Isobaric beam purification for high precision Penning trap mass spectrometry of radioactive isotope beams with SWIFT

被引:17
|
作者
Kwiatkowski, A. A. [1 ,2 ]
Bollen, G. [1 ,2 ]
Redshaw, M. [1 ]
Ringle, R. [1 ]
Schwarz, S. [1 ]
机构
[1] Michigan State Univ, Natl Superconducting Cyclotron Lab, E Lansing, MI 48824 USA
[2] Michigan State Univ, Dept Phys & Astron, E Lansing, MI 48824 USA
基金
美国国家科学基金会;
关键词
Penning trap; Mass spectrometry; Beam purification; Atomic mass; Radioactive isotope beams; ION-CYCLOTRON-RESONANCE; FOURIER-TRANSFORM EXCITATION; QUADRUPOLE EXCITATION; MOTION; LEBIT; ACCURACY; JYFLTRAP; BUNCHER; SYSTEM; COOLER;
D O I
10.1016/j.ijms.2014.09.016
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
The Stored Waveform Inverse Fourier Transform (SWIFT) technique offers general and versatile broadband, mass-selective excitation schemes. The method has hitherto been unharnessed for high precision Penning trap mass spectrometry, in which the atomic mass is determined, and is of particular importance at radioactive isotope beam facilities. Herein, we present the first implementation of SWIFT to clean isobaric contaminants at such a spectrometer. The isobaric separation of CO+ and N-2(+) as well as a high-accuracy mass measurement with SWIFT cleaning were demonstrated at the Low Energy Beam and Ion Trap (LEBIT) facility; resolving powers of 4 x 10(4) (Delta nu = 300 Hz) and 2 x 10(5) (Delta nu = 60 Hz) were achieved for Na-23(+) for cleaning times of 20 and 100 ms respectively. These measurements demonstrate the viability of SWIFT for fast, efficient, and universal beam purification at high-precision Penning trap mass spectrometers. (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:9 / 15
页数:7
相关论文
共 50 条
  • [31] A multi-reflection time-of-flight mass separator for isobaric purification of radioactive ion beams
    Robert N. Wolf
    Markus Eritt
    Gerrit Marx
    Lutz Schweikhard
    Hyperfine Interactions, 2011, 199 : 115 - 122
  • [32] High precision isotope ratio monitoring techniques in mass spectrometry
    Brand, WA
    JOURNAL OF MASS SPECTROMETRY, 1996, 31 (03): : 225 - 235
  • [33] LEBIT II: Upgrades and developments for high precision Penning trap mass measurements with rare isotopes
    Redshaw, M.
    Bollen, G.
    Bustabad, S.
    Kwiatkowski, A. A.
    Lincoln, D. L.
    Novario, S. J.
    Ringle, R.
    Schwarz, S.
    Valverde, A. A.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2013, 317 : 510 - 516
  • [34] The Ramsey method in high-precision mass spectrometry with Penning traps: Theoretical foundations
    Kretzschmar, Martin
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2007, 264 (2-3) : 122 - 145
  • [35] The Ramsey method in high-precision mass spectrometry with Penning traps: Experimental results
    George, S.
    Blaum, K.
    Herfurth, F.
    Herlert, A.
    Kretzschmar, M.
    Nagy, S.
    Schwarz, S.
    Schweikhard, L.
    Yazidjian, C.
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2007, 264 (2-3) : 110 - 121
  • [36] Precision mass measurements using a penning trap and highly charged ions produced in an electron beam ion source
    Borgenstrand, H
    Carlberg, C
    Rouleau, G
    Schuch, R
    Soderberg, F
    Beebe, E
    Bergstrom, I
    Liljeby, L
    Paal, A
    Pikin, A
    Bollen, G
    Hartmann, H
    Jertz, R
    Kluge, HJ
    Schwarz, T
    Senne, P
    Mann, R
    PHYSICA SCRIPTA, 1997, T71 : 88 - 95
  • [37] High precision tungsten isotope measurement by thermal ionization mass spectrometry
    Touboul, Mathieu
    Walker, Richard J.
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2012, 309 : 109 - 117
  • [38] Two ions in one trap: Ultra-high precision mass spectrometry?
    Rainville, S
    Thompson, JK
    Pritchard, DE
    2002 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS, CONFERENCE DIGEST, 2002, : 318 - 319
  • [39] Two ions in one trap: Ultra-high precision mass spectrometry?
    Rainville, S
    Thompson, JK
    Pritchard, DE
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2003, 52 (02) : 292 - 296
  • [40] A high intensity electron beam ion trap for charge state boosting of radioactive ion beams
    Marrs, RE
    Slaughter, DR
    APPLICATION OF ACCELERATORS IN RESEARCH AND INDUSTRY, PTS 1 AND 2, 1999, 475 : 322 - 325