High-energy x-ray backlighter spectrum measurements using calibrated image plates

被引:153
作者
Maddox, B. R. [1 ]
Park, H. S. [1 ]
Remington, B. A. [1 ]
Izumi, N. [1 ]
Chen, S. [1 ]
Chen, C. [1 ]
Kimminau, G. [2 ]
Ali, Z. [3 ]
Haugh, M. J. [3 ]
Ma, Q. [4 ]
机构
[1] Lawrence Livermore Natl Lab, Livermore, CA 94550 USA
[2] Univ Oxford, Dept Phys, Clarendon Lab, Oxford OX1 3PU, England
[3] Natl Secur Technol LLC, Livermore, CA 94550 USA
[4] Argonne Natl Lab, DND CAT, Argonne, IL 60439 USA
关键词
DEPENDENCE;
D O I
10.1063/1.3531979
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The x-ray spectrum between 18 and 88 keV generated by a petawatt laser driven x-ray backlighter target was measured using a 12-channel differential filter pair spectrometer. The spectrometer consists of a series of filter pairs on a Ta mask coupled with an x-ray sensitive image plate. A calibration of Fuji (TM) MS and SR image plates was conducted using a tungsten anode x-ray source and the resulting calibration applied to the design of the Ross pair spectrometer. Additionally, the fade rate and resolution of the image plate system were measured for quantitative radiographic applications. The conversion efficiency of laser energy into silver K alpha x rays from a petawatt laser target was measured using the differential filter pair spectrometer and compared to measurements using a single photon counting charge coupled device. (C) 2011 American Institute of Physics. [doi:10.1063/1.3531979]
引用
收藏
页数:10
相关论文
共 15 条
[1]   Bremsstrahlung and Kα fluorescence measurements for inferring conversion efficiencies into fast ignition relevant hot electrons [J].
Chen, C. D. ;
Patel, P. K. ;
Hey, D. S. ;
Mackinnon, A. J. ;
Key, M. H. ;
Akli, K. U. ;
Bartal, T. ;
Beg, F. N. ;
Chawla, S. ;
Chen, H. ;
Freeman, R. R. ;
Higginson, D. P. ;
Link, A. ;
Ma, T. Y. ;
MacPhee, A. G. ;
Stephens, R. B. ;
Van Woerkom, L. D. ;
Westover, B. ;
Porkolab, M. .
PHYSICS OF PLASMAS, 2009, 16 (08)
[2]   Image plates as x-ray detectors in plasma physics experiments [J].
Gales, SG ;
Bentley, CD .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (10) :4001-4003
[3]   X-RAY-ENERGY DEPENDENCE AND UNIFORMITY OF AN IMAGING PLATE DETECTOR [J].
ITO, M ;
AMEMIYA, Y .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 310 (1-2) :369-372
[4]   Application of imaging plates to x-ray imaging and spectroscopy in laser plasma experiments (invited) [J].
Izumi, N. ;
Snavely, R. ;
Gregori, G. ;
Koch, J. A. ;
Park, H. -S. ;
Remington, B. A. .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2006, 77 (10)
[5]   DESIGN OF AN OPTIMAL ROSS FILTER SYSTEM FOR X-RAY-SPECTRA MEASUREMENTS IN THE RANGE OF 8.98-88-KEV [J].
KHUTORETSKY, IV .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (01) :773-775
[6]   Calibration and characterization of single photon counting cameras for short-pulse laser experiments [J].
Maddox, B. R. ;
Park, H. S. ;
Remington, B. A. ;
McKernan, M. .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2008, 79 (10)
[7]   Evaluation of the sensitivity and fading characteristics of an image plate system for x-ray diagnostics [J].
Meadowcroft, A. L. ;
Bentley, C. D. ;
Stott, E. N. .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2008, 79 (11)
[8]   RADIOACTIVITY AND GEOMETRICAL DISTRIBUTION MEASUREMENTS OF ALPHA-EMITTER SPECIMENS WITH THE IMAGING PLATE [J].
MORI, C ;
MATSUMURA, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1992, 312 (1-2) :39-42
[9]   Functional equation for the fading correction of imaging plates [J].
Ohuchi, H ;
Yamadera, A ;
Nakamura, T .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2000, 450 (2-3) :343-352
[10]   Dependence of fading patterns of photo-stimulated luminescence from imaging plates on radiation, energy, and image reader [J].
Ohuchi, H ;
Yamadera, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 490 (03) :573-582