Low vulnerability of the right phrenic nerve to electroporation ablation

被引:145
作者
van Driel, Vincent J. H. M. [1 ]
Neven, Kars [1 ,2 ]
van Wessel, Harri [1 ,3 ]
Vink, Aryan [4 ]
Doevendans, Pieter A. F. M. [1 ,5 ]
Wittkampf, Fred H. M. [1 ]
机构
[1] Univ Med Ctr Utrecht, Dept Cardiol, NL-3508 GA Utrecht, Netherlands
[2] Alfried Krupp Krunkenhaus, Dept Rhythmol, Essen, Germany
[3] St Jude Med, Veenendaal, Netherlands
[4] Univ Med Ctr Utrecht, Dept Pathol, NL-3508 GA Utrecht, Netherlands
[5] ICIN Netherlands Heart Inst, Utrecht, Netherlands
关键词
Ablation; Nervous system; Irreversible electroporation; Phrenic nerve; Complication; CATHETER ABLATION; IRREVERSIBLE ELECTROPORATION; INJURY; MECHANISMS;
D O I
10.1016/j.hrthm.2015.05.012
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND Circular electroporation ablation is a novel ablation modality for electrical pulmonary vein isolation. With a single 200-3 application, deep circular myocardial lesions can be created. However, the acute and chronic effects of this energy source on phrenic nerve (PN) function are unknown. OBJECTIVE The purpose of this study was to analyze nerve vulnerability to electroporation ablation in a porcine model. METHODS In 20 animals (60-75 kg), the course of the right PN was pace-mapped inside the superior caval vein (SCV). Thereafter, a single 200-3 circular electroporation ablation was performed via a multipolar circular catheter in firm contact with the inner SCV wall. RESULTS In 19 of 20 animals, the PN could be captured along an estimated 6-8 cm trajectory above the right atrial contour. Directly after the application, the PN could be captured above the ablation Level in 17 of 19 animals and after maximally 30 minutes in all animals. Fifteen animals were restudied after 3-13 weeks, and PN functionality was unaffected in all. Histological analysis in 5 animals in which the application had been delivered in the muscular sleeve just above the right atrium showed a transmural circular lesion. However, no lesion was found in the other animals in which the application had been delivered in the fibrous section more cranial in the SCV. CONCLUSIONS Electroporation ablation at an energy Level that may create deep myocardial Lesions may spare the targeted right PN. These animal data suggest that electroporation may be a safe ablation modality near the right PN.
引用
收藏
页码:1838 / 1844
页数:7
相关论文
共 19 条
[1]  
[Anonymous], 2011, GUID CAR US LAB AN
[2]   Phrenic nerve injury after catheter ablation: Should we worry about this complication? [J].
Bai, Rong ;
Patel, Dimpi ;
Di Biase, Luigi ;
Fahmy, Tamer S. ;
Kozeluhova, Marketa ;
Prasad, Subramanya ;
Schweikert, Robert ;
Cummings, Jennifer ;
Saliba, Walid ;
Andrews-Williams, Michelle ;
Themistoclakis, Sakis ;
Bonso, Aldo ;
Rossillo, Antonio ;
Raviele, Antonio ;
Schmitt, Claus ;
Karch, Martin ;
Uriarte, Jorge A. Salerno ;
Tchou, Patrick ;
Arruda, Mauricio ;
Natale, Andrea .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2006, 17 (09) :944-948
[3]   Cardiac Ablation via Electroporation [J].
Hong, Jinback ;
Stewart, Mark T. ;
Cheek, Daniel S. ;
Francischelli, David E. ;
Kirchhof, Nicole .
2009 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-20, 2009, :3381-3384
[4]   Cryoballoon Ablation of Atrial Fibrillation [J].
Kuck, Karl-Heinz ;
Fuernkranz, Alexander .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2010, 21 (12) :1427-1431
[5]   A novel nonthermal energy source for surgical epicardial atrial ablation: Irreversible electroporation [J].
Lavee, Jacob ;
Onik, Gary ;
Mikus, Paul ;
Rubinsky, Boris .
HEART SURGERY FORUM, 2007, 10 (02) :E162-E167
[6]   Biophysical injury mechanisms in electrical shock trauma [J].
Lee, RC ;
Zhang, DJ ;
Hannig, J .
ANNUAL REVIEW OF BIOMEDICAL ENGINEERING, 2000, 2 :477-+
[7]   The Effects of Irreversible Electroporation (IRE) on Nerves [J].
Li, Wei ;
Fan, Qingyu ;
Ji, Zhenwei ;
Qiu, Xiuchun ;
Li, Zhao .
PLOS ONE, 2011, 6 (04)
[8]   Fatal End of a Safety Algorithm for Pulmonary Vein Isolation With Use of High-Intensity Focused Ultrasound [J].
Neven, Kars ;
Schmidt, Boris ;
Metzner, Andreas ;
Otomo, Kiyoshi ;
Nuyens, Dieter ;
De Potter, Tom ;
Chun, K. R. Julian ;
Ouyang, Feifan ;
Kuck, Karl-Heinz .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2010, 3 (03) :260-265
[9]   Irreversible electroporation: Implications for prostate ablation [J].
Onik, Gary ;
Mikus, Paul ;
Rubinsky, Boris .
TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2007, 6 (04) :295-300
[10]  
Rubinsky B, 2010, SER BIOMED ENG, P1, DOI 10.1007/978-3-642-05420-4