN JOURNAL OF OTOLARYNGOLOGY, 2018, 39 (01) : 37 - 40
  • [8] Evaluation of long-term stability of monolithic 3D-printed robotic manipulator structures for minimally invasive surgery
    Yannick S. Krieger
    Daniel Ostler
    Korbinian Rzepka
    Alexander Meining
    Hubertus Feussner
    Dirk Wilhelm
    Tim C. Lueth
    International Journal of Computer Assisted Radiology and Surgery, 2020, 15 : 1693 - 1697
  • [9] On the progress of 3D-printed hydrogels for tissue engineering
    Rigoberto C. Advincula
    John Ryan C. Dizon
    Eugene B. Caldona
    Robert Andrew Viers
    Francis Dave C. Siacor
    Reymark D. Maalihan
    Alejandro H. Espera
    MRS Communications, 2021, 11 : 539 - 553
  • [10] A 3D-printed, dynamic, patient-specific knee simulator
    Conconi, Michele
    Sancisi, Nicola
    Backus, Reid
    Argenti, Christian
    Shih, Albert J.
    RAPID PROTOTYPING JOURNAL, 2024, 30 (07) : 1380 - 1392