Intraoperative findings on ECAP-measurement: Normal or special case?

被引:28
作者
Mueller, Alexander [1 ]
Hocke, Thomas [2 ]
Mir-Salim, Parwis [1 ]
机构
[1] Vivantes Hearing Ctr, ORL Dept Friedrichshain Clin, Berlin, Germany
[2] Cochlear Deutschland GmbH & Co KG Hannover, Hannover, Germany
关键词
Cochlear implant; evoked compound action potential; imaging diagnostic; neural response telemetry; NRT; COCHLEAR IMPLANT; SCALA TYMPANI; SYSTEM; THRESHOLDS; DIMENSIONS; AUTONRT(TM); CONTOUR(TM); EXCITATION; PROFILES; SPREAD;
D O I
10.3109/14992027.2014.969410
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Objective: This study aimed to investigate the measurability and threshold level of electrically evoked compound action potentials (ECAPs) in order to provide a baseline for the development of diagnostic references to support aftercare procedures. Design: In this retrospective study, cochlear implant patients were grouped according to the electrode array type and preoperative diagnostic findings from radiological imaging. ECAP measurements were performed intraoperatively on 22 electrodes for each patient resulting in two comparisons: (1) normal vs. pathological findings based on imaging with same electrode array, (2) perimodiolar electrode vs. straight electrode array within the normal group. Study sample: The study sample consisted of 218 ears with a Nucleus (R) implant. Results: In the group with normal radiological imaging, the percentage of measurable ECAPs was higher than in the pathological group. The ECAP thresholds on 21 electrodes were significantly different between the two groups. Furthermore the thresholds were significantly lower for basal electrodes mainly for the two different electrode types. Conclusion: The pathological changes in the inner ear, the type of electrode array, and the electrode position affect the ECAP threshold. Further consideration suggests that a receiver operating characteristic curve can be derived from the results.
引用
收藏
页码:257 / 264
页数:8
相关论文
共 29 条
[11]   A psychophysical forward masking comparison of longitudinal spread of neural excitation in the Contour™ and straight Nucleus® electrode arrays [J].
Cohen, LT ;
Lenarz, T ;
Battmer, RD ;
von Saebelkampf, CB ;
Busby, PA ;
Cowan, RSC .
INTERNATIONAL JOURNAL OF AUDIOLOGY, 2005, 44 (10) :559-566
[12]   Clinical use of a system for the automated recording and analysis of electrically evoked compound action potentials (ECAPs) in cochlear implant patients [J].
Gaertner, Lutz ;
Lenarz, Thomas ;
Joseph, Gert ;
Buechner, Andreas .
ACTA OTO-LARYNGOLOGICA, 2010, 130 (06) :724-732
[13]  
Guedes Mariana Cardoso, 2005, Braz J Otorhinolaryngol, V71, P660
[14]   DIMENSIONS OF THE SCALA TYMPANI IN THE HUMAN AND CAT WITH REFERENCE TO COCHLEAR IMPLANTS [J].
HATSUSHIKA, S ;
SHEPHERD, RK ;
TONG, YC ;
CLARK, GM ;
FUNASAKA, S .
ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1990, 99 (11) :871-876
[15]   Comparison of EAP thresholds with MAP levels in the nucleus 24 cochlear implant: Data from children [J].
Hughes, ML ;
Brown, CJ ;
Abbas, PJ ;
Wolaver, AA ;
Gervais, JP .
EAR AND HEARING, 2000, 21 (02) :164-174
[16]  
Hughes MP, 2006, AUDIOLOGY ONLINE, P1
[17]  
IGARASHI M, 1976, ARCH OTOLARYNGOL, V102, P428
[18]  
Kubik S., 1987, ANATOMIE MENSCHEN, VIII, P625
[19]   TNRT profiles with the Nucleus Research Platform 8 system [J].
Lai, W. K. ;
Dillier, N. ;
Weber, B. P. ;
Lenarz, T. ;
Battmer, R. ;
Gantz, B. ;
Brown, C. ;
Cohen, N. ;
Waltzman, S. ;
Skinner, M. ;
Holden, L. ;
Cowan, R. ;
Busby, P. ;
Killian, M. .
INTERNATIONAL JOURNAL OF AUDIOLOGY, 2009, 48 (09) :645-654
[20]   Longitudinal behaviour of neural response telemetry (NRT) data and clinical implications [J].
Lai, WK ;
Aksit, M ;
Akdas, F ;
Dillier, N .
INTERNATIONAL JOURNAL OF AUDIOLOGY, 2004, 43 (05) :252-263