Rule-based controller based on suction detection for rotary blood pumps

被引:11
作者
Ferreira, Antonio [1 ]
Boston, J. Robert [1 ]
Antaki, James F. [1 ]
机构
[1] Univ Pittsburgh, Dept Elect Engn, Pittsburgh, PA 15260 USA
来源
2007 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-16 | 2007年
关键词
D O I
10.1109/IEMBS.2007.4353205
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A rule-based controller for rotary ventricular assist devices was developed to automatically regulate the pumping speed of the device without introducing suction in the ventricle. The control approach is based on a discriminant analysis function that detects the occurrence of suction, providing the input for the rule-based controller. This controller has been tested in simulations showing the ability to autonomously adjust pump flow according to the patient's level of activity, while sustaining adequate perfusion pressures. The performance of the system (suction detector and controller) was tested for several levels of activity and contractility state of the left ventricle, using a lumped parameter model of the circulatory system coupled with a left ventricular assist device. In an cases, the controller kept cardiac output and mean arterial pressure within acceptable physiologic ranges.
引用
收藏
页码:3978 / 3981
页数:4
相关论文
共 12 条
[1]   Hierarchical control of heart-assist devices [J].
Boston, JR ;
Antaki, JF ;
Simaan, MA .
IEEE ROBOTICS & AUTOMATION MAGAZINE, 2003, 10 (01) :54-64
[2]   A sensorless approach to control of a turbodynamic left ventricular assist system [J].
Choi, S ;
Antaki, JF ;
Boston, JR ;
Thomas, D .
IEEE TRANSACTIONS ON CONTROL SYSTEMS TECHNOLOGY, 2001, 9 (03) :473-482
[3]  
Choi SJ, 1997, P AMER CONTR CONF, P3714, DOI 10.1109/ACC.1997.609538
[4]  
Driankov D., 1993, INTRO FUZZY CONTROL
[5]  
Ferreira A., 2006, P 28 IEEE ANN INT C, P5382
[6]  
FERREIRA A, 2006, P IEEE ICASSP TOUL F, P1064
[7]  
Ferreira A, 2005, IEEE DECIS CONTR P, P897
[8]   Achieving physiologic perfusion with ventricular assist devices: Comparison of control strategies [J].
Giridharan, G ;
Pantalos, G ;
Koenig, S ;
Gillars, K ;
Skliar, M .
ACC: PROCEEDINGS OF THE 2005 AMERICAN CONTROL CONFERENCE, VOLS 1-7, 2005, :3823-3828
[9]   Cleveland Clinic continuous flow blood pump: Progress in development [J].
Golding, L ;
Medvedev, A ;
Massiello, A ;
Smith, W ;
Horvath, D ;
Kasper, R .
ARTIFICIAL ORGANS, 1998, 22 (06) :447-450
[10]  
MAMDANI EH, 1974, P I ELECTR ENG, V121, P1585, DOI DOI 10.1049/PIEE.1974.0328