共 12 条
Stopping power measurements of heavy ions (3 ≤ Z1 ≤ 14) in Mylar foil by time-of-flight spectrometry
被引:14
|作者:
Ammi, H.
[1
]
Pineda-Vargas, C. A.
[2
,3
]
Mammeri, S.
[1
]
Msimanga, M.
[2
]
Ourabah, S.
[1
]
Dib, A.
[1
]
机构:
[1] Ctr Rech Nucl Alger, 2 Bd Frantz Fanon,BP 399, Algiers, Algeria
[2] IThemba Labs, ZA-7129 Cape Town, South Africa
[3] CPUT, Fac Hlth & Wellness Sci, ZA-7535 Bellville, South Africa
来源:
关键词:
Stopping power;
Time-of-flight;
Energy elastic recoil detection analysis;
ENERGY-LOSS MEASUREMENTS;
O-16;
IONS;
POLYPROPYLENE FOILS;
KAPTON;
HAVAR;
NICKEL;
C-12;
POLYCARBONATE;
SOLIDS;
MATTER;
D O I:
10.1016/j.nimb.2010.12.020
中图分类号:
TH7 [仪器、仪表];
学科分类号:
0804 ;
080401 ;
081102 ;
摘要:
Heavy ions elastic recoil detection analysis coupled with time of flight spectrometer (HIERDA_ToF-E) have been used to measure energy loss of charged particles in thin absorber. The stopping power of heavy ions has been determined in Mylar for Si-28, (27)m, (24)mg, F-19, O-16, C-12 and Li-7 ions over a continuous range of energies 0.14-0.80 MeV/nucleon. The ions were recoils from the bombardment of different samples (Si, MgO. Al2O3, LiF and C) with a 27.5 MeV Kr+ beam. The energy loss of the recoil atoms is measured with and without additional foils placed in front of a Surface Barrier Detector (SBD). The energy of individual ions is determined from its ToF data; the exit energy after the stopping foil is measured using the SBD detector. We have compared our stopping values to those predicted by SRIM-2008 computer code, ICRU-73 stopping data tables, MSTAR calculations and to the published data from literature. The results show good agreement with limited existing data but indicate a large deviation among the predicted theoretical values at the low energy side of the stopping maximum peak. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:386 / 391
页数:6